Medical device recalls: July 2008
The FDA posted 141 medical device recall records in July 2008, from 32 recall events by 29 companies. 141 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Stryker Howmedica Osteonics Corp. · 2 events, 18 records
- AGFA Corp. · 2 events, 2 records
- Arjo, Inc. · 2 events, 2 records
- Edwards Lifesciences, Llc · 1 event, 51 records
- Beckman Coulter Inc · 1 event, 18 records
- Medtronic Perfusion Systems · 1 event, 12 records
- Zimmer Inc. · 1 event, 6 records
- Parks Medical Electronics Inc · 1 event, 4 records
FDA root cause
- Packaging process control · 61
- Software design · 23
- Process control · 21
- Nonconforming Material/Component · 20
- Device Design · 7
- Labeling mix-ups · 3
- Component design/selection · 3
- Release of Material/Component prior to receiving test results · 1
Not yet classified (141)
The FDA has posted these records but has not published a class for them yet.
- Z-1556-2008 · Cordis Corporation · posted Jul 31, 2008 · Terminated
Data transcription error. The 10 affected lots of Catalog Number C10040SL were manufactured with a prior version of the guidewire lumen material, which is not validated with the current manufacturing process. This condition may result in a separation of…
- Z-1569-2008 · Abbott Laboratories · posted Jul 31, 2008 · Terminated
Testing material becomes unstable at the recommeded storage temperature. Some of the Architect Free T4 MasterCheck and Total T4 MasterCheck lots are not stable at the recommended storage condition of 2 to 8 degrees Celsius through expiration dating.…
- Z-1570-2008 · Abbott Laboratories · posted Jul 31, 2008 · Terminated
Testing material becomes unstable at the recommeded storage temperature. Some of the Architect Free T4 MasterCheck and Total T4 MasterCheck lots are not stable at the recommended storage condition of 2 to 8 degrees Celsius through expiration dating.…
- Z-1460-2008 · Varian Medical Systems Oncology Systems · posted Jul 30, 2008 · Terminated
Incorrect Image Orientation: The image orientation tag may not be set correctly if the patient has been scanned with an orientation other than Head First Spine. When such images are imported to a treatment planning system, they may appear with either a…
- Z-1549-2008 · Smiths Medical Asd Inc · posted Jul 30, 2008 · Terminated
Misbranded/mispackaged device was distributed. Product labeled to contain 72 inch Standard Drip ( IV) Administration Sets, actually contained 60 inch (IV) administration sets with a' Y' site.
- Z-1550-2008 · Zimmer Inc. · posted Jul 30, 2008 · Terminated
Sterility compromised. Lack of assurance of sterility, as the packages may not have been sealed on the manufacturer's end.
- Z-1551-2008 · Zimmer Inc. · posted Jul 30, 2008 · Terminated
Sterility compromised. Lack of assurance of sterility, as the packages may not have been sealed on the manufacturer's end.
- Z-1552-2008 · Zimmer Inc. · posted Jul 30, 2008 · Terminated
Sterility compromised. Lack of assurance of sterility, as the packages may not have been sealed on the manufacturer's end.
- Z-1553-2008 · Zimmer Inc. · posted Jul 30, 2008 · Terminated
Sterility compromised. Lack of assurance of sterility, as the packages may not have been sealed on the manufacturer's end.
- Z-1554-2008 · Zimmer Inc. · posted Jul 30, 2008 · Terminated
Sterility compromised. Lack of assurance of sterility, as the packages may not have been sealed on the manufacturer's end.
- Z-1555-2008 · Zimmer Inc. · posted Jul 30, 2008 · Terminated
Sterility compromised. Lack of assurance of sterility, as the packages may not have been sealed on the manufacturer's end.
- Z-1557-2008 · Arjo, Inc. · posted Jul 30, 2008 · Terminated
Scale could break and fall potentially causing injuries to someone who is on it. There is a potential for the mast scale load cell ssembly to fracture and allow the mast/boom to fall. Should a patient be in the lift at the time, the patient could fall…
- Z-2005-2008 · Medtronic International Trade Inc · posted Jul 30, 2008 · Terminated
Medtronic is advising that certain Trillium-coated Medtronic cardiopulmonary bypass (CBP) products (e.g., oxygenators, reservoirs, pumps and tubing packs) were manufactured with heparin batches contaminated with oversulfated chondroitin sulfate (OSCS).…
- Z-2037-2008 · GE Healthcare Integrated IT Solutions · posted Jul 30, 2008 · Terminated
Patient information mix-up: When the PFILS application, configured with a backup server has experienced a network interruption, PFILS verify could result in the Centricity Perinatal application recording non-identifiable patient information to the…
- Z-1373-2008 · Biotronik Inc · posted Jul 29, 2008 · Terminated
Timing Feature Error: An implementation error allows the usage of the V-V timing feature, which is not approved for use in the U.S.
