GE Healthcare Proteus XR/a Radiographic X-ray System, Model… recallZ-0842-2008 · GE Healthcare · initiated Nov 1, 2007
Terminated All recalls posted in June 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The warning label that is required by 21 CFR 1020.30(j) was not on the control console of the Proteus XR/a Radiographic X-ray System, Model 2259976.
FDA root cause category: Process control
What the company is doing
GE issued a Field Modification Instruction to their field service representatives, released in December 2007, with instructions to replace the labels on the console with the appropriate labeling. Consignees will be visited by GE field service representatives beginning December 2007. All consignees will have the required labels affixed to their units by July 31, 2008.
Recall details
- Recall number
- Z-0842-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Nov 1, 2007
- Posted by the FDA
- Jun 10, 2008
- Recalling firm
- GE Healthcare
- Quantity
- 400 Units
- Product code
- KPR
- FDA submissions
- K993090
- Distribution
- Worldwide Distribution, including USA, Canada, Mexico, France, Hong Kong, Italy, Korea, China, Poland, UK, New Zealand, Saudi Arabia, Taiwan, Turkey, and Australia.
Product description
GE Healthcare Proteus XR/a Radiographic X-ray System, Model Number 2259976; previously marketed as the Silhouette FC, GE Healthcare, Waukesha, WI 53188.
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
00000010838HL1, 00000011061HL9, 00000011067HL6, 00000011185HL6, 00000012500HL5, 00000012667HL2, 00000013129HL2, 00000013372HL8, 00000013385HL0, 00000013543HL4, 00000013557HL4, 00000013800HL8, 00000014099HL6, 00000014104HL4, 00000014319HL8, 00000014327HL1, 00000014667HL0, 00000014677HL9, 00000015076HL3, 00000015383HL3, 00000015386HL6, 00000015400HL5, 00000015738HL8, 00000015744HL6, 00000015745HL3, 00000015750HL3, 00000015760HL2, 00000016174HL5, 00000016383HL2, 00000016386HL5, 00000016696HL7, 00000016707HL2, 00000016708HL0, 00000016709HL8, 00000016930HL0, 00000016954HL0, 00000016955HL7, 00000017119HL9, 00000017130HL6, 00000017131HL4, 00000017476HL3, 00000017477HL1, 00000017478HL9, 00000017480HL5, 00000017490HL4, 00000017492HL0, 00000017493HL8, 00000017599HL2, 00000017604HL0, 00000017608HL1, 00000017610HL7, 00000017613HL1, 00000017616HL4, 00000017788HL1, 00000017801HL2, 00000018476HL2, 00000018479HL6, 00000018483HL8, 00000019041HL3, 00000019048HL8, 00000019095HL9, 00000019098HL3, 00000019104HL9, 00000019110HL6, 00000019241HL9, 00000019244HL3, 00000019413HL4, 00000019419HL1, 00000019489HL4, 00000019496HL9, 00000019499HL3, 00000019501HL6, 00000019502HL4, 00000019669HL1, 00000019801HL0, 00000019886HL1, 00000019894HL5, 00000019959HL6, 00000020091HL5, 00000020092HL3, 00000020239HL0, 00000020243HL2, 00000020374HL5, 00000020967HL6, 00000021064HL1, 00000021197HL9, 00000021201HL9, 00000021204HL3, 00000021343HL9, 00000021344HL7, 00000021349HL6, 00000021552HL5, 00000021911HL3, 00000021921HL2, 00000021938HL6, 00000021953HL5, 00000022033HL5, 00000022333HL9, 00000022337HL0, 00000022412HL1, 00000022477HL4, 00000022600HL1, 00000022795HL9, 00000022931HL0, 00000023116HL7, 00000023122HL5, 00000023126HL6, 00000023412HL0, 00000023415HL3, 00000024386HL5, 00000024483HL0, 00000024588HL6, 00000024939HL1, 00000024940HL9, 00000025426HL8, 00000025475HL5, 00000025476HL3, 00000025477HL1, 00000025572HL9, 00000027104HL9, 00000027867HL1, 00000028057HL8, 00000028453HL9, 00000028524HL7, 00000028845HL6, 00000029098HL1, 00000029145HL0, 00000029502HL2, 00000029550HL1, 00000029634HL3, 00000029640HL0, 00000029969HL3, 00000030108HL5, 00000030113HL5, 00000030229HL9, 00000030232HL3, 00000030236HL4, 00000030633HL2, 00000030848HL6, 00000030920HL3, 00000030921HL1, 00000030976HL5, 00000031172HL0, 00000031175HL3, 00000031565HL5, 00000031702HL4, 00000031811HL3, 00000031928HL5, 00000031929HL3, 00000031930HL1, 00000032309HL7, 00000032311HL3, 00000032317HL0, 00000032319HL6, 00000032322HL0, 00000032850HL0, 00000032912HL8, 00000032916HL9, 00000032919HL3, 00000032920HL1, 00000033330HL2, 00000033388HL0, 00000033391HL4, 