Stryker Corporation · PEEK INTRALINE ANCHOR, 5.5MM WITH 2 STRANDS #2 FORCE FIBER · Class 2 · Implant
All models of PEEK IntraLine (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 21
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 07613154644568), received between Sep 1, 2020 and Oct 21, 2025 — classified as 21 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (13 reports).
5 other recalls cover devices cleared under the same FDA submission (K071157); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation13
- Break11
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- Insufficient Information3
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2025Break; Material Fragmentation; Break; Material Fragmentation
IT WAS REPORTED THAT THE PROXIMAL END OF THE ANCHOR BROKE DURING INSERTION AT THE POINT WHERE THE ANCHOR INSERTER ENDS IN THE ANCHOR. A NEW ANCHOR WAS USED TO COMPLETE THE PROCEDURE. ALL BROKEN PARTS WERE REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23353123 ↗MalfunctionOct 21, 2025Break; Material Fragmentation; Break; Material Fragmentation
IT WAS REPORTED THAT THE PROXIMAL END OF THE ANCHOR BROKE DURING INSERTION AT THE POINT WHERE THE ANCHOR INSERTER ENDS IN THE ANCHOR. A NEW ANCHOR WAS USED TO COMPLETE THE PROCEDURE. ALL BROKEN PARTS WERE REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23353122 ↗MalfunctionOct 21, 2025Break; Material Fragmentation; Break; Material Fragmentation
IT WAS REPORTED THAT THE PROXIMAL END OF THE ANCHOR BROKE DURING INSERTION AT THE POINT WHERE THE ANCHOR INSERTER ENDS IN THE ANCHOR. A NEW ANCHOR WAS USED TO COMPLETE THE PROCEDURE. ALL BROKEN PARTS WERE REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23351123 ↗MalfunctionSep 10, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THERE WAS BREAKAGE OF PEEK PART OF PEEK INTRALINE ANCHOR POST DRILLING AND TAPPING. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23022295 ↗MalfunctionSep 10, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THERE WAS BREAKAGE OF PEEK PART OF PEEK INTRALINE ANCHOR POST DRILLING AND TAPPING. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23022294 ↗MalfunctionSep 10, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THERE WAS BREAKAGE OF PEEK PART OF PEEK INTRALINE ANCHOR POST DRILLING AND TAPPING. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23022286 ↗MalfunctionJun 5, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT WHILE PLACING IN THE ANCHOR POST TAPPING WHILE FIXATING THE SCREW STARTED TO PULL OFF. THIS WAS A BREAKAGE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22154558 ↗MalfunctionOct 2, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20355605 ↗MalfunctionOct 2, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20355599 ↗MalfunctionAug 15, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17552569 ↗MalfunctionAug 15, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17552566 ↗MalfunctionAug 8, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17484509 ↗MalfunctionAug 7, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17476062 ↗MalfunctionDec 21, 2022Break
IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16024580 ↗MalfunctionDec 15, 2022Break
IT WAS REPORTED THAT THE ANCHOR BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 15988485 ↗MalfunctionNov 10, 2021Break
IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12789405 ↗MalfunctionDec 17, 2020Break
IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11032213 ↗MalfunctionDec 16, 2020Break; Break
IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11016258 ↗MalfunctionDec 9, 2020Break
IT WAS REPORTED THAT THE ANCHOR BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 10979957 ↗MalfunctionSep 1, 2020Break
IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10477804 ↗MalfunctionSep 1, 2020Break
IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10477683 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071157. One submission often covers many models, so check the product and lot details in each record.
- Z-0618-2011 · Aug 20, 2009 · TerminatedStryker Titanium Intraline Anchor, Model Numbers 3910400050, 3910400055, 3910400060, 3910400065, Soft Eyelet Anchor, manufactured by Stryker Endoscopy, Puerto Rico
- Z-1534-2008 · Feb 7, 2008 · TerminatedStryker Peek Intraline Anchor, 5.5 mm with Needles, Model number 3910-400-015 Used for tissue fixation to bone in certain areas.
- Z-1533-2008 · Feb 7, 2008 · TerminatedStryker Peek Intraline Anchor, 5.5 mm, Model number 3910-400-010 Used for tissue fixation to bone in certain areas.
- Z-1536-2008 · Feb 7, 2008 · TerminatedStryker Peek Intraline Anchor, 6.5 mm with Needles, Model number 3910-400-025 Used for tissue fixation to bone in certain areas.
- Z-1535-2008 · Feb 7, 2008 · TerminatedStryker Peek Intraline Anchor, 6.5 mm, Model number 3910-400-020 Used for tissue fixation to bone in certain areas.
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026