Medical device recalls: February 2010
The FDA posted 90 medical device recall records in February 2010, from 41 recall events by 32 companies. 90 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Stryker Howmedica Osteonics Corp. · 5 events, 15 records
- Abbott Point of Care Inc. · 3 events, 4 records
- Stryker Spine · 2 events, 5 records
- Remel, Inc · 2 events, 3 records
- Siemens Healthcare Diagnostics, Inc. · 2 events, 2 records
- Becton Dickinson Medical Systems · 1 event, 19 records
- Merit Medical Systems, Inc. · 1 event, 6 records
- Cybernet Systems Corp. · 1 event, 4 records
FDA root cause
- Process control · 30
- Other · 9
- Component change control · 8
- Packaging · 6
- Nonconforming Material/Component · 6
- Component design/selection · 6
- Package design/selection · 4
- Software design · 3
Not yet classified (90)
The FDA has posted these records but has not published a class for them yet.
- Z-0861-2010 · Bard Access Systems · posted Feb 26, 2010 · Terminated
After activation of the safety mechanism, the tip of the needle protrudes from the safety sheath creating the potential for needlestick injury.
- Z-0862-2010 · Bard Access Systems · posted Feb 26, 2010 · Terminated
After activation of the safety mechanism, the tip of the needle protrudes from the safety sheath creating the potential for needlestick injury.
- Z-0864-2010 · Siemens Healthcare Diagnostics, Inc. · posted Feb 26, 2010 · Terminated
Siemens Healthcare Diagnostics, Inc. has confirmed high accuracy shifts with Dimension Vista URCA Flex reagent cartridge (various lots) when calibrating with some vials from CHEM 1 calibrator lot 9DM001.
- Z-0865-2010 · Abbott Point of Care Inc. · posted Feb 26, 2010 · Terminated
Abbott Point of Care has determined that it is possible under certain circumstances that some test records could have their status update messages mis-matched with another record, resulting in the incorrect patient name, order number of comment being…
- Z-0866-2010 · Maquet Cardiovascular, LLC · posted Feb 26, 2010 · Terminated
Labeling mix up: Shelf Carton label for Exxcel Soft Thin Wall Vascular Grafts, lot # 12633379 were mislabeled as Exxcel Soft Standard Wall Grafts. Bleeding could occur if wrong product is used.
- Z-0867-2010 · Bio-Rad Laboratories Inc · posted Feb 26, 2010 · Terminated
Product update kit CD rom will not upload. As a result, the device cannot be run.
- Z-0868-2010 · Medtronic Inc · posted Feb 26, 2010 · Terminated
Medtronic has identified a small number of Autotransfusion One Source Kits and MAXPAC/Total System lots with incorrect Use Before Dates (UBD) on the outer kit package label. The UBD on the outer label of the kit/pack does not accurately reflect the UBD of…
- Z-0869-2010 · Medtronic Inc · posted Feb 26, 2010 · Terminated
Medtronic has identified a small number of Autotransfusion One Source Kits and MAXPAC/Total System lots with incorrect Use Before Dates (UBD) on the outer kit package label. The UBD on the outer label of the kit/pack does not accurately reflect the UBD of…
- Z-0870-2010 · Remel, Inc · posted Feb 26, 2010 · Terminated
There is an error in Version 1.0.77 of the electronic code compendium (ERIC) software. The error is limited to the database corresponding to the R8311007 RapID Yeast Plus System. The database erroneously includes Candida dubliniensis as well as displays…
- Z-0850-2010 · Beckman Coulter Inc · posted Feb 25, 2010 · Terminated
Beckman Coulter has become aware of a potential issue in which the Sample ID information embedded in the barcode affixed to a slide made by the Coulter GEN*S and LH SlideMaker does not match the text.
- Z-0851-2010 · Beckman Coulter Inc · posted Feb 25, 2010 · Terminated
Beckman Coulter has become aware of a potential issue in which the Sample ID information embedded in the barcode affixed to a slide made by the Coulter GEN*S and LH SlideMaker does not match the text.
- Z-0852-2010 · Cybernet Systems Corp. · posted Feb 25, 2010 · Terminated
A software problem renders these devices incapable of capturing lung function, oxygen saturation and pulse data collected from multiple tests in accordance with the instructions for use.
