Skip to content

Medtronic Venous Cannula with bent tip wire, for vascular… recallZ-1923-2008 · Medtronic Perfusion Systems · initiated May 9, 2008

Terminated All recalls posted in July 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it had provided some batches to Medtronic that were contaminated with oversulfated chondroitin sulfate (OSCS).

FDA root cause category: Nonconforming Material/Component

What the company is doing

Medtronic issued a Press Release on 05/07/2008. An Urgent Medical Device Recall letter was sent to customers on May 07, 2008. The letter identifies affected devices, explains the situation and requests that the impacted product be returned to Medtronic for disposal. The end user will be requested to complete a Recall Certificate acknowledging the receipt of this information.

Recall details

Recall number
Z-1923-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 9, 2008
Posted by the FDA
Jul 24, 2008
Recalling firm
Medtronic Perfusion Systems
Quantity
12 US
Product code
DWF
FDA submissions
K842375
Distribution
Worldwide Distribution: USA, Australia, Belgium, Canada, Czech Republic, France, Italy, Japan, Norway, and Taiwan.

Product description

Medtronic Venous Cannula with bent tip wire, for vascular cardiopulmonary bypass. Catalog Number: CB67532

Code information (models, lots, serial numbers)

Catalog / Lot: CB67532 / 9816217

Other devices under the same FDA submission

2 devices cleared under K842375 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 48100).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page