IMPACT Catheters, Balloon Dilatation and Esophageal… recallZ-0509-2010 · Numed Inc · initiated Sep 23, 2009

Terminated All recalls posted in December 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Numed Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm received a complaint from the sole direct account that while performing an inventory review it was observed that the packaging of the IMPACT catheter was cracked and deteriorating. The product was still in its packaging and never used. The customer referred to it as looking like "paint crackle." The risk is that for continued sterility to be assured, the packaging must remain intact.

FDA root cause category: Labeling design

What the company is doing

An informal e-mail was sent to B. Braun to give them notice not to distribute any Impact devices until further notice. This was followed up by a formal URGENT MEDICAL DEVICE RECALL letter (dated 9/24/09) to the sole consignee. The products are to be returned to the manufacturer, including any products further distributed. Customer questions should be directed to the recall contact (replace " AT " with @) nlaflesh ATnumedusa.com

Recall details

Recall number
Z-0509-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 23, 2009
Posted by the FDA
Dec 15, 2009
Recalling firm
Numed Inc
Quantity
7,367 units domestically
Product code
LIT
FDA submissions
K931009, K983373
Distribution
One distributor in Pennsylvania

Product description

IMPACT Catheters, Balloon Dilatation and Esophageal Dilatation Catheter, Model Number 307, STERILE, Rx only --- Common/Usual Name: PTA and Esophageal Catheter --- Manufactured for: B. Braun Interventional Systems Inc., 824 Twelfth Avenue Bethlehem, PA18018 --- Manufactured by: NuMED, Inc. 2880 Main Street, Hopkinton, NY 12965 --- The Impact catheter is packaged in a straight sheath and then heat sealed in two Tyvek pouches. The inner pouch is the one that is deteriorating when exposed to the sunlight. The outer pouch retains its seal so the sterility is not compromised. --- 510 K931009 - Device Listing #D003206 and 510 K983373 - Device Listing # D003212, Class II device.

Code information (models, lots, serial numbers)

Show the full code information (9k characters)

