Skip to content

Roche/Hitachi Modular E Module immunoassay analyzer, Roche… recallZ-0165-2009 · Roche Diagnostics Corp. · initiated Jun 24, 2008

Terminated All recalls posted in October 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software bug may result in pipetting from an incorrect reagent pack and/or assigning calibration curve parameters incorrectly.

FDA root cause category: Other

What the company is doing

The consignees were notified of the problem via letter dated 6/24/08 and were provided instructions for a workaround.

Recall details

Recall number
Z-0165-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jun 24, 2008
Posted by the FDA
Oct 22, 2008
Recalling firm
Roche Diagnostics Corp.
Product code
JJE
FDA submissions
K961481
Distribution
Nationwide Distribution --- including states of Florida, New Jersey and North Carolina.

Product description

Roche/Hitachi Modular E Module immunoassay analyzer, Roche Diagnostics, Ind., IN; GMMI Nos. 04998642001 and 03617505001.

Code information (models, lots, serial numbers)

All units with software version 08-02.

Other devices under the same FDA submission

2 devices cleared under K961481 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page