Skip to content

Scorpio NRG Cruciate Retaining Femoral, Right # 7 recallZ-0471-2010 · Stryker Howmedica Osteonics Corp. · initiated Oct 2, 2009

Terminated All recalls posted in December 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It was determined that a change made in the dimensions of the collar component of the packaging resulted in the inner blister material being overstretched causing a potential hole in the sterile barrier system.

FDA root cause category: Other

What the company is doing

An "Urgent Product Recall" letter was sent on October 2, 2009 by Federal Express to Stryker branches and Hospital Risk Management Departments in the United States as well as foreign consignees. The letter described the issue, potential hazard and actions for affected accounts. Consignees were instructed to examine their inventory and return recalled product as instructed in the notification letter to Stryker Orthopaedics. In addition, consignees were instructed to reconcile affected product by completing the attached Product Accountability form (with spreadsheet) and faxing to 1-201-831-6069. Direct questions to Stryker Orthopaedics by calling 1-201-972-2100.

Recall details

Recall number
Z-0471-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 2, 2009
Posted by the FDA
Dec 1, 2009
Recalling firm
Stryker Howmedica Osteonics Corp.
Quantity
77 of total product
Product code
JWH
FDA submissions
K030978
Distribution
Worldwide Distribution -- United States (MA, NJ, IN, MI, CA and PA), Hong Kong, Germany, Netherlands, Japan, Korea, Switzerland, Malaysia, and the United Kingdom.

Product description

Scorpio NRG Cruciate Retaining Femoral, Right # 7; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilized in total knee arthroplasty (TKA). The femoral component is attached to the bottom of the femoral bone and provides the bearing surface to the tibial insert.

Code information (models, lots, serial numbers)

Catalog number 80-4407R, Lot code MHJ09J.

Other devices under the same FDA submission

11 devices cleared under K030978 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 53588).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page