Medical device recalls: April 2009
The FDA posted 137 medical device recall records in April 2009, from 52 recall events by 41 companies. 137 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Abbott Laboratories · 3 events, 8 records
- GE Medical Systems, LLC · 3 events, 8 records
- Pointe Scientific, Inc. · 2 events, 3 records
- Smiths Medical MD, Inc. · 2 events, 3 records
- Stryker Howmedica Osteonics Corp. · 2 events, 3 records
- Biomet, Inc. · 2 events, 2 records
- DiaDexus, Inc · 2 events, 2 records
- Immuno Concepts Inc · 2 events, 2 records
FDA root cause
- Process control · 39
- Process design · 32
- Software design · 16
- Device Design · 11
- Other · 9
- Packaging process control · 6
- Labeling design · 6
- Employee error · 6
Not yet classified (137)
The FDA has posted these records but has not published a class for them yet.
- Z-1182-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1183-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1184-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1185-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1186-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1187-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1188-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1189-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1190-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1191-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1192-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1193-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1194-2009 · Stryker Instruments Div. of Stryker Corporation · posted Apr 30, 2009 · Terminated
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
- Z-1308-2009 · Sybaritic, Inc · posted Apr 30, 2009 · Terminated
Sybaritic stated that no 510(k) has been filed for the device (Lumi 8 Intense LED Light Facial System). No injuries or illness reported.
- Z-1319-2009 · Tri-State Hospital Supply Corporation · posted Apr 30, 2009 · Terminated
Compromised sterility, as the package seals may be inadequate.
- Z-1320-2009 · Tri-State Hospital Supply Corporation · posted Apr 30, 2009 · Terminated
Compromised sterility, as the package seals may be inadequate.
- Z-1321-2009 · Tri-State Hospital Supply Corporation · posted Apr 30, 2009 · Terminated
Compromised sterility, as the package seals may be inadequate.
- Z-1322-2009 · Tri-State Hospital Supply Corporation · posted Apr 30, 2009 · Terminated
Compromised sterility, as the package seals may be inadequate.
- Z-1197-2009 · Beckman Coulter Inc · posted Apr 28, 2009 · Terminated
Beckman Coulter incorrectly shipped Access Hybritech PSA calibrators containing dual calibration cards (Hybritech and WHO) to some Access Hybritech PSA Customers. The WHO calibration for PSA is not currently approved in the United States.
- Z-1208-2009 · Wright Medical Technology Inc · posted Apr 28, 2009 · Terminated
The package insert lists 3 sterilization procedures, 2 of which are not validated. Pre-vacuum steam sterilization is validated.
- Z-1234-2009 · Dentsply Worldwide Headqrt · posted Apr 28, 2009 · Terminated
The product is manufactured with a left-hand thread instead of a right hand thread. This error could cause tooth breakage if the posts were removed without knowing that the threads were reversed.
- Z-1235-2009 · Siemens Medical Solutions USA, Inc · posted Apr 28, 2009 · Terminated
Incorrect and/or incomplete data: Several potential safety issues while working with the syngo Image product including, but not limited to, data loss, errors in patient data due to different patient IDs, delays in image availability, inaccurate results…
- Z-1266-2009 · Siemens Medical Solutions USA, Inc · posted Apr 28, 2009 · Terminated
An issue concerning archived image data that is sporadically saved with corrupted data may occur. This issue may occur when syngo Imaging is configured to archive image data on a network file system (NFS) connected via a connected archive (Mount Point…
- Z-1123-2009 · Berchtold Corp. · posted Apr 27, 2009 · Terminated
Difficulty in pressing the button needed to release a bed part resulted in an injury to a caregiver's finger which became stuck in metal latch under the gas cylinder of the bed piece.
- Z-1200-2009 · Pointe Scientific, Inc. · posted Apr 27, 2009 · Terminated
The R1 reagent has microbial contamination.
