Roche/Hitachi Modular Analytics System, Modular E Module… recallZ-0868-2008 · Roche Diagnostics Corp. · initiated Mar 21, 2008

Terminated All recalls posted in August 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Possible mismatch between patient and result.

FDA root cause category: Process change control

What the company is doing

Consignees were notified via Urgent Medical Device Correction letter dated 3/21/08, which instructed them to not manually order QC test selections while the system is in operation mode and that they can only manually order QC when the system is in Stand-by mode. Consignees are instructed to consult with their physicians or pathologist to determine specific clinical implications for their patients. The firm states that this issue will be resolved in the next version of software.

Recall details

Recall number
Z-0868-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 21, 2008
Posted by the FDA
Aug 13, 2008
Recalling firm
Roche Diagnostics Corp.
Quantity
11 of all models.
Product code
JJE
FDA submissions
K961481
Distribution
Nationwide Distribution --- including states of Florida, Georgia, Indiana, New Jersey and North Carolina.

Product description

Roche/Hitachi Modular Analytics System, Modular E Module GMMI immunoassay analyzer; Catalog numbers 04998642001 and 03617505001. Intended for qualitative and quantitative in vitro determinations using a wide variety of tests. Roche Diagnostics Corporation, 9115 Hague Road, Indianapolis, IN, 46250

Code information (models, lots, serial numbers)

All units with software version 08-02.

Other devices under the same FDA submission

2 devices cleared under K961481 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.