Skip to content

The OptiVantage DH Injection System is a contrast delivery… recallZ-1694-2008 · Mallinckrodt Inc · initiated Apr 7, 2008

Terminated All recalls posted in August 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mallinckrodt Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Screws can begin to loosen and shear off to the point of the J-Bow falling from the suspension system. This may cause injury to patients or hospital staff.

FDA root cause category: Component design/selection

What the company is doing

On April 18, 2008, Mallincrodt Inc. issued an Urgent Device Correction letter informing customers to check their inventory, and determine if they have any defective system units. If they observe any loose screws, sheared screws and/or elongated opening(s) around the screws, they are to immediately take the system out of service and contact the firm's service representative. Covidien's customers were all asked to inspect their inventories and to complete an 'Effectiveness Check Return Form' whether or not the product is defective. However, only those system units observed to have loose screws, sheared screws, and/or elongated openings around the screws, will be scheduled to receive correction by a Covidien Service Representative.

Recall details

Recall number
Z-1694-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 7, 2008
Posted by the FDA
Aug 15, 2008
Recalling firm
Mallinckrodt Inc
Quantity
783 Units
Product code
IZQ
FDA submissions
K063503
Distribution
Product is distributed in the US throughout the following states: AL, AR, AZ, CA, CO ,CT, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, and WV.

Product description

The OptiVantage DH Injection System is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images when used with computed tomography equipment. The J-Bow component is used to connect the powerhead to a suspension system.

Code information (models, lots, serial numbers)

Catalog #V8400, Part No. 241078 & Catalog #V8402, Part #241079

Other devices under the same FDA submission

2 devices cleared under K063503 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page