2.0 LACTOSORB SYSTEM 915-2111Recalls & reported problems
Biomet Microfixation, Inc · Class 2 · Implant
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00841036054564), received between Mar 19, 2020 and Mar 2, 2026 — classified as 5 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K955729, K971870); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Break2
- Material Deformation2
- Fracture1
Patient problems
- Unspecified Infection3
- Inflammation2
- Loss of fracture reduction2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMar 2, 2026Break; Fracture; Material Deformation
IT WAS REPORTED THAT APPROXIMATELY 12 DAYS POST-IMPLANTATION, THE PATIENT EXPERIENCED BONE FLAP COLLAPSE NECESSITATING A SECOND CRANIOTOMY. DURING THE REOPERATION, INTRAOPERATIVE FINDINGS INCLUDED SEVERE DEFORMATION OF THE LEFT UPPER CONNECTOR PLATE AND COMPLETE FRACTURE OF THE LEFT LOWER CONNECTOR PLATE. THE SCREWS REMAINED INTACT, AND ONLY THE FOUR-HOLE CONNECTOR PLATE WAS DAMAGED. THE PATIENT HAD NO HISTORY OF TRAUMA, AND THE PROSTHESIS WAS NOT RESHAPED USING A WATER BATH DURING THE INITIAL SURGERY. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Loss of fracture reduction
MAUDE report 24484441 ↗InjuryMar 2, 2026Break; Material Deformation
IT WAS REPORTED THAT APPROXIMATELY 12 DAYS POST-IMPLANTATION, THE PATIENT EXPERIENCED BONE FLAP COLLAPSE NECESSITATING A SECOND CRANIOTOMY. DURING THE REOPERATION, INTRAOPERATIVE FINDINGS INCLUDED SEVERE DEFORMATION OF THE LEFT UPPER CONNECTOR PLATE AND COMPLETE FRACTURE OF THE LEFT LOWER CONNECTOR PLATE. THE SCREWS REMAINED INTACT, AND ONLY THE FOUR-HOLE CONNECTOR PLATE WAS DAMAGED. THE PATIENT HAD NO HISTORY OF TRAUMA, AND THE PROSTHESIS WAS NOT RESHAPED USING A WATER BATH DURING THE INITIAL SURGERY. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Loss of fracture reduction
MAUDE report 24484419 ↗InjuryDec 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT INFECTION AND INFLAMMATION WERE DISCOVERED NINE (9) DAYS FOLLOWING IMPLANTATION OF PLATES AND SCREWS AND A REMOVAL WAS PLANNED TWO (2) WEEKS FOLLOWING IMPLANTATION. THE SURGEON IS UNCERTAIN AS TO THE CAUSE OF THE INFECTION AND THINKS IT IS POSSIBLE THAT THE PLATES AND SCREWS ARE THE CAUSE OF THE INFECTION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Inflammation; Unspecified Infection
MAUDE report 11086856 ↗InjuryDec 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT INFECTION AND INFLAMMATION WERE DISCOVERED NINE (9) DAYS FOLLOWING IMPLANTATION OF PLATES AND SCREWS AND A REMOVAL WAS PLANNED TWO (2) WEEKS FOLLOWING IMPLANTATION. THE SURGEON IS UNCERTAIN AS TO THE CAUSE OF THE INFECTION AND THINKS IT IS POSSIBLE THAT THE PLATES AND SCREWS ARE THE CAUSE OF THE INFECTION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Inflammation; Unspecified Infection
MAUDE report 11086853 ↗InjuryMar 19, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THREE (3) BIO-ABSORBABLE PLATES WERE REMOVED DUE TO INFECTION APPROXIMATELY TWO (2) WEEKS FOLLOWING IMPLANTATION DURING A MIDFACE RECONSTRUCTION PROCEDURE. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 9853284 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K955729, K971870. One submission often covers many models, so check the product and lot details in each record.
- Z-1872-2008 · Mar 19, 2008 · TerminatedLactosorb Trauma Plating System, part numbers: 915-2100, 915-2423, 915-2113, 915-2810, 915-2418, 915-2105, 915-2832 and CP-1323. The system is used in trauma and reconstructive su
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-DRF32Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.