Model 8596SC Intrathecal Catheter Pump Segment Revision Kit recallZ-2172-2008 · Medtronic Neuromodulation · initiated Apr 14, 2008
Terminated All recalls posted in August 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period.
FDA root cause category: Device Design
What the company is doing
Medtronic informed all Belgium implanting and managing physicians of the SynchroMed II infusion system. Each customer was mailed a Dear Healthcare Professional letter (Medtronic Urgent: Field Safety Notice) on 4/3/08 providing important safety information and patient management recommendations regarding the device incompatibility. The firm recommends that healthcare professionals do not use SC catheters if a connection with an implantable access port is required. Also, if a continuous infusion trial is required with use of an implantable access port, they should use a model 8711 intrathecal catheter. The firm states that for patients who are currently undergoing a continuous trial with an implantable port, replacing the catheter connector before connecting to the SynchroMedd II using the Medtronic model 8578 revision kit is recommended. Medtronic also recommends discussing the risks of connection failure with patients. Local field reps will personally follow-up with affected customers to confirm and document notification.
Recall details
- Recall number
- Z-2172-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 14, 2008
- Posted by the FDA
- Aug 27, 2008
- Recalling firm
- Medtronic Neuromodulation
- Quantity
- 232 total from all models
- Product code
- LKK
- FDA submissions
- P860004
- Distribution
- International Distribution --- country of Belgium.
Product description
Model 8596SC Intrathecal Catheter Pump Segment Revision Kit Medtronic Inc., Minneapolis, MN 55432 A part of the SynchroMed II Infusion system, which is designed to restore and provide intrathecal administration of drug therapy.
Code information (models, lots, serial numbers)
All lots numbers included
Other devices under the same FDA submission
51 devices cleared under P860004 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,929 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,303 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,119 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,054 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,084 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,588 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,291 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths2,242 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall4 deaths1,964 reports
- SynchroMed® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths1,920 reports
- SYNCHROMED™ III 8667-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 death1,937 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 deaths1,766 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 48616).
- Z-2171-2008 · TerminatedMedtronic 8578 Sutureless Pump Connector Revision Kit Medtronic Inc., Minneapolis, MN 55432 A part of the SynchroMed II Infusion system, which is designed to restore and provide i…
- Z-2173-2008 · TerminatedModel 8709SC Indura 1P One-Piece Intrathecal Catheter Medtronic Inc., Minneapolis, MN 55432 A part of the SynchroMed II Infusion system, which is designed to restore and provide …
- Z-2174-2008 · TerminatedModel 8731SC Intrathecal Catheter Medtronic Inc., Minneapolis, MN 55432 A part of the SynchroMed II Infusion system, which is designed to restore and provide intrathecal administr…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page