All Impact catheters distributed between 10/01/2004 and 9/23/2009 are affected: Lot numbers IM-0843, IM-0893, IM-0894, IM-0895, IM-0896, IM-0897,  IM-0898, IM-0899, IM-0900, IM-0901, IM-0902, IM-0903, IM-0904, IM-0905,  IM-0906, IM-0907, IM-0908, IM-0909, IM-0910, IM-0914, IM-0775, IM-0788,  IM-0807, IM-0817, IM-0823, IM-0827, IM-0843, IM-0845, IM-0846, IM-0852,  IM-0857, IM-0868, IM-0869, IM-0890, IM-0891, IM-0912, IM-0915, IM-0916,  IM-0917, IM-0918, IM-0919, IM-0920, IM-0921, IM-0922, IM-0923, IM-0924, IM-0925, IM-0926, IM-0927, IM-0943, IM-0944, IM-0945, IM-0946, IM-0644, IM-0793, IM-0814, IM-0849, IM-0884, IM-0928, IM-0929, IM-0930, IM-0931, IM-0932, IM-0933, IM-0934, IM-0935, IM-0936, IM-0937, IM-0938, IM-0939, IM-0940, IM-0941, IM-0942, IM-0644, IM-0854, IM-0855, IM-0866, IM-0878, IM-0889, IM-0905, IM-0924, IM-0947, IM-0948, IM-0949, IM-0950, IM-0951, IM-0952, IM-0953, IM-0954, IM-0955, IM-0956, IM-0957, IM-0958, IM-0959, IM-0960, IM-0961, IM-0962, IM-0963, IM-0964, IM-0644, IM-0831, IM-0851,  IM-0864, IM-0876, IM-0887, IM-0902, IM-0933, IM-0935, IM-0937, IM-0965, IM-0966, IM-0967, IM-0968, IM-0969, IM-0970, IM-0971, IM-0829, IM-0895,  IM-0897, IM-0898, IM-0917, IM-0918, IM-0931, IM-0949, IM-0972, IM-0973, IM-0974, IM-0975, IM-0976, IM-0977, IM-0978, IM-0986, IM-0979, IM-0980, IM-0981, IM-0982, IM-0983, IM-0984, IM-0985, IM-0882, IM-0915, IM-0928, IM-0945, IM-0951, IM-0954, IM-0967, IM-1002, IM-1003, IM-0987, IM-0988, IM-0989, IM-0990, IM-0991, IM-0992, IM-0993, IM-0994, IM-0995, IM-0996,  IM-0997, IM-0998, IM-0999, IM-1000, IM-1001, IM-0913, IM-0941, IM-1004, IM-1005, IM-1006, IM-1007, IM-1008, IM-1009, IM-1010, IM-1011, IM-1012, IM-1013, IM-1014, IM-1015, IM-1016, IM-0926, IM-0972, IM-1017, IM-1018, IM-0963, IM-0965, IM-1019, IM-0861, IM-0896, IM-0987, IM-1004, IM-1020, IM-1021, IM-1022, IM-1023, IM-1024, IM-1025, IM-1026, IM-1029, IM-1027, IM-1028, IM-1030, IM-1031, IM-1032, IM-1033, IM-1034, IM-1035, IM-0850, IM-0877, IM-0883, IM-0896, IM-0929, IM-0952, IM-0959, IM-0969, IM-0976, IM-0980, IM-1037, IM-1038, IM-1039, IM-1040, IM-1041, IM-1042, IM-1043, IM-1044, IM-0777, IM-0790, IM-0892, IM-0909, IM-0942, IM-0988, IM-1019, IM-1021, IM-0885, IM-0907, IM-0940, IM-0957, IM-0971, IM-0977, IM-0993, IM-1028, IM-1033, IM-1054, IM-1055, IM-1056, IM-1057, IM-1058, IM-0772, IM-0826, IM-0832, IM-0865, IM-0867, IM-0871, IM-0894, IM-0911, IM-0923, IM-0939, IM-0948, IM-0970, IM-0975, IM-0991, IM-1024, IM-1059, IM-1060, IM-1061, IM-1062, IM-1063, IM-1064, IM-1065, IM-1066, IM-0899, IM-0943, IM-1045, IM-0824, IM-0955, IM-0989, IM-1026, IM-1048, IM-1050, IM-1052, IM-1049, IM-1046, IM-1047, IM-1051, IM-1053, IM-0862, IM-0875, IM-0908, IM-0914, IM-0927, IM-0944, IM-0946, IM-0973, IM-1000, IM-1013, IM-1017, IM-1020, IM-1023, IM-1031, IM-1037, IM-1045, IM-1064, IM-1065, IM-1067,  IM-1068, IM-1069, IM-1070, IM-1071, IM-1072, IM-1073, IM-1074, IM-1075, IM-0982, IM-1020, IM-1043, IM-1060, IM-1066, IM-1076, IM-1077, IM-1078,  