- Z-1572-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1573-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1574-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1575-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1576-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1577-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1578-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1579-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1580-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1581-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1582-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1583-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1584-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1585-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1586-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1587-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 26, 2008 · Terminated
Sharp edges will cause a glove to tear and cut someone. Sharp edges on the slots machines in select lots of the Triathlon TS Femoral Trials, sizes 1-8, have been discovered.
- Z-1375-2008 · Biomet, Inc. · posted Jul 25, 2008 · Terminated
Mislabeled: The packaging of the two products was mixed up, resulting in the products being labeled with an incorrect size. If implanted, the product may not fit properly, which could require repeat surgery. The problem could also result in a delay in the…
- Z-1376-2008 · Biomet, Inc. · posted Jul 25, 2008 · Terminated
Mislabeled: The packaging of the two products was mixed up, resulting in the products being labeled with an incorrect size. If implanted, the product may not fit properly, which could require repeat surgery. The problem could also result in a delay in the…
- Z-1566-2008 · Abbott Diagnostic International, Ltd. · posted Jul 24, 2008 · Terminated
High Control value out of range after calibration. --AN INCREASED FREQUENCY OF HIGH CONTROL VALUES BEYOND THE UPPER LIMIT OF THE RANGE SPECIFIED BY THE PACKAGE INSERT.
- Z-1571-2008 · Hitachi Medical Systems America Inc · posted Jul 24, 2008 · Terminated
Miscalculation reading. A software error in the firm's EUB-5500/HI VISION diagnostic scanning system causes a miscalculation of the left ICA/CCA ratio when using the Carotid Calculation package for patient scans.
- Z-1919-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1920-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1921-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1922-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1923-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1924-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1925-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1926-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1927-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1928-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1929-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1930-2008 · Medtronic Perfusion Systems · posted Jul 24, 2008 · Terminated
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it…
- Z-1372-2008 · OEC Medical Systems, Inc · posted Jul 23, 2008 · Terminated
No Premarket Notification: Product does not have FDA 510(k) approval for use.
- Z-1473-2008 · Medline Industries Inc · posted Jul 23, 2008 · Terminated
Overheating of Components: There is the potential for an electrical component in the oxygen concentrator to overheat, causing damage to the unit and making it nonfunctional.
- Z-1474-2008 · Medline Industries Inc · posted Jul 23, 2008 · Terminated
Overheating of Components: There is the potential for an electrical component in the oxygen concentrator to overheat, causing damage to the unit and making it nonfunctional.
- Z-1479-2008 · Abbott Spine · posted Jul 23, 2008 · Terminated
Screw Head Dissociation: The polyaxial screw head may dissociate from screw shaft during surgery. Also, the polyaxial screw driver is experiencing difficulty (broken/bent driver shafts, difficulty attaching screws to the driver and jamming of the collet.)
- Z-1901-2008 · Levitronix, Llc · posted Jul 23, 2008 · Terminated
Stops pumping: Interruption of CentriMag System support may occur when using a Valleylab Force FX-C Electrocautery Unit
- Z-1902-2008 · Levitronix, Llc · posted Jul 23, 2008 · Terminated
Stops pumping: Interruption of CentriMag System support may occur when using a Valleylab Force FX-C Electrocautery Unit
- Z-1308-2008 · Medtronic Neuromodulation · posted Jul 22, 2008 · Terminated
Incorrect Connector Pin: The Medtronic 8596SC Intrathecal Catheter Pump Segment Revision kits were packaged with the incorrect connector pin.
- Z-1233-2008 · Arjo, Inc. · posted Jul 18, 2008 · Terminated
The ceiling lift may not stop lateral movement after releasing the left (or right) action button on a ceiling lift equipped with an infrared hand control.
- Z-1917-2008 · Celsus Laboratories Inc · posted Jul 18, 2008 · Terminated
Contamination: The firm learned that up to 2.2% oversulfated chondroitin was found in a batch of crude heparin they received from their supplier.
- Z-1235-2008 · American Medical Systems · posted Jul 16, 2008 · Terminated
Missing part: The units in two lots of In-Fast Ultra Transvaginal Bladder Neck Support System may contain only one bone screw, but should contain two as indicated on the labeling. Two screws are required to complete the procedure. The missing screw…
- Z-1315-2008 · Draeger Medical, Inc. · posted Jul 16, 2008 · Terminated
Detachment of Component: The chassis on some machines were manufactured out of specifications. As a result, the caster may break loose from the chassis.
