Skip to content

Malibu Polyaxial Reduction Screw, 6.5 x 45mm Part Number… recallZ-1926-2009 · SeaSpine, Inc · initiated Mar 4, 2009

Terminated All recalls posted in September 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from SeaSpine, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

SeaSpine has received reports relating to rust on the internal washer contained within the titanium seat assembly of certain Malibu Polyaxial and Malibu Reduction Screws.

FDA root cause category: Mixed-up of materials/components

What the company is doing

In an Urgent Voluntary Product Recall Notification letter dated March 5th, 2009, SeaSpine informed consignees about the reason for recall and asked them to immediately check the products in their possession and hospital inventories to determine if they have any of the affected product. Alternatively, there were told if they would prefer to return to SeaSpine all Malibu Screw inventory that is in their possession, SeaSpine will sort out the affected product upon receipt and will replace their inventory with Malibu Screws that conform to all SeaSpine specifications. The part number and lot are laser marked on the outside of the seat assembly of each screw. SeaSpine asked their consignees to return affected product to them immediately. Attached was a verification form, that they were told to complete and return even if they did not have any product to return. Furthermore their local Sales Manager could assist them in completing the form. They were also encouraged to contact their Sales Manager or Steve Rybka at (760) 727-8399 x202 if they had any questions regarding the recall, any of their products, or would like assistance with the recall.

Recall details

Recall number
Z-1926-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 4, 2009
Posted by the FDA
Sep 3, 2009
Recalling firm
SeaSpine, Inc
Quantity
1,116 Screws in total
Product code
KWP
FDA submissions
K051942
Distribution
Worldwide Distribution -- USA, including states of AL, AZ, CA, CO, FL, GA, IN, MD, MI, MO, MS, NM, OK, TN, TX, VA, and WI and country of Greece.

Product description

Malibu Polyaxial Reduction Screw, 6.5 x 45mm Part Number 13-6545 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System is to provide immobilization in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine.

Code information (models, lots, serial numbers)

Lot Number: A1728A

Other devices under the same FDA submission

8 devices cleared under K051942 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 52115).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page