Malibu 12-6545Recalls & reported problems
Seaspine Orthopedics Corporation · Malibu Polyaxial Screw,6.5 x 45mm · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10889981001155), received between Sep 12, 2019 and Mar 16, 2020 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (4 reports).
11 other recalls cover devices cleared under the same FDA submission (K051942); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Integrity Problem4
- Corroded3
Patient problems
- No Known Impact Or Consequence To Patient3
- Patient Problem/Medical Problem1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 16, 2020Corroded; Material Integrity Problem; Corroded; Material Integrity Problem
ON (B)(6) 2020 SEASPINE WAS INFORMED OF A MALIBU SPINAL SYSTEM REMOVAL SURGERY. THE INDEX SURGERY WAS (B)(6) 2008. THE PRIMARY REASON FOR THE REMOVAL WAS PATHOLOGY AT AN ADJACENT LEVEL AND NOT DUE TO FAILURE OF THE IMPLANT. DURING THE REMOVAL, CORROSION WAS DISCOVERED WHERE THE ROD AND SCREW INTERFACED. THE DISTRIBUTOR EXPRESSED THE SURGEON'S CONCERN REGARDING THE MALIBU SCREW RECALL THAT TOOK PLACE SHORTLY AFTER THE INDEX SURGERY. THE REFERENCED RECALL WAS DUE TO THE MANUFACTURER USING THE WRONG METAL INSIDE THE TULIP HEAD OF THE SCREW.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9835491 ↗MalfunctionMar 16, 2020Corroded; Material Integrity Problem; Corroded; Material Integrity Problem
ON (B)(6) 2020 SEASPINE WAS INFORMED OF A MALIBU SPINAL SYSTEM REMOVAL SURGERY. THE INDEX SURGERY WAS (B)(6) 2008. THE PRIMARY REASON FOR THE REMOVAL WAS PATHOLOGY AT AN ADJACENT LEVEL AND NOT DUE TO FAILURE OF THE IMPLANT. DURING THE REMOVAL, CORROSION WAS DISCOVERED WHERE THE ROD AND SCREW INTERFACED. THE DISTRIBUTOR EXPRESSED THE SURGEON'S CONCERN REGARDING THE MALIBU SCREW RECALL THAT TOOK PLACE SHORTLY AFTER THE INDEX SURGERY. THE REFERENCED RECALL WAS DUE TO THE MANUFACTURER USING THE WRONG METAL INSIDE THE TULIP HEAD OF THE SCREW. PLEASE SEE ADDITIONAL INFORMATION CONCERNING RECALL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9835423 ↗MalfunctionMar 14, 2020Corroded; Material Integrity Problem; Corroded; Material Integrity Problem
ON FEBRUARY 18, 2020, SEASPINE WAS INFORMED OF A MALIBU SPINAL SYSTEM REMOVAL SURGERY. THE INDEX SURGERY WAS (B)(6) 2008. THE PRIMARY REASON FOR THE REMOVAL WAS PATHOLOGY AT AN ADJACENT LEVEL AND NOT DUE TO FAILURE OF THE IMPLANT. DURING THE REMOVAL, CORROSION WAS DISCOVERED WHERE THE ROD AND SCREW INTERFACED. THE DISTRIBUTOR EXPRESSED THE SURGEON'S CONCERN REGARDING THE MALIBU SCREW RECALL THAT TOOK PLACE SHORTLY AFTER THE INDEX SURGERY. THE REFERENCED RECALL WAS DUE TO THE MANUFACTURER USING THE WRONG METAL INSIDE THE TULIP HEAD OF THE SCREW. PLEASE SEE ADDITIONAL INFORMATION CONCERNING RECALL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9832612 ↗MalfunctionSep 12, 2019Material Integrity Problem; Material Integrity Problem
ON (B)(6) 2019 A PATIENT CONTACTED THE MANUFACTURER ALLEGING A POST-OPERATIVE SCREW FAILURE WITH A POSTERIOR FIXATION SYSTEM. INFORMATION PROVIDED FROM THE PATIENT DETERMINED THE INDEX SURGERY DATE ((B)(6) 2010) AND MODEL NUMBER (12-6545 MALIBU POLYAXIAL SCREW). IT IS UNKNOWN HOW THE SCREW FAILURE WAS DISCOVERED. IT IS UNKNOWN IF A REVISION IS SCHEDULED. NO RADIOGRAPHS OR CASE SHEET IS AVAILABLE. PATIENT WAS ADVISED TO CONTACT A PHYSICIAN.
Patient problems: Patient Problem/Medical Problem
MAUDE report 9014504 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051942. One submission often covers many models, so check the product and lot details in each record.
- Z-0938-2014 · Jan 22, 2014 · TerminatedMalibu Polyaxial Screwdriver, 8.0 Plus, Model Number 91-2109; Malibu Polyaxial Screwdriver, Standard, Model Number 91-2110. 510(k): 91-2109 K061342. 91-2110 K051942. The M
- Z-1920-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Screw, 6.5 x 40mm Part Number 12-6540 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System is to pro
- Z-1926-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Reduction Screw, 6.5 x 45mm Part Number 13-6545 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System
- Z-1919-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Screw, 6.5 x 35mm Part Number 12-6535 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System is to pr
- Z-1923-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Screw, 7.5 x 35mm Part Number 12-7535 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System is to pro
- Z-1921-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Screw, 6.5 x 45mm Part Number 12-6545 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System is to pro
- Z-1927-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Reduction Screw, 6.5 x 50mm Part Number 13-6550 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System
- Z-1922-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Screw, 6.5 x 50mm Part Number 12-6550 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System is to pro
- Z-1925-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Reduction Screw, 6.5 x 40mm Part Number 13-6540 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System
- Z-1924-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Screw, 7.5 x 70mm Part Number 12-7570 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System is to pro
- Z-1918-2009 · Mar 4, 2009 · TerminatedMalibu Polyaxial Screw, 4.5 x 30mm Part Number 12-4530 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System is to pr
Other Orthosis, spondylolisthesis spinal fixation devices
- CD HORIZON® Spinal System 7540020Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation257 reports
- MANTIS Redux 48289999Stryker Corporation · Orthosis, spondylolisthesis spinal fixation76 reports
- EXPEDIUM TOP NOTCH 179771555Depuy Spine, LLC · Orthosis, spondylolisthesis spinal fixation41 reports
- EXPEDIUM 179712650Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation31 reports
- EXPEDIUM VERSE 199721530Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4 5 48130000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4.5 System - XIA CT 48850000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation27 reports
- CD HORIZON® Spinal System 7576545Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation26 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.