GUIDANT VITALITY DS VR Model T135 VR recallZ-1621-2009 · Boston Scientific CRM Corp · initiated Mar 4, 2009

Terminated All recalls posted in July 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific CRM Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

In April 2007, Boston Scientific CRM communicated with physicians regarding the potential for reduced Elective Replacement Indicator (ERI) to Battery End of Life (EOL) time in a subset of implantable defibrillators due to degradation of a low-voltage capacitor. Since that time, the April 2007 advisory population has not experienced any clinically significant changes to either the rate of occurrenc

FDA root cause category: Nonconforming Material/Component

What the company is doing

Boston Scientific CRM notified Doctors via sales representatives of an "Urgent Medical Device Information" letter dated March 4, 2009. The letter described the problem and included Description of Issue, Rate of Occurrence, Patient Management Recommendations, Devices Affected, Warranty Program & Further Information. A Boston Scientific " Product Advisory" letter dated March 4, 2009, was also made available to physicians to distribute to patients. The letter informed the patient that their defibrillator was included in a group of devices that might experience a performance issue. For further questions, Contact Boston Scientific CRM at 1-800-227-3422.

Recall details

Recall number
Z-1621-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 4, 2009
Posted by the FDA
Jul 17, 2009
Recalling firm
Boston Scientific CRM Corp
Quantity
1271
Product code
LWP
FDA submissions
P960040
Distribution
Worldwide Distribution - US including PR and states of AL, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI and WY and countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ITALY, JAPAN, LEBANON, NETHERLANDS, NEW ZEALAND, POLYNESIA, PORTUGAL, SPAIN, SWITZERLAND, TURKEY and UNITED KINGDOM.

Product description

GUIDANT VITALITY DS VR Model T135 VR; EL Model T127 DR; 2ICD Model T165 DR; 2ICD Model T175 VR; 2EL ICD Model T167 DR; 2 EL ICD Model T177 VR; DS Model T125 DR. Implantable Cardioverter Defibrillator, Sterile Lot. Guidant Corporation, Hamline Avenue North, St Paul, MN 551124-6798. Indicated for patients who are at risk for sudden cardiac death due to ventricular arrhythmias and may require pacing support.

Code information (models, lots, serial numbers)

Show the full code information (7k characters)

