Treadmill models TM55, TM65, and CR60 (brand names include… recallZ-1757-2010 · Cardiac Science Corporation · initiated Mar 30, 2009
Terminated All recalls posted in June 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardiac Science Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
28 stress and rehab treadmills manufactured between September 2006 and January 2009 may not have received adequate electrical safety testing prior to being shipped from the factory. If these components break down in the field, a patient or operator could be exposed to excess touch currents on the surface of the product. Because safety testing may not have allowed for complete evaluation of the el
FDA root cause category: Other
What the company is doing
On April 3, 2009 the recalling firms notified consignees by certified letter to domestic customers and via FedEx to international customers with delivery confirmation signature required. Customers were advised to remove treadmill from service until it is tested by a Field Service Engineer. For customers with questions or those that have not already scheduled a service call with Cardiac Science Field Service Engineers, please contact the firm at 1.800.426.0337, option 1.
Recall details
- Recall number
- Z-1757-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 30, 2009
- Posted by the FDA
- Jun 4, 2010
- Recalling firm
- Cardiac Science Corporation
- Quantity
- 28 treadmills
- Product code
- IOL
- FDA submissions
- K003576, K082173, K864013, K990866
- Distribution
- Distributed domestically (including Puerto Rico) to 20 end-user accounts and 2 distributor accounts; Distributed internationally to: Canada, Mexico, Portugal, and Saudi Arabia.
Product description
Treadmill models TM55, TM65, and CR60 (brand names include Quinton, Burdick, and Cardiac Science).
Code information (models, lots, serial numbers)
Lot/Serial Numbers: 11040077414 ,11040077801 ,11040077859 ,11040077861 ,11040077969 ,11040077977 ,11040077974 ,11040078366 ,11040078367 ,11040078375 ,11040078376 ,11040078379 ,11040078385 ,11040078384 ,11040078368 ,11040078395 ,11040078452 ,11040078891 ,11040079014 ,11040079161 ,11040079568 ,11040079567 ,11040079569 ,11040079671 ,11040079734 ,11040079939 ,11040079880 ,11040080124
Other devices under the same FDA submission
17 devices cleared under K003576, K082173, K864013, K990866 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Q-Stress QSERV-6AA-XXXXBMortara Instrument, Inc. · Electrocardiograph2 recalls1 report
- Q-Stress QS6-AXTDSMortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Q-Stress QS-6BD-ACAABMortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Q-Stress QS-6ED-AEAABMortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Q-Stress QS-6IA-AEDABMortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Q-Stress QS6-ATTD1Mortara Instrument, Inc. · Electrocardiograph1 recall4 reports
- Q-Stress QS6-ATTDXMortara Instrument, Inc. · Electrocardiograph1 recall3 reports
- Q-Stress QR-6AC-XXXXXMortara Instrument, Inc. · Electrocardiograph1 recall0 reports
- Q-Stress QS-6EB-BXDXAMortara Instrument, Inc. · Electrocardiograph1 recall0 reports
- Q-Stress QSERV-6AA-XXXXAMortara Instrument, Inc. · Electrocardiograph1 recall0 reports
- Q-Stress QS-6GD-XGDAAMortara Instrument, Inc. · Electrocardiograph1 recall0 reports
- Q-Stress QS-6AA-DDCAAMortara Instrument, Inc. · Electrocardiograph1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page