Skip to content

Smiths Medical BCI(R) Advisor(R) Vital Signs Monitor, model… recallZ-1326-2009 · Smiths Medical PM, Inc. · initiated Dec 22, 2008

Terminated All recalls posted in May 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical PM, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Recent improvements were made to the Advisor case plastic resulting in the inserts being deeply embedded into the plastic. This affects the ability of the monitor to be adequately secured on the wall mount bracket or roll stand bracket, resulting in the monitor falling off. There have been no adverse events or injuries reported.

FDA root cause category: Device Design

What the company is doing

An Urgent Product Safety Information, Voluntary Recall Notification, dated 12/22/08, was sent to affected consignees. The notification contained a Safety Action Bulletin which provided actions to be taken. The Safety Action Bulletin described the issue, requested customer to contact Smiths Medical to ensure proper mounting hardware is utilized for those consignees using a roll stand or wall mount. If the consignee further distributed the product, they were asked to forward on the Safety Action Bulletin to the end user. Page 3 of the Safety Action Bulletin was asked to be completed and returned to Smiths Medical PM, Inc. A second "Urgent Product safety Information Voluntary Recall Notification" letter dated January 16, 2009 was sent to consignees. The second letter included a risk statement associated with the recalled product.

Recall details

Recall number
Z-1326-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 22, 2008
Posted by the FDA
May 15, 2009
Recalling firm
Smiths Medical PM, Inc.
Quantity
143 monitors (17 US, 126 OUS)
Product code
CCK
FDA submissions
K010770, K011177, K030407, K982279
Distribution
Worldwide Distribution -- USA including states of FL, GA, NJ, NY, and TN and countries of Argentina, Brazil, Ecuador, Greece, Peru, Romania, Russia, Singapore, Slovenia, Trinidad and Tobago, United Kingdom, and Uruguay.

Product description

Smiths Medical BCI(R) Advisor(R) Vital Signs Monitor, model 9200, catalog numbers: 920654225, 920674225, 920674235, 925454220, 925454235, 925454320, 925454325, 925454330, 925454335, 925457325, 925474325, 925654225, 925654325, 925754325, 925754335, and 92M774325. Intended to be used in the ICU, CCU, OR, ER, RR, Labor and Delivery rooms, special procedure labs and other areas of a hospital or clinic where low-end monitoring systems are needed.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

serial numbers: AM08070014, AM08070015, AM08080004, AM08080005, AM08080006, AM08080007, AM08080008, AM08080009, AM08080010, AM08080011, AM08080012, AM08080013, AM08080014, AM08080015, AM08080016, AM08080017, AM08080018, AM08080019, AM08080032, AM08080033, AM08080034, AM08080035, AM08080036, AM08080037, AM08080038, AM08080039, AM08080040, AM08080041, AM08080042, AM08080043, AM08080044, AM08080045, AM08080046, AM08080047, AM08080048, AM08080049, AM08080050, AM08080051, AM08080052, AM08080075, AM08080076, AM08080077, AM08080078, AM08080079, AM08090058, AM08090059, AM08090060, AM08090061, AM08090069, AM08090070, AM08090071, AM08090072, AM08090073, AM08090074, AM08090075, AM08090076, AM08090077, AM08090078, AM08090079, AM08090080, AM08090081, AM08090082, AM08090083, AM08090084, AM08090085, AM08090086, AM08090092, AM08090093, AM08090094, AM08090095, AM08090096, AM08090097, AM08100000, AM08100001, AM08100002, AM08100003, AM08100004, AM08100005, AM08100006, AM08100007, AM08100008, AM08100042, AM08100043, AM08100045, AM08100051, AM08100052, AM08100053, AM08100054, AM08100056, AM08100075, AM08100076, AM08100077, AM08100078, AM08100080, AM08100082, AM08100083, AM08100085, AM08100086, AM08100087, AM08100088, AM08100089, AM08100091, AM08100092, AM08100093, AM08100094, AM08100095, AM08100096, AM08100097, AM08100098, AM08100099, AM08100100, AM08100101, AM08100146, AM08110000, AM08110001, AM08110003, AM08110004, AM08110005, AM08110008, AM08110009, AM08110010, AM08110011, AM08110012, AM08110013, AM08110014, AM08110015, AM08110016, AM08110018, AM08110019, AM08110026, AM08110027, AM08110028, AM08110029, AM08110030, AM08110031, AM08110034, AM08110035, AM08110036, AM08110037, AM08110054, AM08110055, AM08110056, and AM08110058.

Other devices under the same FDA submission

2 devices cleared under K010770, K011177, K030407, K982279 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.