Skip to content

Synthes Zero-P StarDrive Screwdriver Shaft, T8… recallZ-1348-2009 · Synthes Spine · initiated Feb 10, 2009

Terminated All recalls posted in June 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Synthes Spine or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Tip breakage, due to inadequate instructions for use. The Technique Guide provided incorrect information related to use of the Torque Limiting Attachment during screw insertion and final tightening.

FDA root cause category: Labeling design

What the company is doing

All affected Synthes Spine Sales Consultants were notified of the Labeling Correction via email on February 10, 2009. They were instructed to review changes with their surgeons. They were supplied with an electronic version of the corrected Technique Guide on March 11, 2009. All affected Synthes Sales Consultants and Customers were notified of this labeling change via distribution of the updated Technique Guide and a notification letter. This packet was sent via USPS certified mail on March 26, 2009. To ensure a broad and accurate distribution of notification, affected consultants and customers are defined as having been shipped Zero-P 3.0 mm Titanium Cervical Spine Locking Screws. These devices are used only with the Zero-P system.

Recall details

Recall number
Z-1348-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Feb 10, 2009
Posted by the FDA
Jun 10, 2009
Recalling firm
Synthes Spine
Quantity
1766 units
Product code
ODP
FDA submissions
K072981
Distribution
Distributed Nationwide and Canada.

Product description

Synthes Zero-P StarDrive Screwdriver Shaft, T8, self-retaining, quick coupling, for use in spinal fusion surgery. Catalog # 03.617.902 Technique Guide.

Code information (models, lots, serial numbers)

Catalog # 03.617.902 - All lots

Other devices under the same FDA submission

6 devices cleared under K072981 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page