CalMed Pericardial Sump, with 1/8"/0.3175 cm Connector… recallZ-1175-2008 · California Medical Laboratories Inc · initiated Jan 24, 2008
Terminated All recalls posted in June 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from California Medical Laboratories Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Fragment of material: The product may have a loose, stainless steel stringer flash located at the distal tip of the product.
FDA root cause category: Process control
What the company is doing
CalMed emailed letters to consignees on 1/24/2008 notifying customers of the recall. The letter advises that that the following Medical Devices are subject to an immediate Voluntary Recall for Correction: Pericardial Sump (Catalog #SU-20601), Pericardial Sump (Catalog #SU-20602), and Malleable Pericardial Sump (Catalog #SU-20802). It is asked that further distribution or use of any remaining product should cease immediately. Where appropriate, the firm asks that consignees notify their own customers who received the product of the same. The Product is being recalled due to a potentially loose, stainless steel stringer flash located at the distal tip of the Product. Customers are asked to contact the firm's Quality Assurance Manager immediately for purposes of issuing a Returned Goods Authorization and coordinating the return of the Subject Medical Devices such that they may be inspected and nonconforming product removed. A response is requested via e-mail.
Recall details
- Recall number
- Z-1175-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jan 24, 2008
- Posted by the FDA
- Jun 11, 2008
- Recalling firm
- California Medical Laboratories Inc
- Quantity
- 31,124 units of all models
- Product code
- DTS
- FDA submissions
- K982891
- Distribution
- Worldwide Distribution - USA, Turkey, Brazil, Germany, Italy, Spain, Philippines, South Africa, Japan, Israel, Saudi Arabia, United Arab Emirates, India, Ireland, China, Hungary, Poland, Taiwan, Belgium, Pakistan, Lithuania, Greece, Canada, Syria, Israel, Austria, Denmark, Bangladesh, Costa Rica, Mexico, Spain, and Sultanate of Oman.
Product description
CalMed Pericardial Sump, with 1/8"/0.3175 cm Connector Single Use Only. Do Not Reuse. REF SU-20601; CalMed Laboratories, Costa Mesa, CA 92626
Code information (models, lots, serial numbers)
Catalog Number SU-20601, Lot Number S03482
Other devices under the same FDA submission
4 devices cleared under K982891 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Suction tip SU-xxxxxLivanova USA, Inc. · SUCKER, CARDIOTOMY RETURN, CARDIOPULMONARY BYPASS9 reports
- Suction tip SU-xxxxxLivanova USA, Inc. · SUCKER, CARDIOTOMY RETURN, CARDIOPULMONARY BYPASS6 reports
- Suction tip SU-xxxxxLivanova USA, Inc. · SUCKER, CARDIOTOMY RETURN, CARDIOPULMONARY BYPASS2 reports
- Suction tip SU-xxxxxLivanova USA, Inc. · SUCKER, CARDIOTOMY RETURN, CARDIOPULMONARY BYPASS2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 46745).
- Z-1176-2008 · TerminatedCalMed Pericardial Sump, with 1/4"/0.635 cm Connector, Single Use Only. Do Not Reuse; REF SU-20602; CalMed Laboratories, Costa Mesa, CA 92626
- Z-1177-2008 · TerminatedCalMed Malleable Pericardial Sump, Single Use Only. Do Not Reuse. REF SU-20802; CalMed Laboratories, Costa Mesa, CA 92626
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page