Biomet Offset Tibial Tray Adaptor, 5.0 MM, sterile, Biomet… recallZ-1061-2009 · Biomet, Inc. · initiated Feb 13, 2009
Terminated All recalls posted in April 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A 2.5mm locking insert was packaged with the 5.0mm Offset Tibial Tray Adaptor instead of the 5.0mm locking insert.
FDA root cause category: Packaging process control
What the company is doing
Consignees were notified by "Field Safety Corrective Action" letter dated February 13, 2009 to conduct a recall to the hospital level and to locate and return the recalled product to Biomet Orthopedics. Customers were also instructed to complete the enclosed "Fax Back Response Form" and return it via fax to Biomet Orthopedics at 1-574-372-1683. Direct questions about this recall to 1-800-348-9500 or 1-574-372-3983, Monday through Friday, 8 a.m. to 5 p.m.
Recall details
- Recall number
- Z-1061-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 13, 2009
- Posted by the FDA
- Apr 7, 2009
- Recalling firm
- Biomet, Inc.
- Quantity
- 17
- Product code
- JDW
- FDA submissions
- K010212
- Distribution
- Worldwide Distribution including Germany, Italy, Netherlands, Spain and the United Kingdom.
Product description
Biomet Offset Tibial Tray Adaptor, 5.0 MM, sterile, Biomet Orthopedics, Inc., Warsaw, IN, REF 141491. The Biomet Offset Tibial Tray Adaptor is a knee prosthesis intended for cemented use in patients with the following indications: painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved, correction of valgus, varus or posttraumatic deformity, correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.
Code information (models, lots, serial numbers)
Lot 644290.
Other devices under the same FDA submission
19 devices cleared under K010212 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BIOMET KNEE SYSTEM 141215Biomet Orthopedics, LLC · Prosthesis, knee, patello/femorotibial, semi-constrained, uhmwpe, pegged, cemented, polymer/metal/polymer1 recall3 reports
- Biomet® Knee System 141483Biomet Orthopedics, LLC · Prosthesis, knee, patello/femorotibial, semi-constrained, uhmwpe, pegged, cemented, polymer/metal/polymer2 reports
- Biomet® Knee System 141211Biomet Orthopedics, LLC · Prosthesis, knee, patello/femorotibial, semi-constrained, uhmwpe, pegged, cemented, polymer/metal/polymer2 reports
- Biomet® Knee System 141212Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER2 reports
- Biomet® Knee System 141213Biomet Orthopedics, LLC · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive2 reports
- BIOMET KNEE SYSTEM 141212Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- Biomet® Knee System 141491Biomet Orthopedics, LLC · Prosthesis, knee, patello/femorotibial, semi-constrained, uhmwpe, pegged, cemented, polymer/metal/polymer1 report
- BIOMET KNEE SYSTEM 141216Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- BIOMET KNEE SYSTEM 141213Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- Biomet Knee System 141264Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Biomet Knee System 141263Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- BIOMET KNEE SYSTEM 141481Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