CADD 21-6400-51Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of CADD (123)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This CADD infusion pump from ICU Medical is used to deliver fluids or medication over time. Reports mostly describe alarms that interrupt infusions, under- or over-delivery of medication, and air bubbles in the tubing. There are 7,541 reports, including 7 deaths and 33 injuries. No recalls name this device. Reports fell to 127 in the last 12 months from 327 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 7,543
- Deaths reported
- 7
- Injuries reported
- 33
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7,543 adverse event reports that list this exact device (DI 10610586019548), received between Apr 17, 2017 and Sep 28, 2026 — classified as 7 deaths, 33 injuries, 7,503 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (3,307 reports).
2 other recalls cover devices cleared under the same FDA submission (K982838); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Alarm System3,307
- Insufficient Flow or Under Infusion583
- Inaccurate Delivery432
- Output Problem426
- Infusion or Flow Problem403
- Audible Prompt/Feedback Problem364
- Volume Accuracy Problem337
- Excess Flow or Over-Infusion326
Patient problems
- No Clinical Signs, Symptoms or Conditions3,806
- No Patient Involvement1,821
- No Information906
- No Consequences Or Impact To Patient650
- No Known Impact Or Consequence To Patient521
- Insufficient Information514
- Headache24
- Nausea18
Recent adverse event reports
The 25 most recent of 7,543 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Insufficient Flow or Under Infusion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50925452 ↗MalfunctionSep 22, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859642 ↗MalfunctionSep 17, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50809925 ↗MalfunctionSep 15, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50779272 ↗MalfunctionAug 25, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50439251 ↗MalfunctionAug 18, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50305570 ↗MalfunctionAug 11, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50209626 ↗MalfunctionJun 26, 2026Insufficient Flow or Under Infusion; Failure to Infuse
IT WAS REPORTED THAT THE CADD-LEGACY ONE AMBULATORY INFUSION PUMP EXPERIENCED AN NDPWR (NO DISPOSABLE ALARM - PUMP WON'T RUN) ALARM DURING INFUSION, RESULTING IN UNDER-DELIVERY OF 5-FU (5-FLUOROURACIL). THE PATIENT DELAYED REPORTING THE EVENT FOR MORE THAN 8 HOURS, AND NO TROUBLESHOOTING WAS PERFORMED. THE PROGRAMMED RATE WAS 52 ML PER 24 HOURS, AND REMAINING VOLUME WAS 81.5 ML. THERE WERE PATIENT INVOLVEMENT AND NO HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25636439 ↗MalfunctionJun 5, 2026Failure to Deliver
IT WAS REPORTED THAT THE PUMP ALARMED WITH A ¿NO DISPOSABLE¿ ALARM DURING USE. THE PATIENT ATTEMPTED TROUBLESHOOTING BY CLAMPING THE TUBING AND REMOVING AND REINSERTING THE CASSETTE TWICE; HOWEVER, THE ALARM PERSISTED, AND AIR BUBBLES WERE NOTED. THE DEVICE WAS REPORTED TO UNDER-DELIVER. THE EVENT OCCURRED WHILE THE DEVICE WAS IN USE WITH A PATIENT IN THE HOME SETTING. THERE WAS PATIENT INVOLVEMENT; HOWEVER, NO HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25415102 ↗MalfunctionMay 28, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THE PUMP EXHIBITED A NO DISPOSABLE AND CLAMP TUBING ALARM AND UNDER INFUSED. THERE WAS PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25322605 ↗MalfunctionMay 20, 2026Inaccurate Delivery
IT WAS REPORTED THAT THE DEVICE WAS DELIVERING INCORRECT VOLUME IN CONTINUOUS MODE AND IT REQUIRED CALIBRATION. THERE WAS UNKNOWN PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25229961 ↗MalfunctionMay 19, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THE DEVICE HAD INCORRECT VOLUME DELIVERED IN CONTINUOUS MODE AND REQUIRED CALIBRATION. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25216628 ↗MalfunctionMay 18, 2026Failure to Infuse
IT WAS REPORTED THAT THE DEVICE WAS SHOWING LAST ERROR CODE 1871 INDICATING A MOTOR TIMEOUT. PER REPORTER, THE ISSUE WAS DISCOVERED DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25202616 ↗MalfunctionMay 11, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THE PUMP WAS UNDER INFUSING AND THE SCREEN WAS FAILING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25135472 ↗MalfunctionMay 5, 2026Insufficient Flow or Under Infusion
AN EVENT INVOLVING A CADD PUMP WAS REPORTED IN WHICH UNDER DELIVERY WAS IDENTIFIED. THE PUMP WAS PROGRAMMED TO INFUSE 110ML AT A RATE OF 0.6 ML/HOUR AND INDICATED THAT 100.5 ML HAD BEEN DELIVERED WITH 9.5 ML REMAINING. HOWEVER, THE REGISTERED NURSE OBSERVED THAT THE MEDICATION BAG CONTAINED MORE THAN THE INDICATED REMAINING VOLUME. PHARMACY REVIEW CONFIRMED THAT 27.3 ML OF BLINATUMOMAB REMAINED IN THE BAG. THERE WAS PATIENT INVOLVEMENT AND NO HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25081423 ↗MalfunctionMay 5, 2026Excess Flow or Over-Infusion
AN EVENT INVOLVING A CADD PUMP WAS REPORTED IN WHICH OVER DELIVERY WAS SUSPECTED. THE PATIENT STATED THE PUMP RAN SLOWER THAN PROGRAMMED. BASED ON CALCULATIONS, MORE VOLUME REMAINED THAN EXPECTED AT DISCONNECTION. THERE WAS PATIENT INVOLVEMENT AND NO HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25081363 ↗MalfunctionMay 5, 2026Collapse
