Skip to content

Piccolo Chemistry Analyzer software version 6.121 with… recallZ-2287-2008 · Abaxis Inc · initiated Apr 21, 2008

Terminated All recalls posted in September 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abaxis Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples

FDA root cause category: Process control

What the company is doing

The firm issued a notification letter to its consignees on April 21, 2008. The firm intends to follow up with phone calls and provide a software upgrade for the products. Contact Abaxis Customer Service at 1-510-675-6500 for assistance.

Recall details

Recall number
Z-2287-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 21, 2008
Posted by the FDA
Sep 17, 2008
Recalling firm
Abaxis Inc
Quantity
237 kits
Product code
JLB
FDA submissions
K992140
Distribution
Product was distributed Worldwide.

Product description

Piccolo Chemistry Analyzer software version 6.121 with Electrolyte Panel Lot #8041AC5, model number 400-1022 in polyfoil pouch. Piccolo Electrolyte Panel is a single-use, self-contained reagent disk intended for use with the Piccolo and PiccoloXpress chemistry analyzers, for in vitro qualitative determination.

Code information (models, lots, serial numbers)

software version 6.121 with Electrolyte Panel Lot #8041AC5, expiration date 2009-01-21

Other devices under the same FDA submission

5 devices cleared under K992140 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 48709).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page