IM-1079, IM-1080, IM-1081, IM-1082, IM-1083, IM-1084, IM-1085, IM-0919,  IM-0994, IM-1029, IM-1040, IM-1042, IM-1051, IM-1074, IM-1086, IM-1087, IM-1088, IM-1089, IM-1090, IM-1091, IM-1092, IM-1093, IM-1094, IM-1095,  IM-1096, IM-1097, IM-1032, IM-1038, IM-1039, IM-1048, IM-1056, IM-1098, IM-1099, IM-1100, IM-1101, IM-1102, IM-1103, IM-1104, IM-1105, IM-1106, IM-1107, IM-1108, IM-0784, IM-1049, IM-1062, IM-1071, IM-1081, IM-1109, IM-1110, IM-1111, IM-1113, IM-1114, IM-1115, IM-1116, IM-1117, IM-1112, IM-1050, IM-1069, IM-1079, IM-1094, IM-1118, IM-1119, IM-1120, IM-1121, IM-1122, IM-1123, IM-1124, IM-1125, IM-1126, IM-1002, IM-1006, IM-1022, IM-1054, IM-1067, IM-1082, IM-1087, IM-1089, IM-1099, IM-1100, IM-1109, IM-1137, IM-1127, IM-1128, IM-1129, IM-1131, IM-1132, IM-1133, IM-1134,  IM-1135, IM-1136, IM-0932, IM-1046, IM-1092, IM-1100, IM-1110, IM-1138, IM-1139, IM-1140, IM-1141, IM-1142, IM-1143, IM-1144, IM-1145, IM-1146, IM-0932, IM-1096, IM-1101, IM-1105, IM-1107, IM-1113, IM-1120, IM-1124, IM-1127, IM-1147, IM-1148, IM-1149, IM-1150, IM-1151, IM-1152, IM-1153,  IM-1154, IM-1155, IM-1156, IM-1157, IM-0886, IM-0903, IM-0936, IM-0985, IM-1044, IM-1112, IM-1116, IM-1126, IM-1159, IM-1158, IM-1160, IM-1161, IM-1162, IM-1163, IM-1164, IM-1068, IM-1090, IM-1025, IM-1149, IM-1161,  IM-1165, IM-1166, IM-1167, IM-1168, IM-1169, IM-1170, IM-1171, IM-1172, IM-0968, IM-1009, IM-1080, IM-1159, IM-1174, IM-0981, IM-1102, IM-1132, IM-1142, IM-1155, IM-1173, IM-1175, IM-1176, IM-1177, IM-1178, IM-1179, IM-1180, IM-1186, IM-0958, IM-0960, IM-0983, IM-0999, IM-1063, IM-1111, IM-1153, IM-1160, IM-1168, IM-1179, IM-1181, IM-1182, IM-1183, IM-1184, IM-1185, IM-1187, IM-1188, IM-1189, IM-1190, IM-1191, IM-1192, IM-1193, IM-1194, IM-1195, IM-1196, IM-1197, IM-1198, IM-1199, IM-1200, IM-1201, IM-1156, IM-1164, IM-1174, IM-1182, IM-1202, IM-1203, IM-1204, IM-1205, IM-1206, IM-1207, IM-1208, IM-1063, IM-1070, IM-1121, IM-1123, IM-1151, IM-1162, IM-1186, IM-1188, IM-1209, IM-1210, IM-1211, IM-1212, IM-1213, IM-1115, IM-1141, IM-1171, IM-1199, IM-1205, IM-1206, IM-1218, IM-1016, IM-1157, IM-1171, IM-1172, IM-1192, IM-1213, IM-1222, IM-1214, IM-1215, IM-1216, IM-1217, IM-1219, IM-1220, IM-1221, IM-1230, IM-1231, IM-1018, IM-1167, IM-1169, IM-1184, IM-1198, IM-1218, IM-1232, IM-1233, IM-1223, IM-1224, IM-1225, IM-1226, IM-1227, IM-1228, IM-1229, IM-1246, IM-1234, IM-1235, IM-1236, IM-1238, IM-1086, IM-1098, IM-1195, IM-1204, IM-1214,  