- Z-0853-2010 · Cybernet Systems Corp. · posted Feb 25, 2010 · Terminated
A software problem renders these devices incapable of capturing lung function, oxygen saturation and pulse data collected from multiple tests in accordance with the instructions for use.
- Z-0854-2010 · Cybernet Systems Corp. · posted Feb 25, 2010 · Terminated
A software problem renders these devices incapable of capturing lung function, oxygen saturation and pulse data collected from multiple tests in accordance with the instructions for use.
- Z-0855-2010 · Cybernet Systems Corp. · posted Feb 25, 2010 · Terminated
A software problem renders these devices incapable of capturing lung function, oxygen saturation and pulse data collected from multiple tests in accordance with the instructions for use.
- Z-0856-2010 · Beckman Coulter Inc. · posted Feb 25, 2010 · Terminated
This recall was initiated because of software issue was identified in System Software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red warning event and stopping sample…
- Z-0857-2010 · CareFusion 203, Inc. · posted Feb 25, 2010 · Terminated
CareFusion is recalling the EnVe Ventilator because the EnVe Ventilator does not fully meet applicable standards for immunity to electro-Magnetic Interference (EMI) when tested in the laboratory environment. CareFusion has identified certain device…
- Z-0858-2010 · Siemens Healthcare Diagnostics, Inc. · posted Feb 25, 2010 · Terminated
Tests exhibit falsely elevated results.
- Z-0859-2010 · Pointe Scientific, Inc. · posted Feb 25, 2010 · Terminated
The firm initiated the recall due to the presence of visible contamination of the product.
- Z-0860-2010 · Pointe Scientific, Inc. · posted Feb 25, 2010 · Terminated
The firm initiated the recall due to the presence of visible contamination of the product.
- Z-0812-2010 · Abbott Point of Care Inc. · posted Feb 24, 2010 · Terminated
The storage temperature on the label on the side of the i-Stat Portable Clinical Analyzer shipping box does not match product specification.
- Z-0822-2010 · Diagnostic Hybrids Inc · posted Feb 24, 2010 · Terminated
During QC post release monitoring, the firm determined that the monolayer culture of rabbit kidney cells previously distributed by the firm, appeared to contain another type of cells (monkey kidney cells).
- Z-0823-2010 · Stryker Spine · posted Feb 24, 2010 · Terminated
The Xia 3 8.5mm Polyaxial screws, lot A91109 were anodized turquoise instead of Fuchsia. The turquoise color is reserved for the 9.5 mm Xia 3 screws.
- Z-0825-2010 · Synthes USA (HQ), Inc. · posted Feb 24, 2010 · Terminated
There is the potential for the bender/cutter attachment to continue heating after release of the power button. There is potential for patient and/or staff injury if continuous heating is unrecognized.
- Z-0826-2010 · Encore Medical, Lp · posted Feb 24, 2010 · Terminated
Partial lot of product was not manufactured to specification. Central locking screw was missing threads. There is a potential for dissociation because the screw is not able to be used as a secondary fixation method.
- Z-0827-2010 · Gibson Laboratories Inc · posted Feb 24, 2010 · Terminated
Discrepancies between the product package label and technical insert are conflicting and may be misleading to user. The technical insert included the statement for In Vitro Diagnostic Use. The product insert should have indicated for Research Use Only…
- Z-0828-2010 · Gibson Laboratories Inc · posted Feb 24, 2010 · Terminated
Discrepancies between the product package label and technical insert are conflicting and may be misleading to user. The technical insert included the statement for In Vitro Diagnostic Use. The product insert should have indicated for Research Use Only…
- Z-0841-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 24, 2010 · Terminated
Stryker Orthopaedics became aware that there is the potential for damage to the sterile packaging of the product during transport.
- Z-0842-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 24, 2010 · Terminated
Stryker Orthopaedics became aware that there is the potential for damage to the sterile packaging of the product during transport.
- Z-0843-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 24, 2010 · Terminated
Stryker Orthopaedics became aware that there is the potential for damage to the sterile packaging of the product during transport.