00000033408HL6, 00000033437HL5, 00000033509HL1, 00000033679HL2, 00000033681HL8, 00000033682HL6, 00000033727HL9, 00000033731HL1, 00000033884HL8, 00000033970HL5, 00000033971HL3, 00000033983HL8, 00000033984HL6, 00000034218HL8, 00000034420HL0, 00000034469HL7, 00000034470HL5, 00000034520HL7, 00000034644HL5, 00000034694HL0, 00000034762HL5, 00000034764HL1, 00000034913HL4, 00000034938HL1, 00000035052HL0, 00000035101HL5, 00000035159HL3, 00000035191HL6, 00000035244HL3, 00000035246HL8, 00000035247HL6, 00000035329HL2, 00000035445HL6, 00000035544HL6, 00000035547HL9, 00000035632HL9, 00000035641HL0, 00000036038HL8, 00000036068HL5, 00000036071HL9, 00000036100HL6, 00000036126HL1, 00000036128HL7, 00000036265HL7, 00000036362HL2, 00000036363HL0, 00000036431HL5, 00000036624HL5, 00000036625HL2, 00000036674HL0, 00000036698HL9, 00000036702HL9, 00000036882HL9, 00000036983HL5, 00000037104HL7, 00000037123HL7, 00000037351HL4, 00000037352HL2, 00000037353HL0, 00000037396HL9, 00000037596HL4, 00000037919HL8, 00000038423HL0, 00000038771HL2, 00000038772HL0, 00000039125HL0, 00000039581HL4, 00000039612HL7, 00000040366HL7, 00000040493HL9, 00000040728HL8, 00000040865HL8, 00000041290HL8, 00000041291HL6, 00000041511HL7, 00000041768HL3, 00000042367HL3, 00000042649HL4, 00000042650HL2, 00000042651HL0, 00000042701HL3, 00000042880HL5, 00000043029HL8, 00000043184HL1, 00000043265HL8, 00000043665HL9, 00000043727HL7.00000043747HL5, 00000043751HL7, 00000044548HL6, 00000045475HL1, 00000045651HL7, 00000015925HL1, 00000019969HL5, 00000024476HL4, 00000027957HL0, 00000030025HL1, 00000030028HL5, 00000032711HL4, 00000035548HL7, 00000036428HL1, 00000036883HL7, 00000020371HL1, 00000021582HL2, 00000031922HL8, 00000033316HL1, 00000042706HL2, 00000020735HL7, 00000022513HL6, 00000022926HL0, 00000024376HL6, 00000031909HL5, 00000040521HL7, 00000041017HL5, 00000011544HL4, 00000014336HL2, 00000016171HL1, 00000016545HL6, 00000017123HL1, 00000017483HL9, 00000023189HL4, 00000028833HL2, 00000030875HL9, 00000030924HL5, 00000031916HL0, 00000031918HL6, 00000032909HL4, 00000033730HL3, 00000033974HL7, 00000034268HL3, 00000036165HL9, 00000042001HL8, 00000016177HL8, 00000031554HL9, 00000036626HL0, 00000013383HL5, 00000014323HL0, 00000019485HL2, 00000020232HL5, 00000010836HL5, 00000019679HL0, 00000023546HL5, 00000023547HL3, 00000030926HL0, 00000041996HL0, 00000042388HL9, 00000042537HL1, 00000011063HL5, 00000015391HL6, 00000017138HL9, 00000021058HL3, 00000030918HL7, 00000040086HL1, 00000010328HL3, 00000013375HL1, 00000014335HL4, 00000015065HL6, 00000015081HL3, 00000015747HL9, 00000016371HL7, 00000016701HL5, 00000017125HL6, 00000017481HL3, 00000019482HL9, 00000019486HL0, 00000019678HL2, 00000019680HL8, 00000019788HL9, 00000020081HL6, 00000020224HL2, 00000021639HL0, 00000022032HL7, 00000023118HL3, 00000023181HL1, 00000024432HL7, 00000029259HL9, 00000031908HL7, 00000031914HL5, 00000032231HL3, 00000032897HL1, 00000033724HL6, 00000033725HL3, 00000033968HL9, 00000034260HL0, 00000034459HL8, 00000034462HL2, 00000034783HL1, 00000035377HL1, 00000035378HL9, 00000036563HL5, 00000037254HL0, 00000038190HL5, 00000016175HL2, 00000019042HL1, 00000013796HL8, 00000011182HL3, 00000032313HL9, 00000012475HL0, 00000020222HL6, 00000029743HL2, 00000033726HL1, 00000020732HL4, 00000020737HL3, 00000012674HL8, 00000021551HL7, 00000031689HL3, 00000010651HL8, 00000015757HL8, 00000022520HL1, 00000015082HL1, 00000015153HL0, 00000021644HL0, 00000013373HL6, 00000015158HL9, 00000022606HL8, 00000022607HL6, 00000024489HL2, 00000024590HL2, 00000024671HL0, 00000024689HL2, 00000030446HL9, 00000030460HL0, 00000030927HL8, 00000036429HL9, 00000015387HL4, 00000039893HL3, 00000039895HL8, 00000039962HL6, 00000040002HL8, 00000040072HL1, 00000040164HL6, 00000040167HL9, 00000022402HL2, 00000036266HL5, 00000036914HL0, 00000037252HL4, 00000013564HL0.
Other devices under the same FDA submission
1 device cleared under K993090 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Proteus XRaGE Hualun Medical Systems Co., Ltd · System, x-ray, stationary2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.