- Z-1201-2009 · Pointe Scientific, Inc. · posted Apr 27, 2009 · Terminated
The R1 reagent has microbial contamination.
- Z-1069-2009 · Abbott Laboratories · posted Apr 23, 2009 · Terminated
Some syringes were improperly assembled and are therefore not usable, due to an incorrect nut on the syringe shell.
- Z-1205-2009 · DiaDexus, Inc · posted Apr 23, 2009 · Terminated
Premature Expiration-- the product is projected to fall short of expected shelf life, and may result in inaccurate results.
- Z-0994-2009 · Cyberonics, Inc · posted Apr 22, 2009 · Terminated
Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference resulting in the loss of stimulation.
- Z-0995-2009 · Cyberonics, Inc · posted Apr 22, 2009 · Terminated
Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference, resulting in the loss of stimulation.
- Z-1001-2009 · GE Medical Systems, LLC · posted Apr 21, 2009 · Terminated
GE Healthcare has recently became aware of excessive pressure applied by the collimator pressure sensitive cover of Infinia (Infinia, Infinia Hawkeye, Infinia Hawkeye 4) that may impact patient safety. It was reported at a customer site that the pressure…
- Z-1002-2009 · GE Medical Systems, LLC · posted Apr 21, 2009 · Terminated
GE Healthcare has recently became aware of excessive pressure applied by the collimator pressure sensitive cover of Infinia (Infinia, Infinia Hawkeye, Infinia Hawkeye 4) that may impact patient safety. It was reported at a customer site that the pressure…
- Z-1003-2009 · GE Medical Systems, LLC · posted Apr 21, 2009 · Terminated
GE Healthcare has recently became aware of excessive pressure applied by the collimator pressure sensitive cover of Infinia (Infinia, Infinia Hawkeye, Infinia Hawkeye 4) that may impact patient safety. It was reported at a customer site that the pressure…
- Z-1004-2009 · Smith & Nephew Inc · posted Apr 21, 2009 · Terminated
Liners measuring 28 mm ID and were mismarked as 32 mm.
- Z-0987-2009 · Smiths Medical MD, Inc. · posted Apr 20, 2009 · Terminated
Smiths Medical noticed an increased trend in leakage associated with 100 ml CADD Medication Cassette Reservoirs Reorder Number 21-7002-24 Lot Numbers 213X18 and 214X18. If a leak occurs in a CADD medication Cassette Reservoir, it may allow spillage of…
- Z-1202-2009 · Pointe Scientific, Inc. · posted Apr 17, 2009 · Terminated
The product performance fails due to precipitation of the R2 component.
- Z-1203-2009 · Hamilton Co · posted Apr 17, 2009 · Terminated
Some tips may be occluded and are unable to aspirate or dispense liquid.
- Z-1207-2009 · Biomet, Inc. · posted Apr 17, 2009 · Terminated
A 30 ml tube is packaged with the kit instead of the labeled 60 ml tube.
- Z-1238-2009 · Edwards Lifesciences, Llc · posted Apr 17, 2009 · Terminated
Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".
- Z-1239-2009 · Edwards Lifesciences, Llc · posted Apr 17, 2009 · Terminated
Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".
- Z-1240-2009 · Edwards Lifesciences, Llc · posted Apr 17, 2009 · Terminated
Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".
- Z-1241-2009 · Edwards Lifesciences, Llc · posted Apr 17, 2009 · Terminated
Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".
- Z-1242-2009 · Edwards Lifesciences, Llc · posted Apr 17, 2009 · Terminated
Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".
- Z-1243-2009 · Edwards Lifesciences, Llc · posted Apr 17, 2009 · Terminated
Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".
- Z-1250-2009 · Cardiovascular Systems, Inc. · posted Apr 17, 2009 · Terminated
CSI is recalling 10 devices due to a mislabeling of the shelf life as shown on the device labeling. The labeling shown on the affected devices represents a 24 month shelf life rather than a 6 month shelf life that should have been indicated.
