Why it was recalled

Leakage-- Due to a manufacturing error, some products may have a fold in the innermost layer, causing disruption in the layering process and may result in blow holds or exposed wire braids, which may result in further damage.

FDA root cause category: Process control

What the company is doing

Penumbra, Inc. initiated Recall activities on May 29, 2009. The Firm notified consignees via telephone and email of the problem and asked them to return the product to Penumbra. For further information, contact Penumbra, Inc. at 1-510-748-3223.

Recall details

Recall number
Z-2049-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jul 2, 2009
Posted by the FDA
Sep 8, 2009
Recalling firm
Penumbra Inc.
Quantity
9 units
Product code
DQY
FDA submissions
K083125
Distribution
Worldwide Distribution -- US (IA, NJ, NY and VA), and Canada.

Product description

Penumbra Neuron 5F Select Catheter, Neuro Deliver/Intracranial Access System Catheter, manufactured by Penumbra, Inc., Alameda, CA. Intended for the introduction of interventional devices into the peripheral, coronary and neuro vasculature.

Code information (models, lots, serial numbers)

Catalog Number: PNS5f130H1; Lot Number: F15174.