- Z-0844-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 24, 2010 · Terminated
Stryker Orthopaedics became aware that there is the potential for damage to the sterile packaging of the product during transport.
- Z-0845-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 24, 2010 · Terminated
Stryker Orthopaedics became aware that there is the potential for damage to the sterile packaging of the product during transport.
- Z-0846-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 24, 2010 · Terminated
Stryker Orthopaedics became aware that there is the potential for damage to the sterile packaging of the product during transport.
- Z-0813-2010 · Abbott Point of Care Inc. · posted Feb 23, 2010 · Terminated
Labeling discrepancy: The manual instruction state that the line should be flushed with 5 mL of saline and the first 5 mL of blood or six dead space volumes should be discarded. The Cartridge Test Information sheets that it should be 2 mL of blood - which…
- Z-0814-2010 · Abbott Point of Care Inc. · posted Feb 23, 2010 · Terminated
Labeling discrepancy: The manual instruction state that the line should be flushed with 5 mL of saline and the first 5 mL of blood or six dead space volumes should be discarded. The Cartridge Test Information sheets that it should be 2 mL of blood - which…
- Z-0817-2010 · Remel, Inc · posted Feb 23, 2010 · Terminated
The products may give false positive results with some non-Staphylococcus aureus strains
- Z-0818-2010 · Remel, Inc · posted Feb 23, 2010 · Terminated
The products may give false positive results with some non-Staphylococcus aureus strains
- Z-0819-2010 · Veridex, LLC · posted Feb 23, 2010 · Terminated
Veridex has received 4 complaints regarding reagent bottles having been incorrectly placed or duplicate bottles places within the reagent tray of CellSearch Circulating Tumor Cell Kit. Delayed detection of disease relapse or progression may occur due to…
- Z-0798-2010 · Neurovision Medical Products Inc · posted Feb 22, 2010 · Terminated
This recall has been initiated due to a conductive glue used to attach the lead wire to electrode, which has been found in a small number of electrodes, to cause a short circuit in one of the channels. This is due to migration of the adhesive from one pad…
- Z-0801-2010 · Merit Medical Systems, Inc. · posted Feb 22, 2010 · Terminated
Degradation of the outer portion of the delivery catheter may prevent proper deployment of the biliary stent.
- Z-0802-2010 · Merit Medical Systems, Inc. · posted Feb 22, 2010 · Terminated
Degradation of the outer portion of the delivery catheter may prevent proper deployment of the biliary stent.
- Z-0803-2010 · Merit Medical Systems, Inc. · posted Feb 22, 2010 · Terminated
Degradation of the outer portion of the delivery catheter may prevent proper deployment of the biliary stent.
- Z-0804-2010 · Merit Medical Systems, Inc. · posted Feb 22, 2010 · Terminated
Degradation of the outer portion of the delivery catheter may prevent proper deployment of the biliary stent.
- Z-0805-2010 · Merit Medical Systems, Inc. · posted Feb 22, 2010 · Terminated
Degradation of the outer portion of the delivery catheter may prevent proper deployment of the biliary stent.
- Z-0806-2010 · Merit Medical Systems, Inc. · posted Feb 22, 2010 · Terminated
Degradation of the outer portion of the delivery catheter may prevent proper deployment of the biliary stent.
- Z-0808-2010 · Hospira Inc. · posted Feb 22, 2010 · Terminated
There is a potential for fluid to leak from the backcheck valve which could result in inaccurate delivery and occupational exposure.
- Z-0809-2010 · Hospira Inc. · posted Feb 22, 2010 · Terminated
There is a potential for fluid to leak from the backcheck valve which could result in inaccurate delivery and occupational exposure.
- Z-0810-2010 · Biomet 3i, LLC · posted Feb 22, 2010 · Terminated
Package labeled as sterile may have an incomplete or partial seal thereby possibly compromising sterility.
