Avanta Fluid Management Injection System, Avanta… recallZ-1199-2010 · Medrad Inc · initiated Jan 6, 2010

Terminated All recalls posted in March 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medrad Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The product may be defective resulting in a reduction of the saline delivery rate and inadequate air purging.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The recalling firm issued an Urgent Medical Device Recall letter dated 1/6/10 to all customers that received the affected lots. The customers were instructed to determine whether or not they had the affected product lots. Customers are to discontinue use and contact MEDRAD Customer Support or their MIP Rep to make arrangements for return of the recalled product. Any affected product returned will be replaced at no charge to the customer. Customers are also to complete and sign the attached form and fax it back to the firm. Questions concerning this recall are to be directed to Deb Easler, Product Manager, at 412-767-2400.

Recall details

Recall number
Z-1199-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jan 6, 2010
Posted by the FDA
Mar 23, 2010
Recalling firm
Medrad Inc
Quantity
2625 pieces
Product code
DXT
FDA submissions
K050456
Distribution
Worldwide Distribution -- USA, including stated of AL, AZ, CA, DC, FL, GA, IL, KS, MD, MO, NM, NY, OH,OK, PA, TN, TX, and WI. and countries of Belgium, Germany, and Italy.

Product description

Avanta Fluid Management Injection System, Avanta Multi-Patient Disposable Set (MPAT) High Pressure Check Valve, Catalog number AVA 500 MPAT, Material number 3018231.

Code information (models, lots, serial numbers)

Lot numbers 820007 and 820008

Other devices under the same FDA submission

3 devices cleared under K050456 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.