Skip to content

AxSOS T8 Screwdriver, 3.0mm Locking Set. REF 702746. This… recallZ-1062-2009 · Stryker Howmedica Osteonics Corp. · initiated Nov 24, 2008

Terminated All recalls posted in April 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential Breakage: Stryker Orthopaedics became aware that there is the potential for the tip of the AxSOS Screwdriver and the AxSOS Torque Limiter to break.

FDA root cause category: Other

What the company is doing

"Urgent Product Recall" letters were sent to Stryker Orthopaedics Branches/Agencies, Chief of Orthopaedics, Hospital Risk Management and Surgeons on November 24, 2008 via Federal Express. The letters described the issue and provided the following instructions: 1. Examine your inventory and determine if you have any of the affected product, 2. Remove and quarantine affected product from your inventory, 3. Complete the attached Product Recall Acknowledgement Form (and fax to Stryker Orthopaedics at 201-831-6069), and 4. Return the affected product to your Stryker Orthopaedics Sales representative. Direct questions to your Stryker Orthopaedics Sales Representative or call Stryker Orthopaedics at 201-831-5000.

Recall details

Recall number
Z-1062-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 24, 2008
Posted by the FDA
Apr 13, 2009
Recalling firm
Stryker Howmedica Osteonics Corp.
Quantity
467 units
Product code
HRS
FDA submissions
K061012
Distribution
Nationwide Distribution.

Product description

AxSOS T8 Screwdriver, 3.0mm Locking Set. REF 702746. This product is non sterile. Distributed in the USA by Howmedica Osteonics Corp, Mahwah, NJ. The screwdrivers of the AxSOS system are multiple use instruments used to insert screws into bones.

Code information (models, lots, serial numbers)

All lot codes.

Other devices under the same FDA submission

3 devices cleared under K061012 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 50850).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page