Smiths Medical BCI(R) Advisor(R) Vital Signs Monitor, model… recallZ-1326-2009 · Smiths Medical PM, Inc. · initiated Dec 22, 2008
Terminated All recalls posted in May 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical PM, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Recent improvements were made to the Advisor case plastic resulting in the inserts being deeply embedded into the plastic. This affects the ability of the monitor to be adequately secured on the wall mount bracket or roll stand bracket, resulting in the monitor falling off. There have been no adverse events or injuries reported.
FDA root cause category: Device Design
What the company is doing
An Urgent Product Safety Information, Voluntary Recall Notification, dated 12/22/08, was sent to affected consignees. The notification contained a Safety Action Bulletin which provided actions to be taken. The Safety Action Bulletin described the issue, requested customer to contact Smiths Medical to ensure proper mounting hardware is utilized for those consignees using a roll stand or wall mount. If the consignee further distributed the product, they were asked to forward on the Safety Action Bulletin to the end user. Page 3 of the Safety Action Bulletin was asked to be completed and returned to Smiths Medical PM, Inc. A second "Urgent Product safety Information Voluntary Recall Notification" letter dated January 16, 2009 was sent to consignees. The second letter included a risk statement associated with the recalled product.
Recall details
- Recall number
- Z-1326-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Dec 22, 2008
- Posted by the FDA
- May 15, 2009
- Recalling firm
- Smiths Medical PM, Inc.
- Quantity
- 143 monitors (17 US, 126 OUS)
- Product code
- CCK
- FDA submissions
- K010770, K011177, K030407, K982279
- Distribution
- Worldwide Distribution -- USA including states of FL, GA, NJ, NY, and TN and countries of Argentina, Brazil, Ecuador, Greece, Peru, Romania, Russia, Singapore, Slovenia, Trinidad and Tobago, United Kingdom, and Uruguay.
Product description
Smiths Medical BCI(R) Advisor(R) Vital Signs Monitor, model 9200, catalog numbers: 920654225, 920674225, 920674235, 925454220, 925454235, 925454320, 925454325, 925454330, 925454335, 925457325, 925474325, 925654225, 925654325, 925754325, 925754335, and 92M774325. Intended to be used in the ICU, CCU, OR, ER, RR, Labor and Delivery rooms, special procedure labs and other areas of a hospital or clinic where low-end monitoring systems are needed.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
serial numbers: AM08070014, AM08070015, AM08080004, AM08080005, AM08080006, AM08080007, AM08080008, AM08080009, AM08080010, AM08080011, AM08080012, AM08080013, AM08080014, AM08080015, AM08080016, AM08080017, AM08080018, AM08080019, AM08080032, AM08080033, AM08080034, AM08080035, AM08080036, AM08080037, AM08080038, AM08080039, AM08080040, AM08080041, AM08080042, AM08080043, AM08080044, AM08080045, AM08080046, AM08080047, AM08080048, AM08080049, AM08080050, AM08080051, AM08080052, AM08080075, AM08080076, AM08080077, AM08080078, AM08080079, AM08090058, AM08090059, AM08090060, AM08090061, AM08090069, AM08090070, AM08090071, AM08090072, AM08090073, AM08090074, AM08090075, AM08090076, AM08090077, AM08090078, AM08090079, AM08090080, AM08090081, AM08090082, AM08090083, AM08090084, AM08090085, AM08090086, AM08090092, AM08090093, AM08090094, AM08090095, AM08090096, AM08090097, AM08100000, AM08100001, AM08100002, AM08100003, AM08100004, AM08100005, AM08100006, AM08100007, AM08100008, AM08100042, AM08100043, AM08100045, AM08100051, AM08100052, AM08100053, AM08100054, AM08100056, AM08100075, AM08100076, AM08100077, AM08100078, AM08100080, AM08100082, AM08100083, AM08100085, AM08100086, AM08100087, AM08100088, AM08100089, AM08100091, AM08100092, AM08100093, AM08100094, AM08100095, AM08100096, AM08100097, AM08100098, AM08100099, AM08100100, AM08100101, AM08100146, AM08110000, AM08110001, AM08110003, AM08110004, AM08110005, AM08110008, AM08110009, AM08110010, AM08110011, AM08110012, AM08110013, AM08110014, AM08110015, AM08110016, AM08110018, AM08110019, AM08110026, AM08110027, AM08110028, AM08110029, AM08110030, AM08110031, AM08110034, AM08110035, AM08110036, AM08110037, AM08110054, AM08110055, AM08110056, and AM08110058.
Other devices under the same FDA submission
2 devices cleared under K010770, K011177, K030407, K982279 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BCI WW1100Icu Medical, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 reports
- BCI 920654230AIcu Medical, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.