CORDIS S.M.A.R.T. Control" Nitinol Stent System recallZ-1556-2008 · Cordis Corporation · initiated Mar 10, 2008
Terminated All recalls posted in July 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cordis Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Data transcription error. The 10 affected lots of Catalog Number C10040SL were manufactured with a prior version of the guidewire lumen material, which is not validated with the current manufacturing process. This condition may result in a separation of the device. If a separation was to occur, and the separated fragment of lumen material was not retrieved on the guidewire, blockage of the vessel
FDA root cause category: Mixed-up of materials/components
What the company is doing
A letter (RE: Medical Device Recall) dated March 11, 2008, and Acknowledgement Form was sent overnight to the following 6 individuals in each account: Dir. Cardiac Cath Lab, Dir. Radiology Services, Dir. Vascular Services, Risk Manager, Materials Dir., Hospital Administrator; requesting that the affected product be set aside, and that the form be reviewed, signed, and faxed back to Cordis. Affected product is to be either returned or the local sales rep should the contacted to facilitate the return of the affected product. All product will be replaced. Customer Service can be contacted if there are additional questions relative to product return, replacement, or availability.
Recall details
- Recall number
- Z-1556-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 10, 2008
- Posted by the FDA
- Jul 31, 2008
- Recalling firm
- Cordis Corporation
- Quantity
- 398 units
- Product code
- NIO
- FDA submissions
- P020036
- Distribution
- Worldwide Distribution --- including USA and countries of Japan and Canada.
Product description
CORDIS S.M.A.R.T. Control" Nitinol Stent System Model Number(s): C10040SL, Manufactured in Mexico. Indicated for improving luminal diameter in patients with symptomatic atherosclerotic disease.
Code information (models, lots, serial numbers)
10 lots. Lot #s: 13326414*, 13326415*, 13327582, 13329185*, 13332369, 13333077*, 13339105*, 13339106*, 13341244, 13343464. (*These six lots have U.S. distribution.)
Other devices under the same FDA submission
38 devices cleared under P020036 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C08040SLCordis US Corp. · STENT, SUPERFICIAL FEMORAL ARTERY7 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C08060MLCordis US Corp. · STENT, SUPERFICIAL FEMORAL ARTERY7 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C08100SLCordis US Corp. · STENT, SUPERFICIAL FEMORAL ARTERY2 deaths7 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C07040MLCordis US Corp. · STENT, SUPERFICIAL FEMORAL ARTERY6 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09040SLCordis US Corp. · STENT, SUPERFICIAL FEMORAL ARTERY1 death6 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C06040MLCordis US Corp. · STENT, SUPERFICIAL FEMORAL ARTERY1 death5 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C06060MLCordis Corporation · STENT, ILIAC5 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C06100MLCordis US Corp. · STENT, ILIAC5 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C06100SLCordis US Corp. · STENT, SUPERFICIAL FEMORAL ARTERY5 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C08060SLCordis US Corp. · STENT, SUPERFICIAL FEMORAL ARTERY1 death5 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page