- Z-1251-2009 · Stryker Howmedica Osteonics Corp. · posted Apr 17, 2009 · Terminated
Instruction for use incorrect: There are no statements or warning on labels or Instructions for Use that 8, 9 and 10 mm diameter MRS Cemented Stems are indicated only for use in the GMRS Proximal Tibial and Distal Femoral Components.
- Z-0988-2009 · Smiths Medical MD, Inc. · posted Apr 16, 2009 · Terminated
Smiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 software. There have been adverse events reported that the amount of Extended Bolus delivered was not…
- Z-0989-2009 · Smiths Medical MD, Inc. · posted Apr 16, 2009 · Terminated
Smiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 software. There have been adverse events reported that the amount of Extended Bolus delivered was not…
- Z-1210-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1211-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1212-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1213-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1214-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1215-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1216-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1217-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1218-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1219-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1220-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1221-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1222-2009 · Arrow International, Inc. · posted Apr 16, 2009 · Terminated
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Z-1046-2009 · Abbott Laboratories · posted Apr 15, 2009 · Terminated
Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
- Z-1047-2009 · Abbott Laboratories · posted Apr 15, 2009 · Terminated
Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
- Z-1048-2009 · Abbott Laboratories · posted Apr 15, 2009 · Terminated
Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
- Z-1049-2009 · Abbott Laboratories · posted Apr 15, 2009 · Terminated
Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
- Z-1050-2009 · Abbott Laboratories · posted Apr 15, 2009 · Terminated
Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
- Z-1051-2009 · Abbott Laboratories · posted Apr 15, 2009 · Terminated
Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
- Z-1198-2009 · Datascope Corporation · posted Apr 15, 2009 · Terminated
Non-functioning cable: the Low-Level Output Cable when connected to a Datascope Intra-aortic Ballon Pump (IABP) and patient monitor does not provide a signal to the patient monitor.
- Z-1173-2009 · Medtronic Inc · posted Apr 14, 2009 · Terminated
Medtronic is recalling one lot of STREAMLINE Bipolar Myocardial Temporary Pacing Leads, Model 6495, Lot JGD880205F. This lot of product contains incorrect inner pouch labeling, instead of 6495 it is labeled 6500. The outer carton, shelf box labeling, and…
- Z-1062-2009 · Stryker Howmedica Osteonics Corp. · posted Apr 13, 2009 · Terminated
Potential Breakage: Stryker Orthopaedics became aware that there is the potential for the tip of the AxSOS Screwdriver and the AxSOS Torque Limiter to break.
- Z-1063-2009 · Stryker Howmedica Osteonics Corp. · posted Apr 13, 2009 · Terminated
Potential Breakage: Stryker Orthopaedics became aware that there is the potential for the tip of the AxSOS Screwdriver and the AxSOS Torque Limiter to break.
- Z-1153-2009 · MAQUET Inc. · posted Apr 12, 2009 · Terminated
The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report.
- Z-1154-2009 · MAQUET Inc. · posted Apr 12, 2009 · Terminated
The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report.
- Z-0986-2009 · Stryker Spine · posted Apr 10, 2009 · Terminated
Reports were received of the ball ring ejecting from the connector's body when other connectors have already been locked down.
- Z-1064-2009 · Abbott Laboratories · posted Apr 10, 2009 · Terminated
Carryover failures in software revisions 1.0ML and 2.0ML result in elevated PLT background count.
- Z-1065-2009 · Roche Diagnostics Corp. · posted Apr 10, 2009 · Terminated
If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.
- Z-1066-2009 · Roche Diagnostics Corp. · posted Apr 10, 2009 · Terminated
If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.
- Z-1067-2009 · Roche Diagnostics Corp. · posted Apr 10, 2009 · Terminated
If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.
