Boston Scientific CONFIENT, model E030, Implantable… recallZ-1555-2010 · Boston Scientific CRM Corp · initiated Mar 15, 2010
Terminated All recalls posted in May 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific CRM Corp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Boston Scientific Corporation has stopped shipment and is retrieving unimplanted devices from US hospital shelves of all its implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) after determining that some manufacturing process changes were not submitted for approval to the US Food & Drug Administration.
FDA root cause category: PMA
What the company is doing
BSC issues press release dated 15 March 2010. FDA press release posted 18 March 2010. On 15 March 2010, BSC Sales Representatives begin removal of devices from health care facilities stock. Starting 22 March 2010 BSC sent a physician letter dated 22 March 2010 to implanting and following physicians via FedEx. The Urgent Medical Device Recall letter described the issue of why devices were being removed from hospital shelves. The letter did NOT recommend any changes to normal follow-up procedures for patients implanted with an ICD or CRT-D device. The letter included ALL potential US models. To be complete and mitigate the possibility of retrieval errors, the ICD and CRT-D US models that may be on customer's shelves were included in the letter. The devices subject to retrieval, however, were those non-implanted devices with domestic customers. For questions, please contact the local representative (1-800-227-3422) for pick up and return or contact Boston Scientific Technical Services at 1.800.CARDIAC (227.3422).
Recall details
- Recall number
- Z-1555-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 15, 2010
- Posted by the FDA
- May 8, 2010
- Recalling firm
- Boston Scientific CRM Corp
- Quantity
- 55
- Product code
- LWP
- FDA submissions
- P960040
- Distribution
- nationwide within US only (OUS distribution not affected).
Product description
Boston Scientific CONFIENT, model E030, Implantable Cardioverter Defibrillator (ICD). The CONFIENT Implantable Cardioverter Defibrillator (ICDs) are designed to detect and terminate ventricular tachycardia (VT) and ventricular fibrillation(VF) and provide bradycardia therapy (atrial and ventricular pacing). Therapies include both low- and high-energy shocks using either a biphasic or monophasic waveform. Sterilized using ethylene oxide. Boston Scientific, 4100 Hamline Avenue, St. Paul, MN 55112-5798. USA.
Code information (models, lots, serial numbers)
Model E030, serial numbers: 504800, 505054, 505073, 506067, 506886, 506961, 507219, 507236, 507255, 507268, 507292, 507304, 507332, 507372, 507381, 507388, 507412, 507424, 507446, 507448, 507449, 507454, 507467, 507527, 507538, 507577, 507594, 507600, 507607, 507614, 507618, 507625, 507728, 507742, 507757, 507785, 507829, 507840, 507882, 507898, 507903, 507904, 507923, 507940, 507943, 507944, 507952, 507960, 507963, 507966, 507984, 508001, 508008, 508059, 508130.
Other devices under the same FDA submission
45 devices cleared under P960040 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- DYNAGEN™ EL ICD DR D152Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)12 deaths1,208 reports
- INOGEN™ EL ICD DR D142Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)6 deaths1,017 reports
- RESONATE™ EL ICD DR D433Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths965 reports
- DYNAGEN™ EL ICD DR D153Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths886 reports
- MOMENTUM™ EL ICD DR D121Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 deaths850 reports
- DYNAGEN™ EL ICD VR D150Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)8 deaths831 reports
- INOGEN™ EL ICD VR D140Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths823 reports
- ENERGEN™ ICD E140Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)3 deaths819 reports
- VIGILANT™ EL ICD VR D232Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)8 deaths750 reports
- ENERGEN™ ICD E142Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)1 death705 reports
- INCEPTA™ ICD E160Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 deaths601 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 55216).
- Z-1556-2010 · TerminatedBoston Scientific LIVIAN Cardiac Resynchronization Therapy Defibrillator (CRT-D), models H220, H225, H227, H229. The LIVIAN family of cardiac resynchronization therapy defibrillat…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page