i-STAT EG7+ cartridges recallZ-1793-2010 · Abbott Point of Care Inc. · initiated Feb 5, 2010
Terminated All recalls posted in June 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Point of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Abbott has determined that there is a dependence of the i-STAT Hct assay on the vertical pitch angle of the i-STAT Analyzer during the blood analysis cycle.
FDA root cause category: Other
What the company is doing
The firm, Abbott Point of Care, sent an "Urgent Field Safety Notice Product Recall" letter dated January 2010 to the customers. Safety Alert /Product Recall letters were also sent by Federal Express commencing on February 5, 2010. The letter describes the product, problem and the action to be taken by the customers. The customers were instructed to replace the i-STAT Hematocrit/HCT and Calculated Hemoglobin/HB /Cartridge and Test Information (CT) Sheet found in their i-STAT System manual with those included in the mailing , and to provide a copy of the letter another facility, if they have forwarded any cartridges. If you should have any questions regarding this information, please contact Abbott Point of Care Technical Support at 800-366-8020 or your Abbott Point of Care representative.
Recall details
- Recall number
- Z-1793-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 5, 2010
- Posted by the FDA
- Jun 11, 2010
- Recalling firm
- Abbott Point of Care Inc.
- Quantity
- 40,446 total cartridges, all varieties
- Product code
- JPI
- FDA submissions
- K940918
- Distribution
- Worldwide distribution: USA and Argentina; Austria; Bahamas; Bolivia; Brazil; Chile; Hong Kong; China; China; Columbia; Costa Rica; Denmark; Dominican Republic; Ecuador; El Salvador; Finland; Georgia; Greece; Honduras; India; India; Ireland; Israel; Italy; Japan; Kuwait; Mauritius; Mexico; Norway; Oman; Panama; Paraguay; Peru; Poland; Portugal; Qatar; Russian Federation; Saudi Arabia; Singapore; South Africa; South Korea; Spain; Switzerland; Taiwan; Thailand; Trinidad and Tobago; Turkey; United Arab Emirates; Uruguay; Yemen; Netherlands Antilles; Venezuela; Australia; Belgium; France; Germany; Italy; New Zealand; Spain; Sweden; and The Netherlands.
Product description
i-STAT EG7+ cartridges The i-stat Hematocrit (Hct) is intended for use in the in vitro quantification of packed red blood cell volume in arterial, venous, or capillary whole blood. Hematocrit is a key indicator of the body's state of hydration, anemia, or severe blood loss, as well as the blood's ability to transport oxygen.
Code information (models, lots, serial numbers)
List # 06F01-01, 02
Other devices under the same FDA submission
5 devices cleared under K940918 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- i-STAT 03P88-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH105 reports
- i-STAT 03P76-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall46 reports
- i-STAT 03P77-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall10 reports
- i-STAT 03P79-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH23 reports
- i-STAT 03P78-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH16 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 54652).
- Z-1791-2010 · Terminatedi-STAT CHEM8+ cartridges The i-stat Hematocrit (Hct) is intended for use in the in vitro quantification of packed red blood cell volume in arterial, venous, or capillary whole blo…
- Z-1794-2010 · Terminatedi-STAT EG6+ cartridges The i-stat Hematocrit (Hct) is intended for use in the in vitro quantification of packed red blood cell volume in arterial, venous, or capillary whole blood…
- Z-1796-2010 · Terminatedi-STAT 6+ cartridges The i-stat Hematocrit (Hct) is intended for use in the in vitro quantification of packed red blood cell volume in arterial, venous, or capillary whole blood. …
- Z-1797-2010 · Terminatedi-STAT EC4+ cartridges The i-stat Hematocrit (Hct) is intended for use in the in vitro quantification of packed red blood cell volume in arterial, venous, or capillary whole blood…
- Z-1798-2010 · Terminatedi-STAT E3+ cartridges The i-stat Hematocrit (Hct) is intended for use in the in vitro quantification of packed red blood cell volume in arterial, venous, or capillary whole blood.…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page