Dimension Vista System StreamLAB Analytical Workcell.… recallZ-1205-2010 · Siemens Healthcare Diagnostics, Inc. · initiated Sep 1, 2009

Terminated All recalls posted in March 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect result: An intermittent low result may be seen with the last test run from an aliquot on a Dimension Vista System when connected to a StreamLAB Analytical Workcell. The sample delivery sequence may deliver insufficient surplus volume to the low sensitivity aliquot (second aliquot). The surplus sample volume is occasionally insufficient to and can cause short sampling on the last test.

FDA root cause category: Software design

What the company is doing

Siemens healthcare Diagnostics issued an Urgent Field Safety Notice letter to each customer regarding the issue dated September 2009. The letter informed the user of the problem and volume change instructions. Customers can call Technical Solutions Center at 800-441-9250

Recall details

Recall number
Z-1205-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 1, 2009
Posted by the FDA
Mar 24, 2010
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
81 units
Product code
CDT
FDA submissions
K043546, K051087
Distribution
Worldwide distribution: USA, Australia, Canada, France, Germany, Malaysia, Netherlands, Spain, and Switzerland.

Product description

Dimension Vista System StreamLAB Analytical Workcell. Dimension Vista System Part Numbers 1000035896, 1000035899, 1000036563, and 1000036567. StreamLAB Analytical Workcell Part Number 746002.901, for in vitro diagnostic use.

Code information (models, lots, serial numbers)

Dimension Vista System Part Numbers 1000035896, 1000035899, 1000036563, and 1000036567. StreamLAB Analytical Workcell Part Number 746002.901 ***All StreamLAB Analytical Workcells connected to Dimension Vista Systems.

Other devices under the same FDA submission

4 devices cleared under K043546, K051087 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.