Medical device recalls: May 2009
The FDA posted 118 medical device recall records in May 2009, from 49 recall events by 43 companies. 118 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Roche Diagnostics Corp. · 2 events, 3 records
- Siemens Medical Solutions USA, Inc · 2 events, 3 records
- Synthes Spine · 2 events, 3 records
- Abbott Laboratories · 2 events, 2 records
- Philips Healthcare Inc. · 2 events, 2 records
- Zimmer Inc. · 2 events, 2 records
- Argon Medical Devices, Inc · 1 event, 41 records
- Boston Scientific Corporation · 1 event, 17 records
FDA root cause
- Other · 43
- Labeling False and Misleading · 18
- Process control · 17
- Device Design · 9
- Software design · 8
- Nonconforming Material/Component · 5
- Process change control · 3
- Component design/selection · 3
Not yet classified (118)
The FDA has posted these records but has not published a class for them yet.
- Z-1224-2009 · Terumo Cardiovascular Systems Corporation · posted May 29, 2009 · Terminated
The monitor may exhibit start-up failure, video/display issues, calibration issues, and partial or complete loss of functionality during use.
- Z-1225-2009 · Medtronic Inc · posted May 29, 2009 · Terminated
Medtronic identified several heart valves (Mosaic Porcine Bioprosthesis with Cinch & Hancock II Bioprosthesis with Cinch) that were inadvertently distributed after exposure to temperatures exceeding the 77 F (25 C) maximum storage temperature published on…
- Z-1226-2009 · Medtronic Inc · posted May 29, 2009 · Terminated
Medtronic identified several heart valves (Mosaic Porcine Bioprosthesis with Cinch & Hancock II Bioprosthesis with Cinch) that were inadvertently distributed after exposure to temperatures exceeding the 77 F (25 C) maximum storage temperature published on…
- Z-1324-2009 · Synthes Spine · posted May 29, 2009 · Terminated
Numbers are etched incorrectly on the product.
- Z-1323-2009 · Biomet Microfixation, Inc. · posted May 27, 2009 · Terminated
This recall is based on the occurrence of known adverse effects identified in the device instructions for use. Bending, fracture, loosening, rubbing and migration of the devices can occur as a result of excessive activity, trauma or load bearing.…
- Z-1392-2009 · Abbott Laboratories · posted May 27, 2009 · Terminated
Abbott has had an increase in complaints due to the receipt of error code 1006 (unable to process test, background read failure), error code 1007 (unable to process test, activated read failure) and incorrect results when using these lots of ARCHITECT…
- Z-1401-2009 · Mindray DS USA, Inc., dba Datascope Patient Monitoring · posted May 27, 2009 · Terminated
Software Anomaly: Software anomaly can occur when the Panorama is attempting to play an audio file, either a physiological or technical alarm sound, at which time the system software may cause the system to freeze due to an interaction in the alarm system.
- Z-1421-2009 · Neuronetics, Inc. · posted May 27, 2009 · Terminated
SenStar Treatment Link may have an open circuit that during use with the NeuroStar TMS Therapy system may cause arcing and heat damage to the SenStar and/or treatment coil. Heating of the SenStar during use has a potential risk of minor scalp burn to…
- Z-1436-2009 · Philips Healthcare Inc. · posted May 27, 2009 · Terminated
Software: A software defect under certain circumstances could lead to the unintended administration of a duplicate dose of a scheduled medication within a single 24 hour period
- Z-1267-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1268-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1269-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1270-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1271-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1272-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1273-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1274-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1275-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1276-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1277-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1278-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1279-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1280-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1281-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1282-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1283-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1284-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1285-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1286-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1287-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1288-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1289-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1290-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1291-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1292-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1293-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1294-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1295-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1296-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1297-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1298-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1299-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1300-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1301-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1302-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1303-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1304-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1305-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1306-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1307-2009 · Argon Medical Devices, Inc · posted May 26, 2009 · Terminated
Hydrophilic guidewires have the potential to demonstrate degradation of the Pebax cladding.
- Z-1399-2009 · Siemens Medical Solutions Diagnostics · posted May 26, 2009 · Terminated
Design Flaw: The redesign of the Digital Fluidics PCB used in IMMULITE 2000 and IMMULITE 2500 is missing a jumper.
- Z-1417-2009 · Remington Medical Inc. · posted May 26, 2009 · Terminated
Incorrect needle (Franseen Needle, FNM-1806) was packaged as a Chiba Needle Marked.
- Z-1418-2009 · Stryker Communications Corp · posted May 26, 2009 · Terminated
Potential for health care provider/hospital staff member to sustain injury to fingers or hands when removing/handling the headrest component of the VERTIER Surgical Table.