IT WAS REPORTED THAT THE PUMP SHOWS "SERVICE NEEDED MESSAGE" AND HAD A BAD SCREEN. THERE WAS NO PATIENT INVOLVEMENT AND HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25081230 ↗MalfunctionMay 4, 2026Device Sensing Problem
IT WAS REPORTED THAT THE PUMP DISPLAYS LAST ERROR CODE 1310. THERE WAS NO PATIENT INVOLVEMENT AND THE ISSUE WAS DISCOVERED DURING TESTING. EVALUATION OF THE RETURNED DEVICE IDENTIFIED A FAULTY DOWNSTREAM OCCLUSION SENSOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25070367 ↗MalfunctionMay 4, 2026Failure to Infuse
THE CUSTOMER WAS REPORTED THAT THE DEVICE HAD AN EVENT OF CONSTANT BEEPING DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT/ NO HARM REPORTED. EVALUATION OF THE RETURNED DEVICE IDENTIFIED THE DEFECTIVE DOWNSTREAM OCCLUSION (DSO) SENSOR WHICH MAY LEAD TO INTERRUPTION OF MEDICATION DURING USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25069790 ↗MalfunctionApr 21, 2026Improper Flow or Infusion
IT WAS REPORTED THAT THE PUMP EXPERIENCED ERROR CODE 1836 INDICATING A MOTOR FAILURE. PER REPORTER THE BATTERIES WERE REMOVED AND REINSERTED BUT THE ALARM RETURNED. THE DEVICE WAS THEN POWERED OFF. THE PROGRAMMED RATE WAS 4.3 ML PER HOUR, THE REMAINING VOLUME WAS 175.3 ML, AND VOLUME GIVEN WAS 24.7 ML. THE EVENT TOOK PLACE AT THE PATIENT¿S HOME. NO PATIENT HARM/ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24940059 ↗MalfunctionApr 17, 2026Improper Flow or Infusion
IT WAS REPORTED THAT THE PUMP SHOWED ERROR CODE 1870 INDICATING A MOTOR TIMEOUT. PER REPORTER, THE SETTINGS WERE ERASED FOLLOWED BY THE SECOND ALARM. THE MEDICATION WAS INFUSED AT RATE OF 2.2 ML PER HOUR. THE REMAINING VOLUME WAS 73.6 AND GIVEN VOLUME WAS 25 ML. THE EVENT OCCURRED DURING INFUSION AT THE PATIENT'S HOME. NO ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24916513 ↗MalfunctionApr 17, 2026Volume Accuracy Problem
IT WAS REPORTED THAT THE DEVICE HAD INCORRECT VOLUME DELIVERED IN CONTINUOUS MODE REQUIRES CALIBRATION. THERE WAS UNKNOWN REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24912161 ↗MalfunctionApr 17, 2026Volume Accuracy Problem
IT WAS REPORTED THAT THE DEVICE HAD INCORRECT VOLUME DELIVERED IN CONTINUOUS MODE REQUIRES CALIBRATION. THERE WAS UNKNOWN REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24912147 ↗MalfunctionApr 17, 2026Volume Accuracy Problem
IT WAS REPORTED THAT THE DEVICE HAD INCORRECT VOLUME DELIVERED IN CONTINUOUS MODE REQUIRES CALIBRATION. THERE WAS UNKNOWN REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24912135 ↗MalfunctionApr 17, 2026Volume Accuracy Problem
IT WAS REPORTED THAT THAT THE DEVICE HAD INCORRECT VOLUME DELIVERED IN CONTINUOUS MODE REQUIRES CALIBRATION. THERE WAS UNKNOWN REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24912118 ↗
FDA clearance and classification
Cleared through 510(k) K982838 on Nov 4, 1998 (decision: Substantially Equivalent). Submitted by Sims Deltec, Inc. as “Cadd-Legacy 1 Ambulatory Infusion System Model Number 6400”. The FDA review took 84 days. FDA record ↗
The FDA regulates product code FRN (pump, infusion) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K982838. One submission often covers many models, so check the product and lot details in each record.
- Z-0307-2018 · Nov 22, 2017 · TerminatedCADD-Legacy 1, Model 6400 Ambulatory Infusion Pump, REF 21-6400-51, PC 10008552-001, Smiths Medical ASD, Inc. Rx ONLY
- Z-1235-2010 · Dec 23, 2009 · TerminatedCADD-LEGACY 1 Model 6400, Ambulatory Infusion Pump, REF 21-6400-51, Rx Only, Smiths Medical MD, Inc.., St. Paul, MN 55112.
Common questions about the CADD 21-6400-51
Has the CADD 21-6400-51 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10610586019548), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the CADD 21-6400-51?
The FDA's MAUDE database lists 7,543 adverse event reports for this device, received between Apr 17, 2017 and Sep 28, 2026. In those reports, the most common specific device problems are “Device Alarm System” (3,307), “Insufficient Flow or Under Infusion” (583) and “Inaccurate Delivery” (432); the most common patient problems are “Headache” (24) and “Nausea” (18). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CADD 21-6400-51?
7 reports are classified as deaths, 33 as injuries and 7,503 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CADD 21-6400-51 FDA approved?
It was cleared through 510(k) K982838 on Nov 4, 1998. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the CADD 21-6400-51, and is it still sold?
The FDA's device database lists Icu Medical, Inc. as the company responsible for it (brand CADD, model 21-6400-51). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Pump, infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,412 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,079 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths50,284 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,947 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,545 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion3 recalls6 deaths13,013 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,232 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page