Skip to content

Roche CoaguChek XS System, CoaguChek XS Plus Care Kit… recallZ-2004-2009 · Roche Diagnostics Operations, inc. · initiated Jul 22, 2009

Terminated All recalls posted in September 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Operations, inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The recommended cleaning procedures for the meter may cause a malfunction and falsely elevated test result reports.

FDA root cause category: Device Design

What the company is doing

Roche Diagnostics Operations, Inc. notified Consignees by letter dated July 22, 2009 informing them of the problem and instructing them to use revised cleaning procedures. For further questions, contact Roche Diagnostics Operations, Inc. at 1-317-521-3194.

Recall details

Recall number
Z-2004-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jul 22, 2009
Posted by the FDA
Sep 3, 2009
Recalling firm
Roche Diagnostics Operations, inc.
Quantity
1,327 in US
Product code
GJS
FDA submissions
K071041
Distribution
Worldwide Distribution

Product description

Roche CoaguChek XS System, CoaguChek XS Plus Care Kit, Roche Diagnostics, Indianapolis, IN; Model Number 05021537001. CoaguChek XS System and CoaguChek XS Plus System are point of care monitoring devices for people taking oral anticoagulation medication.

Code information (models, lots, serial numbers)

All serial numbers.

Other devices under the same FDA submission

1 device cleared under K071041 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 52724).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page