IM-1217, IM-1219, IM-1248, IM-1237, IM-1239, IM-1240, IM-1241, IM-1242, IM-1243, IM-1244, IM-1245, IM-1247, IM-1057, IM-1178, IM-1196, IM-1216, IM-1223, IM-1233, IM-1249, IM-1250, IM-1251, IM-1252, IM-1253, IM-1254, IM-1255, IM-1256, IM-1114, IM-1143, IM-1200, IM-1208, IM-1228, IM-1229, IM-1244, IM-1254, IM-1255, IM-1257, IM-1258, IM-1259, IM-1260, IM-1261, IM-1262, IM-1263, IM-1264, IM-1264, IM-1266, IM-1267, IM-1268, IM-1269, IM-1270, IM-1271, IM-1272, IM-1190, IM-1252, IM-1273, IM-1279, IM-1061, IM-1122, IM-1211, IM-1227, IM-1236, IM-1238, IM-1242, IM-1297, IM-1298, IM-1299, IM-1289, IM-1274, IM-1275, IM-1276, IM-1277, IM-1278, IM-1280, IM-1281, IM-1282, IM-1283, IM-1284, IM-1285, IM-1286, IM-1287, IM-1288, IM-1290, IM-1291, IM-1292, IM-1293, IM-1294, IM-1295, IM-1296, IM-1206, IM-1226, IM-1272, IM-1299, IM-1300, IM-1301, IM-1302, IM-1303, IM-1304, IM-1305, IM-1306, IM-1307, IM-1308, IM-1309, IM-1310, IM-1311, IM-1318, IM-1312, IM-1313, IM-1314, IM-1315, IM-1316, IM-1317, IM-1332, IM-1333, IM-1334, IM-1335, IM-1117, IM-1133, IM-1136, IM-1146, IM-1163, IM-1170, IM-1201, IM-1202, IM-1217, IM-1219, IM-1222, IM-1225, IM-1232, IM-1241, IM-1253, IM-1260, IM-1262, IM-1271, IM-1276, IM-1279, IM-1286, IM-1289,  IM-1293, IM-1319, IM-1320, IM-1321, IM-1322, IM-1323, IM-1324, IM-1325, IM-1326, IM-1327, IM-1224, IM-1250, IM-1328, IM-1329, IM-1330, IM-1331, IM-1336, IM-1337, IM-1338, IM-1339, IM-1340, IM-1341, IM-1342, IM-1343, IM-1344, IM-1345, IM-1346, IM-1347, IM-1348, IM-1349, IM-1232, IM-1235, IM-1311, IM-1318, IM-1322, IM-1350, IM-1351, IM-1352, IM-1353, IM-1354, IM-1355, IM-1356, IM-1357, IM-1358, IM-1360, IM-1361, IM-1362, IM-1359, IM-1363, IM1364, IM-1158, IM-1165, IM-1215, IM-1240, IM-1277, IM-1283, IM-1365, IM-1366, IM-1367, IM-1281, IM-1299, IM-1307, IM-1323, IM-1328, IM-1331, IM-1335, IM-1340, IM-1346, IM-1349, IM-1354, IM-1358, IM-1368, IM-1369, IM-1370, IM-1371, IM-1372, IM-1191, IM-1239, IM-1246, IM-1275, IM-1278, IM-1280, IM-1291, IM-1310, IM-1324, IM-1373, IM-1374, IM-1375, IM-1376, IM-1298, IM-1312, IM-1363, IM-1377, IM-1378, IM-1379, IM-1380, IM-1381, IM-1382, IM-1383, IM-1285, IM-1304, IM-1341, IM-1355, IM-1364,  IM-1367, IM-1376, IM-1384, IM-1385, IM-1386, IM-1387, IM-1255, IM-1257, IM-1266, IM-1274, IM-1301, IM-1338, IM-1348, IM-1363, IM-1385, IM-1388, IM-1389, IM-1390, IM-1391, IM-1392, IM-1393, IM-1394, IM-1308, IM-1320, IM-1347, IM-1359, IM-1395, IM-1396, IM-1397, IM-1398, IM-1399, IM-1400, IM-1401, IM-1290, IM-1333, IM-1402, IM-1403, IM-1404, IM-1405, IM-1406, IM-1407, IM-1408, IM-1337, IM-1350, IM-1365, IM-1409, IM-1410, IM-1411, IM-1412, IM-1413, IM-1354, IM-1375, IM-1390, IM-1414, IM-1415, IM-1416, IM-1359, IM-1360, IM-1366, IM-1372, IM-1387, IM-1392, IM-1417, IM-1418, IM-1419, IM-1420, IM-1421, IM-1423, IM-1422, IM-1273, IM-1282, IM-1336, IM-1371, IM-1380, IM-1405, IM-1409, IM-1424, IM-1425, IM-1426, IM-1427, IM-1428, IM-1429, IM-1430, IM-1431, IM-1432, IM-1433, IM-1434, IM-1367, IM-1435, IM-1436, IM-1438, IM-1439, IM-1440, IM-1310, IM-1321, IM-1441, IM-1442, IM-1443, IM-1444, IM-1445, IM-1446, IM-1351, IM-1373, IM-1402,  IM-1415, IM-1447, IM-1448.

Other devices under the same FDA submission

12 devices cleared under K931009, K983373 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.