Skip to content

Abbott i-STAT ACT Celite Cartridges recallZ-0813-2010 · Abbott Point of Care Inc. · initiated Aug 18, 2009

Terminated All recalls posted in February 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Labeling discrepancy: The manual instruction state that the line should be flushed with 5 mL of saline and the first 5 mL of blood or six dead space volumes should be discarded. The Cartridge Test Information sheets that it should be 2 mL of blood - which is incorrect.

FDA root cause category: Labeling False and Misleading

What the company is doing

Abbott Point of Care issued an "Important Product Information" letter beginning August 18, 2009. Consignees were instructed on how to reconcile all affected product. For further information, contact Abbott Point of Care Technical Support at 1-800-366-8020, Option 1.

Recall details

Recall number
Z-0813-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 18, 2009
Posted by the FDA
Feb 23, 2010
Recalling firm
Abbott Point of Care Inc.
Quantity
696,675 Cartridges
Product code
JBP
FDA submissions
K992571
Distribution
Worldwide distribution.

Product description

Abbott i-STAT ACT Celite Cartridges. In vitro diagnostic test used to monitor moderate and high-level heparin therapy through analysis of arterial and venous whole blood samples.

Code information (models, lots, serial numbers)

Catalog number: 07G01-1

Other devices under the same FDA submission

1 device cleared under K992571 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 52754).