Depuy P.F.C. E Knee System Stabilized Plus Tibial Insert… recallZ-0425-2008 · Depuy Orthopaedics, Inc. · initiated Sep 27, 2007
Terminated All recalls posted in December 2007
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Depuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
FDA root cause category: Packaging process control
What the company is doing
Consignee hospitals and implanting physicians were notified of the recall via an Urgent Information-Recall Notice letter and Dear Doctor letters dated 9/27/07. The hospitals were instructed to determine whether they had any of the recalled products and to contact their sales representative to pick up the product. Depuy sales representatives were notified to collect the products from consignee hospitals via email on 10/1/07. The decision to discuss this product defect with patients who had already been implanted was left to the implanting physicians.
Recall details
- Recall number
- Z-0425-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 27, 2007
- Posted by the FDA
- Dec 6, 2007
- Recalling firm
- Depuy Orthopaedics, Inc.
- Quantity
- 12
- Product code
- NJL
- FDA submissions
- K963117
- Distribution
- Nationwide.
Product description
Depuy P.F.C. E Knee System Stabilized Plus Tibial Insert (STAB PLUS), 71 mm M/L, 47 mm A/P, 15 mm, 3, UHMWPE, sterile; Product Code/REF 96-2732.
Code information (models, lots, serial numbers)
Lot 79063A.
Other devices under the same FDA submission
12 devices cleared under K963117 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- P.F.C. SIGMA 96-2731Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- P.F.C. SIGMA 96-2736Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- P.F.C. SIGMA 96-2721Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- P.F.C. SIGMA 96-2722Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- P.F.C. SIGMA 96-2730Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- P.F.C. SIGMA 96-2732Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- P.F.C. SIGMA 96-2734Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- P.F.C. SIGMA 96-2735Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- P.F.C. SIGMA 96-2741Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- P.F.C. SIGMA 96-2742Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- P.F.C. SIGMA 96-2745Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- P.F.C. SIGMA 96-2754Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 44893).
- Z-0385-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 67 mm M/L, 45 mm A/P, 8.0 mm, 2.5, sterile; Product Code/REF 96-0430.
- Z-0386-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 67 mm M/L, 45 mm A/P, 10.0 mm, 2.5, UHMWPE, sterile; Product Code/REF 96-0431.
- Z-0387-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 67 mm M/L, 45 mm A/P, 12.5 mm, 2.5, UHMWPE, sterile; Product Code/REF 96-0432.
- Z-0388-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 67 mm M/L, 45 mm A/P, 15.0 mm, 2.5, UHMWPE, sterile; Product Code/REF 96-0433.
- Z-0389-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 71 mm M/L, 47 mm A/P, 8.0 mm, 3, UHMWPE, sterile; Product Code/REF 96-0440.
- Z-0390-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 71 mm M/L, 47 mm A/P, 10.0 mm, 3, UHMWPE, sterile; Product Code/REF 96-0441.
- Z-0391-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 71 mm M/L, 47 mm A/P, 12.5 mm, 3, UHMWPE, sterile; Product Code/REF 96-0442.
- Z-0392-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 76 mm M/L, 51 mm A/P, 10.0 mm, 4, UHMWPE, sterile; Product Code/REF 96-0451.
- Z-0393-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 76 mm M/L, 51 mm A/P, 12.5 mm, 4, UHMWPE, sterile; Product Code/REF 96-0452.
- Z-0394-2008 · TerminatedDepuy P.F.C. E Knee SystemStabilized Tibial Insert (STAB), 76 mm M/L, 51 mm A/P, 15.0 mm, 4, UHMWPE, sterile; Product Code/REF 96-0453.
- Z-0395-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Tibial Insert (STAB), 89 mm M/L, 59 mm A/P, 10.0 mm, 6, UHMWPE, sterile; Product Code/REF 96-0471.
