Dimension Vista V-LYTE Fluids: Diluent catalog number K835 recallZ-1039-2010 · Siemens Healthcare Diagnostics, Inc. · initiated Oct 22, 2009

Terminated All recalls posted in March 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Results can elevate after sensor replacement. Sodium urine results can be falsely elevated 15 mmol/L within 24 hours after installation of a new sensor.

FDA root cause category: Other

What the company is doing

The recalling firm issued an Urgent Field Safety Notice dated October 2009 to all affected customers. Customers were given instruction which minimize urine specimen use and report urine specimen results when QC is acceptable. Sensors are to be replaced when urine QC is out. Replacement sensors are available by contacting the firm at 800-241-0420. Please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.

Recall details

Recall number
Z-1039-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 22, 2009
Posted by the FDA
Mar 10, 2010
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
150 units
Product code
JJE
FDA submissions
K051087
Distribution
The product was shipped to medical facilities nationwide. The product was also shipped to Australia, Canada, Belgium, France, Germany, Italy, Japan, Malaysia, New Zealand, Netherlands, Portugal, Spain, South Korea, and Switzerland.

Product description

Dimension Vista V-LYTE Fluids: Diluent catalog number K835

Code information (models, lots, serial numbers)

Catalog number K835

Other devices under the same FDA submission

4 devices cleared under K051087 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 53686).