- Z-1643-2009 · Alcon Laboratories, Inc · posted May 26, 2009 · Terminated
Specific lots of Alcon 25 GA Trocar Blades do not conform to finished product specifications in that oxidation can occur on some of the blades.
- Z-1644-2009 · Alcon Laboratories, Inc · posted May 26, 2009 · Terminated
Specific lots of Alcon 25 GA Trocar Blades do not conform to finished product specifications in that oxidation can occur on some of the blades.
- Z-1645-2009 · Kimberly-Clark Corporation · posted May 26, 2009 · Terminated
The product may contain defects/holes reducing the product's effectiveness as a barrier.
- Z-1646-2009 · Disetronic Medical Systems, Inc. · posted May 26, 2009 · Terminated
The "up" and "down" buttons, which are used for changing the program in the menu or to administer additional insulin through a bolus delivery, may experience intermittent failure or completely cease to operate.
- Z-1256-2009 · HemoSense Inc · posted May 22, 2009 · Terminated
INRatio Monitors that were manufactured with a specific EEPROM memory chip fail during use. The monitors report an error code and do not report a test result.
- Z-1265-2009 · Imaging Sciences International, LLC · posted May 22, 2009 · Terminated
faulty calibration
- Z-1396-2009 · Beckman Coulter Inc · posted May 22, 2009 · Terminated
Beckman Coulter initiated the field action as a result of recent internal testing following the CLSI EP-5A - Evaluation of Precision Performance of Clinical Chemistry Devices protocol, it was determined that the total imprecision for the Access Myoglobin…
- Z-1398-2009 · Abbott Laboratories · posted May 22, 2009 · Terminated
Burn Risk-- when power is cycled to the instrument, there is a possibility that the Mixer Motor Assembly may reach 100 degrees Celsius (212 degrees Fahrenheit ). If the customer touches the motor, burns may occur.
- Z-1638-2009 · Aspen Surgical Products, Inc. · posted May 22, 2009 · Terminated
An adaptor, which may be required when connecting the drain to an external device, was not included in the package.
- Z-1639-2009 · Aspen Surgical Products, Inc. · posted May 22, 2009 · Terminated
An adaptor, which may be required when connecting the drain to an external device, was not included in the package.
- Z-1331-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1332-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1333-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1334-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1335-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1336-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1337-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1338-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1339-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1340-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1341-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1342-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1343-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1344-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1345-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1346-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1347-2009 · Boston Scientific Corporation · posted May 21, 2009 · Terminated
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the…
- Z-1640-2009 · Synthes Spine · posted May 21, 2009 · Terminated
Sterility may be compromised due to a ineffective package design.
- Z-1641-2009 · Synthes Spine · posted May 21, 2009 · Terminated
Sterility may be compromised due to a ineffective package design.
- Z-1642-2009 · Stryker Howmedica Osteonics Corp. · posted May 21, 2009 · Terminated
Packaging burst test values for the Restoration Modular Cone Body family were found to be incorrectly transferred from the finalized validation to the work specification. Testing showed that burst test results for recalled products fell below the validated…
- Z-1196-2009 · Abbott Diabetes Care, Inc. · posted May 20, 2009 · Terminated
The product has the Units of Measure set to mmol/L rather than mg/dL. The blood glucose results could be misinterpreted. This may lead to under treatment and potential for hyperglycemia.
- Z-1393-2009 · Roche Diagnostics Corp. · posted May 20, 2009 · Terminated
Some data alarm flags do not transfer from the analyzer to the laboratory information system, since not all data flags are recognized by the instrument manager/middleware.
- Z-1394-2009 · Roche Diagnostics Corp. · posted May 20, 2009 · Terminated
Some data alarm flags do not transfer from the analyzer to the laboratory information system, since not all data flags are recognized by the instrument manager/middleware.
- Z-1195-2009 · GE Medical Systems, LLC · posted May 19, 2009 · Terminated
GE Healthcare has become aware of a potential image storage failure on Innova 2121 IQ/ 3131 IQ cardiovascular imaging systems that may impact patient safety. This intermittent communication failure on the image storage chain may result in the following…
- Z-1199-2009 · Philips Healthcare Inc. · posted May 19, 2009 · Terminated
Rotary energy select switch in affected M4735A devices may fail and prevent the user from turning the device on, rendering the device unusable for monitoring and defibrillation therapy.
- Z-1395-2009 · Bio-Rad Laboratories Inc · posted May 19, 2009 · Terminated
The wrong barcode number and tube position number may be assigned to samples leading to discrepant sample results.
- Z-1400-2009 · Remel, Inc · posted May 19, 2009 · Terminated
These lots may exhibit false positive reactions with strains of Staphylococcus epidermidis that may result in the mis-identification of Staphylococcus aureus.
