Skip to content

Stryker Peek Intraline Anchor, 6.5 mm with Needles, Model… recallZ-1536-2008 · Stryker Endoscopy · initiated Feb 7, 2008

Terminated All recalls posted in August 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Endoscopy or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product package may crack and cause loss of sterility.

FDA root cause category: Packaging

What the company is doing

Stryker Endoscopy initiated the recall on 02/07/2008 by notifying its consignees via telephone, as well as sending out Urgent: Device Removal notices. Recipients of affected lots are to inspect their inventory for part/lot numbers and quarantine any affected product. Affected products are to be returned to Stryker Endoscopy and Customer Service should be contacted for replacement alternatives. Also, the "Acknowledgment of recipt" card needs to be signed and returned to Stryker Endoscopy.

Recall details

Recall number
Z-1536-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Feb 7, 2008
Posted by the FDA
Aug 6, 2008
Recalling firm
Stryker Endoscopy
Quantity
125 units
Product code
HWC
FDA submissions
K071157
Distribution
Product was distributed to 8 consignees in CA, OH, PA, CO, MA, and NJ.

Product description

Stryker Peek Intraline Anchor, 6.5 mm with Needles, Model number 3910-400-025 Used for tissue fixation to bone in certain areas.

Code information (models, lots, serial numbers)

Lot numbers 07277AE2, 07292AE2

Other devices under the same FDA submission

4 devices cleared under K071157 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 46916).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page