SERIAL NUMBERS MODEL #T125:  115369, 115371, 115379, 115380, 115381, 115402, 115404, 115406, 115407, 115411, 115412, 115415, 115416, 115417, 115418, 115419, 115420, 115421, 115422, 115423, 115426, 115428, 115430, 115431, 115435, 115438, 115439, 115441, 115502, 115503, 115507, 115514, 115517, 115518, 115520, 115527, 115529, 115531, 115533, 115534, 115536, 115538, 115539, 115540, 115541, 115552, 115553, 115554, 115555, 115556, 115557, 115559, 115562, 115565, 115566, 115568, 115569, 115571, 115573, 115574, 115577, 115579, 115581, 115585, 115586, 115587, 115588, 115589, 115590, 115592, 115593, 115594, 115596, 115598, 115599, 115600, 115601, 115603, 115606, 115608, 115633, 115634, 115635, 115638, 115671, 115675, 115690, 115691, 115695, 115697, 115698, 115780, 115842, 115843, 115847, 115851, 115854, 115856, 115857, 115858, 115861, 115862, 115863, 115868, 115872, 115873, 115874, 115875, 115877, 115880, 115881, 115887, 115890, 115891, 115965, 116111, 116587, 116594, 116684, 115403, 115607 and 115636; and  SERIAL NUMBERS MODEL #T127:  103080, 103081, 103082, 103083, 103084, 103086, 103090, 103091, 103092, 103095, 103096, 103097, 103098, 103100, 103101, 103104, 103105, 103106, 103108, 103110, 103111, 103113, 103115, 103116, 103117, 103119, 103120, 103121, 103123, 103125, 103128, 103129, 103133, 103134, 103135, 103138, 103139, 103141, 103142, 103143, 103144, 103145, 103150, 103151, 103152, 103155, 103156, 103157, 103159, 103160, 103165, 103168, 103171, 103172, 103176, 103177 and 103178; and  SERIAL NUMBERS MODEL #T135:  943611, 943613, 943614, 943615, 943631, 943632, 943633, 943634, 943635, 943636, 943638, 943639, 943641, 943642, 943643, 943644, 943645, 943648, 943651, 943652, 943653, 943663, 943665, 943675, 943677, 943679, 943682, 943685, 943715 and 944338; and  SERIAL NUMBERS MODEL #T165: 111413, 111415, 111416, 111417, 111418, 111419, 111422, 111425, 111432, 111436, 111438, 111440, 111442, 111445, 111447, 111450, 111452, 111466, 111469, 111471, 111472, 111479, 111481, 111482, 111483, 111485, 111499, 111501, 111503, 111504, 111507, 111514, 111515, 111516, 111538, 111540, 111544, 111545, 111546, 111547, 111548, 111549, 111553, 111556, 111557, 111626, 111629, 111630, 111631, 111632, 111633, 111634, 111637, 111638, 111640, 111641, 111643, 111644, 111645, 111646, 111652, 111659, 111666, 111668, 111670, 111672, 111677, 111678, 111679, 111681, 111683, 111690, 111693, 111694, 111695, 111698, 111700, 111701, 111702, 111703, 111704, 111706, 111707, 111708, 111709, 111710, 111711, 111713, 111716, 111717, 111718, 111719, 111722, 111723, 111724, 111726, 111727, 111730, 111731, 111734, 111736, 111737, 111738, 111739, 111747, 111749, 111750, 111752, 111753, 111754, 111756, 111757, 111778, 111802, 111804, 111807, 111810, 111819, 111821, 111838, 111839, 111842, 111843, 111844, 111846, 111847, 111850, 111851, 111852, 111853, 111861, 111863, 111864, 111865, 111867, 111872, 111873, 111874, 111875, 111876, 111878, 111879, 111880, 111881, 111883, 111884, 111887, 111889, 111891, 111893, 111895, 111896, 111897, 111898, 111899, 111900, 111901, 111902, 111903, 111906, 111914, 111918, 111919, 111920, 111921, 111923, 111924, 111926, 111927, 111929, 111931, 111932, 111934, 111935, 111936, 111938, 111940, 111941, 111942, 111943, 111944, 111945, 111947, 111948, 111950, 111951, 111952, 111953, 111955, 111956, 111957, 111964, 111965, 111966, 111969, 111971, 111974, 111977, 111978, 111981, 111984, 111988, 111993, 111996, 111999, 112000, 112001, 112002, 112003, 112004, 112005, 112006, 112007, 112008, 112009, 112012, 112013, 112014, 112016. 112018, 112021, 112022, 112023, 112025, 112028, 112030, 112032, 112033, 112038, 112040, 112049, 112050, 112051, 112053, 112054, 112056, 112057, 112058, 112059, 112061, 112065, 112067, 112068, 112069, 112070, 112071, 112072, 112076, 112077, 112078, 112081, 112083, 112084, 112085, 112089, 112090, 112092, 112093, 112099, 112104, 112107, 112111, 112114, 112115, 112117, 112127, 112131, 112132, 112133, 112135, 112137, 112138, 112139, 112140, 112143, 112144, 112145, 112149, 112150, 112152, 112159, 112161, 112162, 112219, 112237, 112240, 112242, 112244, 112247, 112322, 112341, 112350, 112352, 112353, 112354, 112355, 112357, 112359, 112360, 112364, 112634, 112658, 112751, 112777, 112782, 112785, 112786, 112810, 112875, 112879, 113101, 113150, 113301, 113796, 111421, 111448, 111470, 111477, 111484, 111489, 111500, 111541, 111554, 111555, 111651, 111662, 111669, 111696, 111699, 111714, 111715, 111801, 111805, 111806, 111845, 111870, 111871, 111882, 111885, 111888, 111890, 111894, 111997, 112017, 112036, 112039, 112041, 112048, 112055, 112080, 112086, 112087, 112088, 112096, 112097, 112098, 112102, 112105, 112113, 112129, 112141, 112151, 112160, 112238, 112351, 112473, 112474, 112476 and 112491; and  SERIAL NUMBERS MODEL #T167:  103886, 103888, 103890, 103903, 103907, 103912, 103913, 103915, 103917, 103920, 103925, 103926, 103928, 103930, 104268, 104303, 104356, 104357, 104360, 104419, 104423, 104488, 104490, 104502, 104583, 103905, 103906, 103914, 103916, 103927, 104096, 104269, 104420, 104496 and 104497; and  SERIAL NUMBERS MODEL #T175:  110174, 110177, 110178, 110180, 110181, 110198, 110247, 110318, 110321, 110322, 110323, 110324, 110325, 110326, 110327, 110330, 110331, 110345, 110347, 110351, 110354, 110356, 110357, 110358, 110361, 110362, 110364, 110365, 110368, 110372, 110374, 110375, 110378, 110381, 110387, 110429, 110432, 110433, 110435, 110437, 110441, 110443, 110445, 110446, 110447, 110459, 110461, 110462, 110471, 110473, 110474, 110475, 110479, 110482, 110483, 110487, 110488, 110492, 110497, 110498, 110540, 110543, 110544, 110548, 110554, 110562, 110565, 110566, 110567, 110571, 110572, 110573, 110574, 110576, 110577, 110579, 110580, 110581, 110582, 110583, 110589, 110590, 110591, 110592, 110597, 110602, 110603, 110605, 110606, 110607, 110608, 110609, 110612, 110617, 110618, 110619, 110620, 110621, 110623, 110625, 110626, 110628, 110698, 110699, 110740, 110741, 110744, 110747, 110756, 110793, 110857, 110858, 110859, 110860, 110862, 110865, 110867, 110871, 110877, 111296, 111993, 110245, 110246, 110248, 110296, 110320, 110344, 110348, 110359, 110369, 110373, 110379, 110384, 110430, 110434, 110439, 110444, 110451, 110455, 110456, 110460, 110463, 110478, 110489, 110490, 110493, 110494, 110496, 110511, 110542, 110559, 110594, 110694, 110700, 110702, 110739, 110743, 110745, 110750, 110751, 110753, 110754, 110755, 110758, 110787, 110790, 110795, 110796, 110797, 110854, 110868, 110870, 110881, 110933, 112361, 110290, 110388, 110442 and 110484; and  SERIAL NUMBERS MODEL #T177:  104009, 104012, 104013, 104016, 104019, 104021, 104033, 104039, 103990, 103992, 103993, 103998, 103999, 104000, 104004, 104005, 104007, 104008, 104011, 104015, 104018, 104022, 104027, 104028, 104031, 104034, 104040, 104051 and 104052.

Other devices under the same FDA submission

45 devices cleared under P960040 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates