Scorpio CR Basic Femur with Posts recallZ-0466-2010 · Stryker Howmedica Osteonics Corp. · initiated Oct 2, 2009
Terminated All recalls posted in December 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was determined that a change made in the dimensions of the collar component of the packaging resulted in the inner blister material being overstretched causing a potential hole in the sterile barrier system.
FDA root cause category: Other
What the company is doing
An "Urgent Product Recall" letter was sent on October 2, 2009 by Federal Express to Stryker branches and Hospital Risk Management Departments in the United States as well as foreign consignees. The letter described the issue, potential hazard and actions for affected accounts. Consignees were instructed to examine their inventory and return recalled product as instructed in the notification letter to Stryker Orthopaedics. In addition, consignees were instructed to reconcile affected product by completing the attached Product Accountability form (with spreadsheet) and faxing to 1-201-831-6069. Direct questions to Stryker Orthopaedics by calling 1-201-972-2100.
Recall details
- Recall number
- Z-0466-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 2, 2009
- Posted by the FDA
- Dec 1, 2009
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Quantity
- 77 of total product
- Product code
- JWH
- FDA submissions
- K974556
- Distribution
- Worldwide Distribution -- United States (MA, NJ, IN, MI, CA and PA), Hong Kong, Germany, Netherlands, Japan, Korea, Switzerland, Malaysia, and the United Kingdom.
Product description
Scorpio CR Basic Femur with Posts; Stryker Orthopaedics; 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilized in total knee arthroplasty (TKA). The femoral component is attached to the bottom of the femoral bone and provides the bearing surface to the tibial insert.
Code information (models, lots, serial numbers)
Catalog number 70-3007R, Lot code MHHR5W.
Other devices under the same FDA submission
4 devices cleared under K974556 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Scorpio 70-3011LHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Scorpio 70-3011RHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Scorpio 70-4107RHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Scorpio 70-4113RHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 53588).
- Z-0467-2010 · TerminatedScorpio CR Waffle Femur Lfit with posts; Stryker Orthopaedics; 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilized in total…
- Z-0468-2010 · TerminatedScorpio PS Basic Femur; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilized in total knee arthroplast…
- Z-0469-2010 · TerminatedScorpio PS Femur Waffle Posts with LfIt; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilized in total…
- Z-0470-2010 · TerminatedScorpio NRG Cruciate Retaining Femoral, Left/Right # 5; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is ut…
- Z-0471-2010 · TerminatedScorpio NRG Cruciate Retaining Femoral, Right # 7; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilize…
- Z-0472-2010 · TerminatedScorpio NRG Cruciate Retaining Femoral, Left # 8; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilized…
- Z-0473-2010 · TerminatedScorpio NRG Posterior Stabilized Femoral, #3 Left; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilize…
- Z-0474-2010 · TerminatedScorpio NRG Posterior Stabilized Femoral. # 4 left/right; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is …
- Z-0475-2010 · TerminatedScorpio NRG Posterior Stabilized Femoral, # 5 Left; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07340. The Scorpio Femoral Components are a knee component that is utiliz…
- Z-0476-2010 · TerminatedScorpio NRG Posterior Stabilized Femoral, # 7 left/right; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is …
- Z-0477-2010 · TerminatedScorpio NRG Posterior Stabilized Femoral, #9 left; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilize…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page