Roche COBAS c 111 Analyzer with ISE module option, GMMI No.… recallZ-1009-2009 · Roche Diagnostics Corp. · initiated Nov 17, 2008

Terminated All recalls posted in March 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a manufacturing error at the supplier, not all cobas c 111 pinch valve caps are properly glued to the correct position on the pinch valve shaft. Without the glue, the cap can become loose or misaligned. The effect of these faulty valves is that they may not open or close properly which can disturb the fluid transportation within the ISE module resulting in problems with ISE calibration, con

FDA root cause category: Nonconforming Material/Component

What the company is doing

Consignees were notified via an Urgent Medical Device Correction letter dated 11/17/08 and instructed Roche will contact them to make an appointment to inspect their analyzers. A work around was provided, for interim use, in the letter. Customers were instructed to complete and fax back the "ISE Controls Request - Purchased from a Distributor." A no charge replacement product will be shipped to customers.

Recall details

Recall number
Z-1009-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 17, 2008
Posted by the FDA
Mar 24, 2009
Recalling firm
Roche Diagnostics Corp.
Quantity
66
Product code
JJE
FDA submissions
K071211
Distribution
Nationwide Distribution

Product description

Roche COBAS c 111 Analyzer with ISE module option, GMMI No. 04777433001; Roche Diagnostics, Indianapolis, IN. Is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein electrolyte tests. Analytes are measured photometrically or turbidimetrically. The analyzer also has an optional ISE module for measuring sodium, potassium and chloride.

Code information (models, lots, serial numbers)

Serial numbers: 50654, 50694, 50541, 50634, 50251, 50506, 50453, 50688, 50244, 50475, 50597, 50656, 50598, 50635, 50632, 50586, 50320, 50246, 50693, 50692, 50562, 50547, 50624, 50690, 50687, 50680, 50560, 50247, 50538, 50633, 50201, 50612, 50576, 50689, 50681, 50424, 50651, 50661, 50452, 50450, 50290, 50630, 50319, 50250, 50549, 50682, 50683, 50685, 50660, 50545, 50463, 50655, 50539, 50653, 50249, 50563, 50561, 50686, 50649, 50470, 50657, 50448, 50644, 50595, 50575, and 50679.

Other devices under the same FDA submission

2 devices cleared under K071211 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.