Insight rev H software recallZ-1487-2012 · Sandhill Scientific, Inc · initiated Feb 23, 2009
Terminated All recalls posted in May 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sandhill Scientific, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software error may produce incorrect esophageal contraction readings.
FDA root cause category: Software change control
What the company is doing
Sandhill Scientific sent a Urgent Product Recall letter dated February 23, 2009, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were informed All Suite revision H, BioVIEW Analysis Software rev 5.5.1 are affected. Replacement of the Suite revision H, BioVIEW Analysis Software rev 5.5.1, will be provided at no cost to you. Sandhill Scientific will pay for shipping/handling charges associated with replacement of this software. We anticipate that by the first week of March, we will be able to provide you with the Software that resolves this problem. For further questions please call toll free 800-468-4556.
Recall details
- Recall number
- Z-1487-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 23, 2009
- Posted by the FDA
- May 7, 2012
- Recalling firm
- Sandhill Scientific, Inc
- Quantity
- 60 programs
- Product code
- FFX
- FDA submissions
- K012232
- Distribution
- Worldwide Distribution--USA (nationwide) including the states of CA, CO, CT, FL, IA, IL, MA, MD, MI, MN, NC, ND, NE, NM, NY, OH, OR, PA, SC, SD, TX, UT, VA and WA and the countries of Belgium, Canada, Malaysia, Mexico, New Zealand, South-Korea, Switzerland and United Kingdom.
Product description
Insight rev H software. The InSIGHT Gastrointestinal Motility System is intended for use by gastroenterologists, surgeons, and medically trained personnel for esophageal, biliary, and anorectal studies as an aid in documenting and diagnosing digestive motility disorders.
Code information (models, lots, serial numbers)
S98-1925-2H; S98-1925-4H All lots and shipments of the software.
Other devices under the same FDA submission
3 devices cleared under K012232 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Zvu® ZVU-3Diversatek Healthcare, Inc. · System, Gastrointestinal Motility (Electrical)1 recall0 reports
- inSIGHT Ultima H12R-2000Diversatek Healthcare, Inc. · System, Gastrointestinal Motility (Electrical)1 report
- ZepHr Z07-2000-B-BDiversatek Healthcare, Inc. · System, Gastrointestinal Motility (Electrical)1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page