Dimension Vista LOCI Reaction Vessel recallZ-0980-2010 · Siemens Healthcare Diagnostics, Inc. · initiated Nov 24, 2009

Terminated All recalls posted in March 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some vessels may leak. If leakage occurs, patient results will be flagged and are not reportable.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The recalling firm issued an Urgent Field Safety Notice dated 11/2009 to all affected customers. Customers have been told to discard any bags of the affected lot. If the affected lot is the only lot the customer has, customers can examine a vessel from the bag. If the vessel mold cavity identity is B8, customers should discard the bag. Thus customers can use the lot if the cavity mold is anything other than B8. In addition, customers have been told to empty the LOCI vessel holder on all systems and to discard vessels in case any affected vessels have been loaded onto the system. For technical questions or concerns, please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.

Recall details

Recall number
Z-0980-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 24, 2009
Posted by the FDA
Mar 5, 2010
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
408 units
Product code
JJE
FDA submissions
K051087
Distribution
The device was shipped to medical facilities in CA, CO, FL, GA, IL, LA, MD, MI, MS, NE, NC, NJ, NY, OH, SC, TN, TX, VA, WA, and WI. The product was also shipped to Belgium, France, Germany, Italy, Netherlands, Portugal, Spain, and Switzerland.

Product description

Dimension Vista LOCI Reaction Vessel

Code information (models, lots, serial numbers)

Catalog number KS855. Lot number NH29-188-09

Other devices under the same FDA submission

4 devices cleared under K051087 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.