Lactosorb Trauma Plating System, part numbers: 915-2100… recallZ-1872-2008 · Biomet Microfixation, Inc. · initiated Mar 19, 2008
Terminated All recalls posted in September 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet Microfixation, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The outer foil pouch on the Lactosorb Plate system was not properly sealed. The product is labeled as sterile, but the lot was not sterilized. These devices are intended to be implanted into bone and are required to be sterile. A heavy bioburden would increase the potential for bone infection (osteomyelitis).
FDA root cause category: Packaging process control
What the company is doing
On 3/19/08 the firm issued an Urgent Medical Device Recall Notice to its consignees by certified letters with return receipt. All returned products will be held in quarantine until final disposition is determined. If you have any questions contact Rachel Osbeck at 1-800-874-7711 or (904) 741-4400 ext 448.
Recall details
- Recall number
- Z-1872-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 19, 2008
- Posted by the FDA
- Sep 16, 2008
- Recalling firm
- Biomet Microfixation, Inc.
- Quantity
- 117
- Product code
- JEY
- FDA submissions
- K955729, K980927
- Distribution
- Class II Recall - Worldwide Distribution --- USA including states of NC, MI, TX, GA, HI, OK, IN and CA, and countries of Denmark, Germany, Mexico, Italy, Brazil, Japan, Canada, South Africa, China and the Philippines.
Product description
Lactosorb Trauma Plating System, part numbers: 915-2100, 915-2423, 915-2113, 915-2810, 915-2418, 915-2105, 915-2832 and CP-1323. The system is used in trauma and reconstructive surgical procedures in the mid face and craniofacial skeleton. Used to support non-load bearing tissues of cranial maxillofacial anatomy, including but not limited to brow fixation.
Code information (models, lots, serial numbers)
Part numbers: 915-2100, 915-2423, 915-2113, 915-2810, 915-2418, 915-2105, 915-2832 and CP-1323. Lot numbers: 898240, 898330, 898250, 898350, 898320, 951140, 868920, and 978810.
Other devices under the same FDA submission
21 devices cleared under K955729, K980927 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- 2.0 LACTOSORB SYSTEM 915-2301Biomet Microfixation, Inc · PLATE, FIXATION, BONE32 reports
- 1.5 LACTOSORB SYSTEM 915-2316Biomet Microfixation, Inc · PLATE, FIXATION, BONE9 reports
- 2.0 LACTOSORB SYSTEM 915-2111Biomet Microfixation, Inc · PLATE, FIXATION, BONE5 reports
- 2.0 LACTOSORB SYSTEM 915-2101Biomet Microfixation, Inc · Screw, fixation, bone4 reports
- 2.0 LACTOSORB SYSTEM 915-2102Biomet Microfixation, Inc · Screw, fixation, bone4 reports
- 1.5 LACTOSORB SYSTEM 915-2315Biomet Microfixation, Inc · POLYMER, EAR, NOSE AND THROAT, SYNTHETIC, ABSORBABLE3 reports
- 2.0 LACTOSORB SYSTEM 915-2103Biomet Microfixation, Inc · Screw, fixation, bone2 reports
- 2.0 LACTOSORB SYSTEM 915-2110Biomet Microfixation, Inc · PLATE, FIXATION, BONE2 reports
- 1.5 LACTOSORB SYSTEM 915-2314Biomet Microfixation, Inc · PLATE, FIXATION, BONE2 reports
- 1.5 LACTOSORB SYSTEM 915-2413Biomet Microfixation, Inc · Screw, fixation, bone2 reports
- 1.5 LACTOSORB SYSTEM 915-2414Biomet Microfixation, Inc · Screw, fixation, bone2 reports
- 1.5 LACTOSORB SYSTEM 915-2419Biomet Microfixation, Inc · Screw, fixation, bone2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page