Skip to content

ZepHr Z07-2000-B-BRecalls & reported problems

Diversatek Healthcare, Inc. · ZepHr® Z/pH Recorder · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00816734022177), received between Sep 5, 2024 and Sep 5, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K012232); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 5, 2024No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information

    CUSTOMER USER FACILITY REPORT CONTAINED THE FOLLOWING INFORMATION: DESCRIBE THE EVENT OR PROBLEM: THE IMPEDANCE MACHINE WOULD NOT FUNCTION, AND REGISTERED NURSE (RN) DID NOT HAVE A REPLACEMENT MACHINE TO USE, SO THE IMPEDANCE CATHETER WAS NOT PLACED AS PLANNED. VENDOR WAS CALLED TO LET THEM KNOW THAT THE MACHINE NEEDS REPAIRED. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? UPPER GASTROINTESTINAL ENDOSCOPY. WHAT PROBLEM DID THE USER HAVE (CHECK ALL THAT APPLY): DEVICE FAILED (E.G. BROKE, COULDN'T GET IT TO WORK OR STOPPED WORKING)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20146779 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K012232. One submission often covers many models, so check the product and lot details in each record.

  • Z-2521-2023 · Jul 25, 2023 · Open, Classified
    ZVU Functional GI Software, REF: ZVU-3
  • Z-1487-2012 · Feb 23, 2009 · Terminated
    Insight rev H software. The InSIGHT Gastrointestinal Motility System is intended for use by gastroenterologists, surgeons, and medically trained personnel for esophageal, biliary

Other System, Gastrointestinal Motility (Electrical) devices

All System, Gastrointestinal Motility (Electrical) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.