- Z-1068-2009 · Roche Diagnostics Corp. · posted Apr 10, 2009 · Terminated
If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.
- Z-1094-2009 · GE Medical Systems, LLC · posted Apr 10, 2009 · Terminated
There is a possibility for misregistration when brain scans are performed with Infinia with the Hawkeye option. This misregistration could present the potential for misdiagnosis when the patient headrest sags under the weight of a head.
- Z-1095-2009 · GE Medical Systems, LLC · posted Apr 10, 2009 · Terminated
There is a possibility for misregistration when brain scans are performed with Infinia with the Hawkeye option. This misregistration could present the potential for misdiagnosis when the patient headrest sags under the weight of a head.
- Z-1119-2009 · GE Medical Systems, LLC · posted Apr 10, 2009 · Terminated
The metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential for misdiagnosis.
- Z-1120-2009 · GE Medical Systems, LLC · posted Apr 10, 2009 · Terminated
The metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential for misdiagnosis.
- Z-1121-2009 · GE Medical Systems, LLC · posted Apr 10, 2009 · Terminated
The metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential for misdiagnosis.
- Z-1156-2009 · St. Jude Medical · posted Apr 10, 2009 · Terminated
Installation problem: there is a risk that the monitor may not be adequately secured to the stand, resulting in the monitor falling off which may cause injury.
- Z-1071-2009 · Boston Scientific Corp · posted Apr 9, 2009 · Terminated
The marker band may detach from the sheath and a clinical failure could result in an embolism.
- Z-1072-2009 · Boston Scientific Corp · posted Apr 9, 2009 · Terminated
The marker band may detach from the sheath and a clinical failure could result in an embolism.
- Z-1175-2009 · Teleflex Medical · posted Apr 9, 2009 · Terminated
The product has a rough square finish instead of a smooth rounded finish which can cause damage to arteries.
- Z-1176-2009 · Teleflex Medical · posted Apr 9, 2009 · Terminated
The product has a rough square finish instead of a smooth rounded finish which can cause damage to arteries.
- Z-1177-2009 · Teleflex Medical · posted Apr 9, 2009 · Terminated
The product has a rough square finish instead of a smooth rounded finish which can cause damage to arteries.
- Z-1178-2009 · Teleflex Medical · posted Apr 9, 2009 · Terminated
The product has a rough square finish instead of a smooth rounded finish which can cause damage to arteries.
- Z-1179-2009 · Teleflex Medical · posted Apr 9, 2009 · Terminated
The product has a rough square finish instead of a smooth rounded finish which can cause damage to arteries.
- Z-1180-2009 · Teleflex Medical · posted Apr 9, 2009 · Terminated
The product has a rough square finish instead of a smooth rounded finish which can cause damage to arteries.
- Z-1181-2009 · Teleflex Medical · posted Apr 9, 2009 · Terminated
The product has a rough square finish instead of a smooth rounded finish which can cause damage to arteries.
- Z-1096-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1097-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1098-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1099-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1100-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1101-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1102-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1103-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1104-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1105-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1106-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1107-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1108-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1109-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1110-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1111-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1112-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1113-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1114-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1115-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1116-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1117-2009 · C R Bard Inc · posted Apr 8, 2009 · Terminated
Out of specification replacement gastrostomy device was distributed.
- Z-1172-2009 · Medtronic Inc. Cardiac Rhythm Managment · posted Apr 8, 2009 · Terminated
Medtronic has identified an issue related to installation or removal of the Lead Integrity Alert software in EnTrust(R) and EnTrust(R) Escudo" defibrillators. In those devices only, installation or removal of LIA will inadvertently turn off two audible…
- Z-0998-2009 · Cardinal Health, Inc · posted Apr 7, 2009 · Terminated
Internal testing has identified several several scenarios where unexpected System lockups, shutdowns or system errors would render the Sonara unavailable for use until the Sonara can be restarted. These issues can be corrected by installing the Version 6.0…
- Z-0999-2009 · Cardinal Health, Inc · posted Apr 7, 2009 · Terminated
Internal testing has identified several several scenarios where unexpected System lockups, shutdowns or system errors would render the Sonara unavailable for use until the Sonara can be restarted. These issues can be corrected by installing the Version 6.0…
- Z-1061-2009 · Biomet, Inc. · posted Apr 7, 2009 · Terminated
A 2.5mm locking insert was packaged with the 5.0mm Offset Tibial Tray Adaptor instead of the 5.0mm locking insert.
