Skip to content

CADD-LEGACY 1 Model 6400, Ambulatory Infusion Pump recallZ-1235-2010 · Smiths Medical ASD, Inc. · initiated Dec 23, 2009

Terminated All recalls posted in April 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The CADD-Legacy 1 Model 6400 Ambulatory Infusion Pump (Serial Number 370969) was shipped without documentation that the required accuracy testing had been performed during the manufacturing process. If the subject infusion pump did not go through the accuracy testing, and if it does not meet accuracy specifications, there is a potential for over or under-delivery.

FDA root cause category: Employee error

What the company is doing

Smiths Medical telephoned the customer on December 23, 2009, to return the CADD-Legacy 1 Ambulatory Infusion Pump, serial number 370969, since there was no record of the completed accuracy testing for the infusion pump.

Recall details

Recall number
Z-1235-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 23, 2009
Posted by the FDA
Apr 5, 2010
Recalling firm
Smiths Medical ASD, Inc.
Quantity
1
Product code
FRN
FDA submissions
K982838
Distribution
MI

Product description

CADD-LEGACY 1 Model 6400, Ambulatory Infusion Pump, REF 21-6400-51, Rx Only, Smiths Medical MD, Inc.., St. Paul, MN 55112.

Code information (models, lots, serial numbers)

Serial number: 370969

Other devices under the same FDA submission

1 device cleared under K982838 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page