VITALITY DR+ Implantable Cardioverter Defibrillator (ICD)… recallZ-1513-2008 · Boston Scientific CRM Corp · initiated Jan 7, 2008
Terminated All recalls posted in August 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific CRM Corp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Repetitive mechanical stress applied to the titanium case can induce component damage and device malfunction in certain ICD and CRT-D devices implanted subpectorally with the serial number facing the ribs. Devices implanted subcutaneously or in a subpectoral position with the serial number facing away from the ribs are not included in this advisory. This update identifies additional VITALITY ICD
FDA root cause category: Device Design
What the company is doing
An Urgent Medical Device Information Update to May 2006 Product Advisory, dated January 04, 2008, was sent to implanting physicians and physicians following implanted devices beginning Monday, January 07, 2008. The letter described a malfunction that can occur when the ICD is implanted subpectorally with the serial number facing the ribs. Devices implanted subcutaneously or in a subpectoral position with the serial number facing away from the ribs are not included in this advisory. The physicians were asked to review records to determine if an affected device was implanted subpectorally. An AP radiograph should be done to determine specific device orientation for subpectoral implants. If a subpectoral implant is in a susceptible orientation, consider repositioning or replacement for physically active patients who regularly need device therapy. Follow patients with susceptible devices at 3-month intervals.
Recall details
- Recall number
- Z-1513-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jan 7, 2008
- Posted by the FDA
- Aug 21, 2008
- Recalling firm
- Boston Scientific CRM Corp
- Quantity
- BSC CRM estimates that the number of devices implanted in a susceptible orientation is likely less than 1% of the total active implants 25,000 (w.w.)
- Product code
- LWP
- FDA submissions
- P960040
- Distribution
- Nationwide and Worldwide: Bolivia, Panama, Iran, Saudi Arabia, Kuwait, United Arab Rep., Austria, Belgium, Cyrus, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, Ireland, Italy, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland, UK, India, Australia, New Zealand, Malaysia, Singapore, Hong Kong, China, Israel, Turkey, Thailand, and Canada.
Product description
VITALITY DR+ Implantable Cardioverter Defibrillator (ICD), model 1872
Code information (models, lots, serial numbers)
Devices that are implanted subpectorally with the serial number facing the ribs. Implant orientation is not reported to Boston Scientific CRM.
Other devices under the same FDA submission
45 devices cleared under P960040 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- DYNAGEN™ EL ICD DR D152Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)12 deaths1,208 reports
- INOGEN™ EL ICD DR D142Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)6 deaths1,017 reports
- RESONATE™ EL ICD DR D433Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths965 reports
- DYNAGEN™ EL ICD DR D153Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths886 reports
- MOMENTUM™ EL ICD DR D121Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 deaths850 reports
- DYNAGEN™ EL ICD VR D150Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)8 deaths831 reports
- INOGEN™ EL ICD VR D140Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths823 reports
- ENERGEN™ ICD E140Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)3 deaths819 reports
- VIGILANT™ EL ICD VR D232Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)8 deaths750 reports
- ENERGEN™ ICD E142Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)1 death705 reports
- INCEPTA™ ICD E160Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 deaths601 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 46297).
- Z-1514-2008 · TerminatedVITALITY EL, models T127, T167, T177 Implantable Cardioverter Defibrillator (ICD), pulse-generator, dual Chamber
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page