Skip to content

CoaguChek XS Plus Professional Care Kit 04562780001Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630945726), received between Dec 2, 2021 and Oct 17, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Circuit Failure” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K071041); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Circuit Failure1
  • Failure to Transmit Record1
  • Image Display Error/Artifact1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 17, 2023Image Display Error/Artifact; Image Display Error/Artifact

    WE RECEIVED AN ALLEGATION OF A DISPLAY ISSUE WITH COAGUCHEK XS PLUS METER SERIAL NUMBER (B)(6). THE CUSTOMER STATED THE METER DISPLAY LOOKS BLURRY AND SOMETIMES THE DISPLAY IS CHOPPY AND DIFFICULT TO READ. NO MISINTERPRETATION OF RESULTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17949474 ↗
  • MalfunctionApr 21, 2022Failure to Transmit Record; Failure to Transmit Record

    THERE WAS AN ALLEGATION OF AN ISSUE WITH THE SOFTWARE FOR A COAGUCHEK XS PLUS METER. A METER RESULT OF 2.4 INR WAS TAKEN ON (B)(6) 2022 AT 00:03. THE CUSTOMER COULD NOT FIND THE RESULT IN THEIR (RALS) SYSTEM. UPON REVIEW OF THE METER MEMORY, THE RESULT WAS IN THE METER MEMORY AS BEING TAKEN ON (B)(6) 2021. THE RESULT WAS TRANSFERRED TO THE CORRECT PATIENT BUT HAD THE WRONG DATE. THE TEST STRIP LOT NUMBER AND EXPIRATION DATE WERE REQUESTED, BUT NOT PROVIDED. THE PATIENT'S THERAPEUTIC RANGE WAS REQUESTED, BUT NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14170395 ↗
  • MalfunctionDec 2, 2021Circuit Failure; Circuit Failure

    THERE WAS AN ALLEGATION OF A METER MEMORY ISSUE WITH A COAGUCHEK XS PLUS METER. THE CUSTOMER REVIEWED THE METER'S MEMORY AND NOTICED TWO RESULTS FROM TWO PATIENTS ON (B)(6) 2021 WERE MISSING. THE TWO RESULTS WERE 1.9 INR AT 09:29 AND 2.5 INR AT 10:50. THE CUSTOMER CONFIRMED THERE WERE NO MISSING PIXELS FROM THE DISPLAY, THE RESULTS FOUND IN THE METER'S MEMORY WERE ABLE TO BE READ, AND THE DATE WAS SET CORRECTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12915265 ↗

FDA clearance and classification

Cleared through Special 510(k) K071041 on May 11, 2007 (decision: Substantially Equivalent). Submitted by Roche Diagnostics as “Coaguchek Xs Plus System”. The FDA review took 29 days. FDA record ↗

The FDA regulates product code GJS (test, time, prothrombin) under 21 CFR 864.7750 as a Class II device. 21 CFR 864.7750 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071041. One submission often covers many models, so check the product and lot details in each record.

  • Z-0360-2019 · Sep 13, 2018 · Terminated
    CoaguChek XS System Prothrombin time test: CoaguChek XS PT Test 2X24 Strips, Catalog Number: 04625315160; CoaguChek XS PT Test 6 Strips, Catalog Number: 04625374160; CoaguChek XS P
  • Z-1583-2013 · May 16, 2013 · Terminated
    CoaguChek XS Plus System The CoaguChek XS Plus system for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy
  • Z-2004-2009 · Jul 22, 2009 · Terminated
    Roche CoaguChek XS System, CoaguChek XS Plus Care Kit, Roche Diagnostics, Indianapolis, IN; Model Number 05021537001. CoaguChek XS System and CoaguChek XS Plus System are point

Common questions about the CoaguChek XS Plus Professional Care Kit 04562780001

Has the CoaguChek XS Plus Professional Care Kit 04562780001 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04015630945726), based on the FDA recall file dated Nov 27, 2026. However, 3 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the CoaguChek XS Plus Professional Care Kit 04562780001?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Dec 2, 2021 and Oct 17, 2023. In those reports, the most common specific device problems are “Circuit Failure” (1), “Failure to Transmit Record” (1) and “Image Display Error/Artifact” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CoaguChek XS Plus Professional Care Kit 04562780001?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CoaguChek XS Plus Professional Care Kit 04562780001 FDA approved?

It was cleared through Special 510(k) K071041 on May 11, 2007. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the CoaguChek XS Plus Professional Care Kit 04562780001, and is it still sold?

The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand CoaguChek XS Plus Professional Care Kit, model 04562780001). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Test, Time, Prothrombin devices

All Test, Time, Prothrombin devices →