i-STAT G3+ Cartridge recallZ-1541-2009 · Abbott Point of Care Inc. · initiated Feb 23, 2009
Terminated All recalls posted in June 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Point of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Equipment problem: i-STAT cartridges were not properly sealed as the packaging system jammed, causing compromised seals.
FDA root cause category: Packaging process control
What the company is doing
Urgent Recall Notice letters pertaining to the EG7+ Cartridges were sent on February 23, 2009 by first class mail to all customers. The firm found one other affected lot of G3+ cartridges and expanded the recall to include one other lot. Letters for the expansion are dated May 2009. Customers were asked to use the attached form to record all of the box numbers if they have any cartridges from the affected lot. Cartridges from the box numbers listed will be replaced an any remaining unused cartridges from those specific boxes are to be returned as instructed on the form. To receive replacement cartridges, customers are to follow the instructions on the enclosed customer reply form. If any cartridges from the affected lot were forwarded to another facility, a copy of the recall letter needs to be provided to them and the information at the bottom of the customer reply form should be completed. Questions should be directed to 1-800-366-8020, Option 1.
Recall details
- Recall number
- Z-1541-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 23, 2009
- Posted by the FDA
- Jun 18, 2009
- Recalling firm
- Abbott Point of Care Inc.
- Quantity
- 1136 boxes of 25 cartridges each
- Product code
- CHL
- FDA submissions
- K940918
- Distribution
- Worldwide Distribution -- USA, Austria, China, India, Saudi Arabia, South Africa, and Taiwan.
Product description
i-STAT G3+ Cartridge; pH, PCO, PO. HCO2, TCO2, BE sO2; Test includes: PCO2 for calculated HCO3 and PO2 for calculated oxygen saturated/sO2. Catalog Number 220100. Abbott Point of Care Inc., Abbott Park, IL 60064 The iSTAT G3+ Cartridge is intended to be used to quantitatively measure oxygen, carbon dioxide, and pH.
Code information (models, lots, serial numbers)
Lot P08321C, Abbott List number 06F03-01 (USA and Canada), Abbott List Number (rest of world) 06F03-02; Box numbers 1491 and 1534.
Other devices under the same FDA submission
5 devices cleared under K940918 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- i-STAT 03P88-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH105 reports
- i-STAT 03P76-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall46 reports
- i-STAT 03P77-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall10 reports
- i-STAT 03P79-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH23 reports
- i-STAT 03P78-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH16 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 51211).
- Z-1540-2009 · Terminatedi-STAT EG7+ Cartridge; pH, PCO, PO, Na , K, iCa, Hct. Test include: Electrode Measurement Blood-Gases (Pco2, PO2) and Blood Ph, Electrode, Ion Specific, Potassium, Electrode, Ion …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page