- Z-1379-2008 · Smiths Medical ASD, Inc. · posted Jul 16, 2008 · Terminated
Incorrect Oxygen Level: Device may not provide expected oxygen concentration or the expected flow rate due to incorrectly molded part. Patient will not receive proper amounts of air/oxygen when in need.
- Z-1380-2008 · Smiths Medical ASD, Inc. · posted Jul 16, 2008 · Terminated
Incorrect Oxygen Level: Device may not provide expected oxygen concentration or the expected flow rate due to incorrectly molded part. Patient will not receive proper amounts of air/oxygen when in need.
- Z-1381-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 16, 2008 · Terminated
Mislabeled: The labeling indicates a Scorpio CR Femoral Component, however, the device inside the package is actually a Scorpio PS Femoral Component.
- Z-1382-2008 · Stryker Howmedica Osteonics Corp. · posted Jul 16, 2008 · Terminated
Mislabeled: The labeling indicates a Scorpio PS Femoral Component, however, the device inside the package is actually a Scorpio CR Femoral Component.
- Z-2016-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2017-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2018-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2019-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2020-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2021-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2022-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2023-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2024-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2025-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2026-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2027-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2028-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2029-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2030-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2031-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2032-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-2033-2008 · Beckman Coulter Inc · posted Jul 11, 2008 · Terminated
Incorrect Display: In certain modes, the protocol will not initially display the correct data in some plots. In addition, when running EWL files, any edited SampleID1 is not automatically updated in all tubes in the panel.
- Z-1779-2008 · Philips Medical Systems · posted Jul 9, 2008 · Terminated
Buttons Sticking - The On/Off and Shock buttons on the front panel of the device may stick in place when pressed and fail to respond to the button press, causing a delay in the delivery of a defibrillation shock when the system advises a shock.
- Z-1780-2008 · Philips Medical Systems · posted Jul 9, 2008 · Terminated
Buttons Sticking - The On/Off and Shock buttons on the front panel of the device may stick in place when pressed and fail to respond to the button press, causing a delay in the delivery of a defibrillation shock when the system advises a shock.
- Z-1781-2008 · Philips Medical Systems · posted Jul 9, 2008 · Terminated
Buttons Sticking - The On/Off and Shock buttons on the front panel of the device may stick in place when pressed and fail to respond to the button press, causing a delay in the delivery of a defibrillation shock when the system advises a shock.
- Z-0670-2008 · Parks Medical Electronics Inc · posted Jul 3, 2008 · Terminated
Premature battery failure: Wrong acid flux used on circuit board which may lead to premature battery failure of ultrasonic non-directional doppler flowmeter.
- Z-0671-2008 · Parks Medical Electronics Inc · posted Jul 3, 2008 · Terminated
Premature battery failure: Wrong acid flux used on circuit board which may lead to premature battery failure of ultrasonic non-directional doppler flowmeter.
- Z-0672-2008 · Parks Medical Electronics Inc · posted Jul 3, 2008 · Terminated
Premature battery failure: Wrong acid flux used on circuit board which may lead to premature battery failure of ultrasonic non-directional doppler flowmeter.
- Z-0673-2008 · Parks Medical Electronics Inc · posted Jul 3, 2008 · Terminated
Premature battery failure: Wrong acid flux used on circuit board which may lead to premature battery failure of ultrasonic non-directional doppler flowmeter.
- Z-1246-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1247-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1248-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1249-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1250-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1251-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1252-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1253-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1254-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1255-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1256-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1257-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1258-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1259-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1260-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1261-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1262-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1263-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1264-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1265-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1266-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1267-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1268-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1269-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1270-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1271-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1272-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1273-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1274-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1275-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1276-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1277-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1278-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1279-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1280-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1281-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1282-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1283-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1284-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1285-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1286-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1287-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1288-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1289-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1290-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1291-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1292-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1293-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1294-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1295-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1296-2008 · Edwards Lifesciences, Llc · posted Jul 2, 2008 · Terminated
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
- Z-1307-2008 · AGFA Corp. · posted Jul 2, 2008 · Terminated
Misidentification: The orthopedic planning X-ray images for one patient are misidentified as the images for another patient.
- Z-1314-2008 · ConvaTec · posted Jul 2, 2008 · Terminated
Mislabeled: The carton label reads "For postsurgical use following a urostomy procedure." It should read "For postsurgical use following a colostomy or ileostomy procedure."
- Z-1306-2008 · AGFA Corp. · posted Jul 1, 2008 · Terminated
Failure of Data Transmission: No imaging data could be transmitted by the device, because a key value needed for data transmission was incorrectly entered in the affected units by AGFA Healthcare.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.