- Z-0811-2010 · Physio Control, Inc. · posted Feb 22, 2010 · Terminated
Affected devices contain a Biphasic PCBA that may have a solder defect on the H Bridge component that could prevent the device from providing a defibrillation shock. There could be a delay in therapy or prevention of defibrillation therapy if the…
- Z-0780-2010 · Philips Medical Systems (Cleveland) Inc · posted Feb 19, 2010 · Terminated
Polymeric RF connector block on the tabletop overheated and eventually ignited during scanning due to combination of arcing of electrical components associated with the RF trap assembly located below the connector and flame rating of the connector block…
- Z-0795-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 19, 2010 · Terminated
Stryker Orthopaedics became aware that a box packaged and labeled as a Scorpio PS Femoral Waffle (No Post) with LFIT, may actually contain a Scorpio Femoral Waffle (With Post) and LFIT.
- Z-0796-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 19, 2010 · Terminated
The Depth Measuring Gauges have 5 scale marking instead of the required 4 scale markings.
- Z-0793-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 17, 2010 · Terminated
Stryker Orthopaedics learned that the T-handles have the potential to not engage and not attach to the reamer properly.
- Z-0794-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 17, 2010 · Terminated
Stryker Orthopaedics learned that the T-handles have the potential to not engage and not attach to the reamer properly.
- Z-0608-2010 · Solafeet, Inc. · posted Feb 16, 2010 · Terminated
Product failed to have a label that contained a recommended exposure schedule; the operator's manual provided inadequate instructions for use; and label implication may result in overexposure.
- Z-0789-2010 · Stryker Spine · posted Feb 16, 2010 · Terminated
Stryker Spine received reports of packaging failures during shipping for the Jam Shidi instruments, which may compromise the sterility of the product.
- Z-0790-2010 · Stryker Spine · posted Feb 16, 2010 · Terminated
Stryker Spine received reports of packaging failures during shipping for the Jam Shidi instruments, which may compromise the sterility of the product.
- Z-0791-2010 · Stryker Spine · posted Feb 16, 2010 · Terminated
Stryker Spine received reports of packaging failures during shipping for the Jam Shidi instruments, which may compromise the sterility of the product.
- Z-0792-2010 · Stryker Spine · posted Feb 16, 2010 · Terminated
Stryker Spine received reports of packaging failures during shipping for the Jam Shidi instruments, which may compromise the sterility of the product.
- Z-0781-2010 · Cincinnati Sub-Zero Products Inc · posted Feb 4, 2010 · Terminated
Product is labeled as "Sterile". It was observed that incomplete seals were created in the packaging materials during production. When the seal is compromised, the sterility of the product inside the package cannot be assured.
- Z-0782-2010 · Cincinnati Sub-Zero Products Inc · posted Feb 4, 2010 · Terminated
Product is labeled as "Sterile". It was observed that incomplete seals were created in the packaging materials during production. When the seal is compromised, the sterility of the product inside the package cannot be assured.
- Z-0661-2010 · Thomas Medical Products Inc · posted Feb 2, 2010 · Terminated
Radiopaque sheath tip may fracture
- Z-0668-2010 · Ge Healthcare · posted Feb 2, 2010 · Terminated
Failure to properly document the CTDI in the Technical Reference or User Manual.
- Z-0669-2010 · Ge Healthcare · posted Feb 2, 2010 · Terminated
Failure to properly document the CTDI in the Technical Reference or User Manual.
- Z-0750-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0751-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0752-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0753-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0754-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0755-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0756-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0764-2010 · Argon Medical Devices, Inc · posted Feb 2, 2010 · Terminated
One lot of Percutaneous Catheter Introducer Set contained incorrect size needle.
- Z-0779-2010 · Biomet, Inc. · posted Feb 2, 2010 · Terminated
The screw may not have the hex dimension inside of the head of the screw, which will prevent the device from being implantable. This may cause a delay in the procedure while another screw is obtained.
- Z-0829-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0830-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0831-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0832-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0833-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0834-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0835-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0836-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0837-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0838-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0839-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0840-2010 · Becton Dickinson Medical Systems · posted Feb 2, 2010 · Terminated
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Z-0757-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 1, 2010 · Terminated
Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
- Z-0758-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 1, 2010 · Terminated
Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
- Z-0759-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 1, 2010 · Terminated
Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
- Z-0760-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 1, 2010 · Terminated
Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
- Z-0761-2010 · Stryker Howmedica Osteonics Corp. · posted Feb 1, 2010 · Terminated
Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.