BD Q-Syte Extension Set, 15cm Std Bore, P.V. 1.14 ml… recallZ-0751-2010 · Becton Dickinson Medical Systems · initiated Oct 19, 2009
Terminated All recalls posted in February 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson Medical Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
FDA root cause category: Process control
What the company is doing
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall details
- Recall number
- Z-0751-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 19, 2009
- Posted by the FDA
- Feb 2, 2010
- Recalling firm
- Becton Dickinson Medical Systems
- Quantity
- 42,600 units
- Product code
- FPA
- FDA submissions
- K013621
- Distribution
- Nationwide distribution.
Product description
BD Q-Syte Extension Set, 15cm Std Bore, P.V. 1.14 ml Luer-Lok, REF 385101, Becton Dickinson Infusion Therapy Systems Inc.
Code information (models, lots, serial numbers)
Lot numbers: 8305510, 8331944.
Other devices under the same FDA submission
5 devices cleared under K013621 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BD Q-Syte 385100Becton, Dickinson And Company · Set, administration, intravascular1 death971 reports
- BD Q-Syte 385102Becton, Dickinson And Company · Set, administration, intravascular269 reports
- BD Q-Syte 385101Becton, Dickinson And Company · Set, administration, intravascular27 reports
- BD Q-Syte 385108Becton, Dickinson And Company · Set, administration, intravascular18 reports
- Q-Syte 385103Becton, Dickinson And Company · Set, administration, intravascular4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 53749).
- Z-0750-2010 · TerminatedBD Q-Syte, Luer Access Split-Septum, 0.10 ml, REF 385100, Becton Dickinson Infusion Therapy Systems Inc.
- Z-0752-2010 · TerminatedBD Q-Syte Extension Set, 15 cm, P.V. 0.21 ml, REF 385102, Becton Dickinson Infusion Therapy Systems Inc.
- Z-0753-2010 · TerminatedBD Q-Syte Vial Access Adapter, 0.16 ml, REF 385108, Becton Dickinson Infusion Therapy Systems Inc.
- Z-0754-2010 · TerminatedMPS Acacia Extension Set, 15cm RightBore-18, Spin Nut w/BBD Q-Syte, REF 385150,
- Z-0755-2010 · TerminatedMPS Acacia Extension Set, 15cm Small Bore, Spin Nut w/BD Q-Syte, REF 385151
- Z-0756-2010 · TerminatedMPS Acacia Tri-Extension Set, 15cm Small Bore, Spin Nut w/BD Q-Syte, REF 385164
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page