- Z-0396-2008 · TerminatedDepuy P.F.C. E Knee System TC3 Tibial Insert, 2.5, 12.5mm, 67 mm M/L, 45 mm A/P, UHMWPE/Ti-6AI-4V, sterile; Product Code/REF 96-0521.
- Z-0397-2008 · TerminatedDepuy P.F.C. E Knee System TC3 Tibial Insert, 2.5, 15.0mm, 67 mm M/L, 45 mm A/P, UHMWPE/Ti-6AI-4V, sterile; Product Code/REF 96-0522.
- Z-0398-2008 · TerminatedDepuy P.F.C. E Knee System Depuy P.F.C. E Knee System TC3 Tibial Insert, 2.5, 17.5mm, 67 mm M/L, 45 mm A/P, UHMWPE/Ti-6AI-4V, sterile; Product Code/REF 96-0523.
- Z-0399-2008 · TerminatedDepuy P.F.C. E Knee System TC3 Tibial Insert, 2.5, 30.0mm, 67mm M/L, 45 mm A/P, 30.0 mm, UHMWPE/Ti-6AI-4V, sterile; Product Code/REF 96-0527.
- Z-0400-2008 · TerminatedDepuy P.F.C. E Knee System TC3 Tibial Insert, 4, 20.0mm, 75mm M/L, 51mm A/P, UHMWPE/Ti-6AI-4V, sterile; Product Code/REF 96-0544.
- Z-0401-2008 · TerminatedDepuy P.F.C. E Knee System TC3 Tibial Insert, 5, 12.5mm, 83mm M/L, 55mm A/P, UHMWPE/Ti-6AI-4V, sterile; Product Code/REF 96-0551.
- Z-0402-2008 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 2.5, 8 mm, 67mm M/L, 45 mm A/P, UHMWPE, 2.5, sterile; Product Code/REF 96-0630.
- Z-0403-2008 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 3, 8 mm, 71mm M/L, 47 mm A/P, UHMWPE, sterile; Product Code/ REF 96-0640.
- Z-0404-2008 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 3, 10mm, 71mm M/L, 47mm A/P, UHMWPE, sterile; Product Code/REF 96-0641.
- Z-0405-2008 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 4, 8mm, 76mm M/L, 51mm A/P, UHMWPE, sterile; Product Code/REF 96-0650.
- Z-0406-2008 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 4, 10mm, 76mm M/L, 51mm A/P, UHMWPE, sterile; Product Code/REF 96-0651.
- Z-0407-2008 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 5, 15mm, 83mm M/L, 55mm A/P, UHMWPE, sterile; Product Code/REF 96-0663.
- Z-0418-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized All-Polyethylene Tibia, 71mm M/L, 47mm A/P, 15.0mm, 3, UHMWPE, sterile; REF Product Code/96-2643.
- Z-0419-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized All-Polyethylene Tibia, 76mm M/L, 51mm A/P, 8.0mm, 4, UHMWPE, sterile; Product Code/REF 96-2650.
- Z-0420-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized All-Polyethylene Tibia, 76mm M/L, 51mm A/P, 12.5mm, 4, UHMWPE, sterile; Product Code/REF 96-2652.
- Z-0421-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Plus Tibial Insert (STAB PLUS), 64 mm M/L, 43 mm A/P, 15 mm, 2, UHMWPE, sterile; Product Code/REF 96-2712.
- Z-0422-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Plus Tibial Insert (STAB PLUS), 64 mm M/L, 43 mm A/P, 17.5 mm, 2, UHMWPE, sterile; Product Code/REF 96-2713.
- Z-0423-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Plus Tibial Insert (STAB PLUS), 71 mm M/L, 47 mm A/P, 10 mm, 3, UHMWPE, sterile; Product Code/REF 96-2730.
- Z-0424-2008 · TerminatedDepuy P.F.C. E Knee System Stabilized Plus Tibial Insert (STAB PLUS), 71 mm M/L, 47 mm A/P, 12.5 mm, 3, UHMWPE, sterile; Product Code/REF 96-2731.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page