PEEK IntraLine 3910-400-010Recalls & reported problems

Stryker Corporation · PEEK INTRALINE ANCHOR, 5.5MM WITH 2 STRANDS #2 FORCE FIBER · Class 2 · Implant

All models of PEEK IntraLine (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
21
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 07613154644568), received between Sep 1, 2020 and Oct 21, 2025 — classified as 21 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (13 reports).

5 other recalls cover devices cleared under the same FDA submission (K071157); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation13
  • Break11

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Insufficient Information3
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2025Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED THAT THE PROXIMAL END OF THE ANCHOR BROKE DURING INSERTION AT THE POINT WHERE THE ANCHOR INSERTER ENDS IN THE ANCHOR. A NEW ANCHOR WAS USED TO COMPLETE THE PROCEDURE. ALL BROKEN PARTS WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23353123 ↗
  • MalfunctionOct 21, 2025Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED THAT THE PROXIMAL END OF THE ANCHOR BROKE DURING INSERTION AT THE POINT WHERE THE ANCHOR INSERTER ENDS IN THE ANCHOR. A NEW ANCHOR WAS USED TO COMPLETE THE PROCEDURE. ALL BROKEN PARTS WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23353122 ↗
  • MalfunctionOct 21, 2025Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED THAT THE PROXIMAL END OF THE ANCHOR BROKE DURING INSERTION AT THE POINT WHERE THE ANCHOR INSERTER ENDS IN THE ANCHOR. A NEW ANCHOR WAS USED TO COMPLETE THE PROCEDURE. ALL BROKEN PARTS WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23351123 ↗
  • MalfunctionSep 10, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THERE WAS BREAKAGE OF PEEK PART OF PEEK INTRALINE ANCHOR POST DRILLING AND TAPPING. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23022295 ↗
  • MalfunctionSep 10, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THERE WAS BREAKAGE OF PEEK PART OF PEEK INTRALINE ANCHOR POST DRILLING AND TAPPING. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23022294 ↗
  • MalfunctionSep 10, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THERE WAS BREAKAGE OF PEEK PART OF PEEK INTRALINE ANCHOR POST DRILLING AND TAPPING. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23022286 ↗
  • MalfunctionJun 5, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT WHILE PLACING IN THE ANCHOR POST TAPPING WHILE FIXATING THE SCREW STARTED TO PULL OFF. THIS WAS A BREAKAGE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22154558 ↗
  • MalfunctionOct 2, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20355605 ↗
  • MalfunctionOct 2, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20355599 ↗
  • MalfunctionAug 15, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17552569 ↗
  • MalfunctionAug 15, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17552566 ↗
  • MalfunctionAug 8, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17484509 ↗
  • MalfunctionAug 7, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17476062 ↗
  • MalfunctionDec 21, 2022Break

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16024580 ↗
  • MalfunctionDec 15, 2022Break

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15988485 ↗
  • MalfunctionNov 10, 2021Break

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12789405 ↗
  • MalfunctionDec 17, 2020Break

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11032213 ↗
  • MalfunctionDec 16, 2020Break; Break

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11016258 ↗
  • MalfunctionDec 9, 2020Break

    IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10979957 ↗
  • MalfunctionSep 1, 2020Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10477804 ↗
  • MalfunctionSep 1, 2020Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10477683 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071157. One submission often covers many models, so check the product and lot details in each record.

  • Z-0618-2011 · Aug 20, 2009 · Terminated
    Stryker Titanium Intraline Anchor, Model Numbers 3910400050, 3910400055, 3910400060, 3910400065, Soft Eyelet Anchor, manufactured by Stryker Endoscopy, Puerto Rico
  • Z-1534-2008 · Feb 7, 2008 · Terminated
    Stryker Peek Intraline Anchor, 5.5 mm with Needles, Model number 3910-400-015 Used for tissue fixation to bone in certain areas.
  • Z-1533-2008 · Feb 7, 2008 · Terminated
    Stryker Peek Intraline Anchor, 5.5 mm, Model number 3910-400-010 Used for tissue fixation to bone in certain areas.
  • Z-1536-2008 · Feb 7, 2008 · Terminated
    Stryker Peek Intraline Anchor, 6.5 mm with Needles, Model number 3910-400-025 Used for tissue fixation to bone in certain areas.
  • Z-1535-2008 · Feb 7, 2008 · Terminated
    Stryker Peek Intraline Anchor, 6.5 mm, Model number 3910-400-020 Used for tissue fixation to bone in certain areas.

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →