Penumbra Neuron Delivery Catheter 070, percutaneous… recallZ-0001-2010 · Penumbra Inc. · initiated Sep 22, 2009

Terminated All recalls posted in October 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Penumbra Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Failure-- product can kink or ovalize, which may result in difficulty in catheter advancement or delivery of other devices and damage to catheter, which could lead to patient injury or death as a result of portions being retained in the patient.

FDA root cause category: Device Design

What the company is doing

The firm issued an Urgent Voluntary Field Removal Notice letter on 09/22/2009 to its consignees, asking them to inspect their stocks; remove affected product and place in quarantine for return to Penumbra Inc.; and complete and return the attached product identification / return form.

Recall details

Recall number
Z-0001-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 22, 2009
Posted by the FDA
Oct 5, 2009
Recalling firm
Penumbra Inc.
Quantity
2246 units
Product code
DQY
FDA submissions
K082290
Distribution
Worldwide Distribution -- USA, Canada, Hong Kong, Chile, the UK, Denmark, Norway, Sweden, Finland, Netherlands, Germany, Switzerland, Austria, and Spain.

Product description

Penumbra Neuron Delivery Catheter 070, percutaneous catheter, catalog numbers PND6F0701058, PND6F070956, PND6F0701058M, and PND6F070956M; manufactured by Penumbra, Inc., Alameda, CA. Indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

Code information (models, lots, serial numbers)

Lot numbers F13733, F13751, F13773, F13799, F13817, F13826, F13856, F13865, F13880, F13899, F13934, F13960, F13970, F13971, F13980, F13997, F14007, F14028, F14045, F14046, F14047, F14048, F14102, F14103, F14150, F14182, F14183, F14427, F14482, F14560, and F14629.

Other devices under the same FDA submission

9 devices cleared under K082290 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.