- Z-1170-2009 · Cardinal Health 303 dba Cardinal Health · posted Apr 7, 2009 · Terminated
The recall was initiated as a result of a potential for out of specification flow rates. The product may not consistently meet the labeled flow rate range of +/- 15%.
- Z-1171-2009 · Physio Control, Inc. · posted Apr 7, 2009 · Terminated
Sterilization efficacy can not be confirmed for the current methods of sterilization listed in the product labeling.
- Z-1174-2009 · Integra LifeSciences Corp. · posted Apr 7, 2009 · Terminated
Package Seal Integrity: Integra became aware of the potential of open seals in the foil pouch of the Integra Bilayer Matrix Wound Dressings.
- Z-1039-2009 · GE Healthcare · posted Apr 6, 2009 · Terminated
GE Healthcare is providing information about two issues associated with iCentral L-NETO5 and L-NETC05 v4.3, 4.4, 4.5 and 5.0 when used in combination with Dash 2000, 3000, 4000 and 5000; Solar 7000, 8000, 8000M and 8000i; Eagle 3000 and 4000 and ApexPro.…
- Z-1044-2009 · Immuno Concepts Inc · posted Apr 6, 2009 · Terminated
Samples can occasionally run out of the wells and cross contaminate adjacent wells.
- Z-1045-2009 · Immuno Concepts Inc · posted Apr 6, 2009 · Terminated
In certain patient samples, it is difficult to properly discriminate between anti-Sm/RMP positive samples and samples that are only anti-Sm positive.
- Z-1230-2009 · Stryker Medical Div. of Stryker Corporation · posted Apr 5, 2009 · Terminated
The head section backrest (fowler) may not raise up, or, if up, may drop suddenly due to wearing out of the actuator threads on the fowler.
- Z-1231-2009 · Stryker Medical Div. of Stryker Corporation · posted Apr 5, 2009 · Terminated
The head section backrest (fowler) may not raise up, or, if up, may drop suddenly due to wearing out of the actuator threads on the fowler.
- Z-1053-2009 · Zimmer Inc. · posted Apr 2, 2009 · Terminated
The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.
- Z-1054-2009 · Zimmer Inc. · posted Apr 2, 2009 · Terminated
The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.
- Z-1055-2009 · Zimmer Inc. · posted Apr 2, 2009 · Terminated
The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.
- Z-1056-2009 · Zimmer Inc. · posted Apr 2, 2009 · Terminated
The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.
- Z-1057-2009 · Zimmer Inc. · posted Apr 2, 2009 · Terminated
The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.
- Z-1058-2009 · Zimmer Inc. · posted Apr 2, 2009 · Terminated
The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.
- Z-1122-2009 · Trumpf Medical Systems, Inc. · posted Apr 1, 2009 · Terminated
Cracks may form around the screw connections of the cover plates causing brittleness, which may lead to fragments breaking off and falling into the operating area.
- Z-1204-2009 · DiaDexus, Inc · posted Apr 1, 2009 · Terminated
Storage equipment failure at distributor site due to Hurricane Ike caused product to experience a temperature variation and go out of range. The integrity of the product could not be assured.
- Z-1206-2009 · ZOLL Medical Corporation, World Wide Headquarters · posted Apr 1, 2009 · Terminated
Device fails to discharge the defibrillation energy.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.