Skip to content

Depuy P.F.C. E Knee System TC3 Tibial Insert, 2.5, 30.0mm… recallZ-0399-2008 · Depuy Orthopaedics, Inc. · initiated Sep 27, 2007

Terminated All recalls posted in December 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Depuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.

FDA root cause category: Packaging process control

What the company is doing

Consignee hospitals and implanting physicians were notified of the recall via an Urgent Information-Recall Notice letter and Dear Doctor letters dated 9/27/07. The hospitals were instructed to determine whether they had any of the recalled products and to contact their sales representative to pick up the product. Depuy sales representatives were notified to collect the products from consignee hospitals via email on 10/1/07. The decision to discuss this product defect with patients who had already been implanted was left to the implanting physicians.

Recall details

Recall number
Z-0399-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 27, 2007
Posted by the FDA
Dec 6, 2007
Recalling firm
Depuy Orthopaedics, Inc.
Quantity
12
Product code
NJL
FDA submissions
K952830
Distribution
Nationwide.

Product description

Depuy P.F.C. E Knee System TC3 Tibial Insert, 2.5, 30.0mm, 67mm M/L, 45 mm A/P, 30.0 mm, UHMWPE/Ti-6AI-4V, sterile; Product Code/REF 96-0527.

Code information (models, lots, serial numbers)

Lot 79272A.

Other devices under the same FDA submission

76 devices cleared under K952830 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 44893).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page