- Z-1411-2009 · Respironics, Inc. · posted May 19, 2009 · Terminated
Audible alarm failure.
- Z-1255-2009 · Roche Diagnostics Corp. · posted May 18, 2009 · Terminated
Protruding screw threads may cut the user when cleaning the interior of the instrument.
- Z-1261-2009 · Zimmer Inc. · posted May 18, 2009 · Terminated
The package is labeled as containing 330 mm length nails, but actually contains 300 mm length nails.
- Z-1326-2009 · Smiths Medical PM, Inc. · posted May 15, 2009 · Terminated
Recent improvements were made to the Advisor case plastic resulting in the inserts being deeply embedded into the plastic. This affects the ability of the monitor to be adequately secured on the wall mount bracket or roll stand bracket, resulting in the…
- Z-1397-2009 · GE Healthcare Integrated IT Solutions · posted May 15, 2009 · Terminated
A software defect in the Centricity Lab (TriWin) product may incorrectly link patient information to a wrong report. Also, when viewing results in Patient Viewer, and a coagulation result is higher or lower than the linearity limit, then all of the…
- Z-1263-2009 · Draeger Medical, Inc. · posted May 14, 2009 · Terminated
The device may not generate a pilot line disconnect alarm if the pilot line becomes disconnected, which may potentially result in a substantial reduction in tidal volume delivery.
- Z-1327-2009 · Diagnostic Chemicals, Ltd. · posted May 13, 2009 · Terminated
Precipitate in product may cause elevated background absorbance interference.
- Z-1252-2009 · MAQUET Inc. · posted May 12, 2009 · Terminated
The expiratory valve in the SV 300 ventilator may stay in the open position if the printed circuit board fails.
- Z-1262-2009 · Zimmer Inc. · posted May 12, 2009 · Terminated
The surface finish inside of the broach is not to specification and the instrument may corrode if cleaned and sterilized according to instructions.
- Z-1309-2009 · Ethicon, Inc. · posted May 12, 2009 · Terminated
Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.
- Z-1310-2009 · Ethicon, Inc. · posted May 12, 2009 · Terminated
Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.
- Z-1311-2009 · Ethicon, Inc. · posted May 12, 2009 · Terminated
Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.
- Z-1312-2009 · Ethicon, Inc. · posted May 12, 2009 · Terminated
Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.
- Z-1313-2009 · Ethicon, Inc. · posted May 12, 2009 · Terminated
Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.
- Z-1314-2009 · Ethicon, Inc. · posted May 12, 2009 · Terminated
Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.
- Z-1315-2009 · Ethicon, Inc. · posted May 12, 2009 · Terminated
Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.
- Z-1316-2009 · Ethicon, Inc. · posted May 12, 2009 · Terminated
Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.
- Z-1209-2009 · Medtronic Spine LLC, formerly Kyphon Inc · posted May 11, 2009 · Terminated
Product not sterile: 10 units were released for distribution without undergoing sterilization process.
- Z-1391-2009 · Biosite Inc · posted May 8, 2009 · Terminated
The recall was initiated because the use of the affected lot may lead to reporting falsely low values for troponin I and myoglobin, possibly resulting in missed or incorrect diagnosis.
- Z-1233-2009 · Siemens Medical Solutions USA, Inc · posted May 7, 2009 · Terminated
Image may appear dark, requiring repeat image acquisition.
- Z-1328-2009 · Nova Biomedical Corporation · posted May 7, 2009 · Terminated
Erroneous, low creatinine results on CCX analyzer
- Z-1244-2009 · Migada Inc. · posted May 6, 2009 · Terminated
Syringe adaptor is disintegrating when attached to the end of the syringe.
- Z-1253-2009 · Siemens Medical Solutions USA, Inc · posted May 6, 2009 · Terminated
During interventional use and applications, the possibility exists for liquids to enter the systems.
- Z-1254-2009 · Siemens Medical Solutions USA, Inc · posted May 6, 2009 · Terminated
During interventional use and applications, the possibility exists for liquids to enter the systems.
- Z-1232-2009 · GE Medical Systems Information Tecnology · posted May 4, 2009 · Terminated
GE Healthcare is aware of potential network time issues associated with the time change aspects of certain CARESCAPE CIC Pro" software versions. These issues may impact patient safety. No reported patient injuries or illnesses for the use of this device…
- Z-1236-2009 · Cerner Corp · posted May 1, 2009 · Terminated
Specimen labels printing incorrect patient demographics.
- Z-1237-2009 · Cardinal Health 303 dba Cardinal Health · posted May 1, 2009 · Terminated
The Alaris PCA module can potentially infuse above or below the intended infusion dose.
- Z-1317-2009 · Hill-Rom, Inc. · posted May 1, 2009 · Terminated
An e-ring may not have been installed during production, which could result in an inability to latch the siderail.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.