Dimension Total Prostate Specific Antigen (TPSA) Flex… recallZ-1606-2008 · Siemens Healthcare Diagnostics Inc. · initiated Mar 24, 2008

Terminated All recalls posted in June 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and quality control; therefore this issue may not be readily detected upon calibration or during routine processing of Quality Control.

FDA root cause category: Device Design

What the company is doing

The recalling firm issued a letter (Urgent Field Safety Notice) dated 3/25/08 to inform their customers to discontinue use of any remaining inventory, retest using an alternate lot for those samples from post-prostatectomy patients which recovered positive TPSA results less than 1.0 ng/nL. The firm has requested that you call the them for no-charge replacement product at 1-800-441-9250 and/or for technical assistance.

Recall details

Recall number
Z-1606-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 24, 2008
Posted by the FDA
Jun 28, 2008
Recalling firm
Siemens Healthcare Diagnostics Inc.
Quantity
1631 cartons
Product code
MTF
FDA submissions
P000021
Distribution
The product was shipped to medical facilities nationwide. The product was also shipped to Germany, Singapore, Canada, Japan and AE

Product description

Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19714

Code information (models, lots, serial numbers)

Lot number GB9021 exp 2009-01-21 and Lot number GB9029 exp 2009-01-29